GAMMA 3425-1380SRecalls & reported problems
Stryker GmbH · Long Nail Kit R1.5, Ti, RIGHT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613252273844), received between Mar 18, 2019 and Jun 16, 2026 — classified as 15 injuries.
The most frequently reported specific device problem is “Fracture” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture13
- Break10
- Device Damaged by Another Device2
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- Failure of Implant11
- Fall2
- Implant Pain2
- Pain2
- Ambulation or Postural Difficulties1
- Impaired Healing1
- Injury1
- Insufficient Information1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryJun 16, 2026Fracture
AS REPORTED: "A (B)(6) PATIENT, INITIALLY TREATED ON (B)(6) 2026. NO IMPLANTATION ERRORS DETECTED; NAIL FRACTURE AT A BODY WEIGHT OF APPROXIMATELY 70 KG AS EARLY AS (B)(6), I.E. BARELY 8 WEEKS POST-OPERATION. REVISION SURGERY WITH A PROSTHESIS".
Patient problems: Insufficient Information
MAUDE report 25521165 ↗InjuryApr 1, 2025Break; Fracture; Break; Fracture
AS REPORTED: "BROKEN GAMMA 3 NAIL, IMPLANTED (B)(6) 2024, REMOVED TODAY. PRIMARY SURGERY (B)(6) 2024 ANNA-KAJSA HARDING, NO PROBLEMS DURING SURGERY LAB TESTS, NOTHING DEVIANT HYPERTONI, ARTRIAL FIBRILLATION FELL TWICE BEFORE SHE SOUGHT MEDICAL ATTENTION AGAIN."
Patient problems: Fall
MAUDE report 21739997 ↗InjurySep 5, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE IMPLANTATION TOOK PLACE ON (B)(6) 2023 WITHOUT PROBLEMS. IN (B)(6) 2023, A NAIL FRACTURE OCCURRED DURING REHAB.
Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture
MAUDE report 20153763 ↗InjuryAug 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PRESENT SWISS TRAUMA FRACTURE REGISTRY (SWTR) IS A PROSPECTIVE, MULTI-CENTER, OBSERVATIONAL STUDY TO COLLECT DATA ON ALL STRYKER TRAUMA AND EXTREMITIES PRODUCTS WITH THE AIM TO ADD TO THE EXISTING CLINICAL EVIDENCE TO SUPPORT THE SAFETY AND PERFORMANCE OF STRYKER IMPLANTS FOR THE APPROVED INDICATIONS FOR USE. THE SWTR CAPTURES DATA EVALUATING THE REAL-WORLD USAGE OF STRYKER IMPLANTS AT TRAUMA CENTERS WITH THE CAPACITY, TRAINING, AND EXPERIENCE TO GATHER CLINICALLY RELEVANT STANDARD OF CARE SAFETY AND PERFORMANCE ENDPOINT DATA. DURING THE REVIEW OF THE REPORT, DELAYED UNION WAS NOTED FOR THIS PATIENT DURING THE FOLLOW-UP VISIT.
Patient problems: Impaired Healing
MAUDE report 19938446 ↗InjuryJul 30, 2024Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device
AS REPORTED: ''BROKEN GAMMA3 LONG NAIL. REVISION WAS ON (B)(6) 2024'."
Patient problems: Failure of Implant
MAUDE report 19860935 ↗InjuryFeb 27, 2024Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device
AS REPORTED: THE PATIENT'S ATTORNEY CONTACTED STRYKER WITH THE FOLLOWING ISSUE: THE PATIENT WAS OPERATED ON AT THE BEGINNING OF (B)(6) 2023 FOR A FRACTURE OF THE FEMORAL NECK AND A GAMMA NAIL WAS INSERTED. ON (B)(6) 2023, THE GAMMA NAIL BROKE IN THE AREA OF THE RIGHT THIGH. THE PATIENT SUBSEQUENTLY HAD AN INPATIENT STAY FROM (B)(6) 2023 TO (B)(6) 2023.
Patient problems: Failure of Implant
MAUDE report 18788837 ↗InjuryJan 9, 2024Break; Fracture; Break; Fracture
AS REPORTED BY THE HOSPITAL RISK MANAGER: "THE PATIENT REPORTS HAVING BEEN OPERATED ON (B)(6) 2023, IN AT ANOTHER HOSPITAL, FOR A FRACTURE OF THE PROXIMAL FEMUR WITH A GAMMA 3 LONG NAIL. IN (B)(6) 2023, THE PATIENT WENT TO ANOTHER EMERGENCY ROOM AND FOUND THAT THE FEMORAL FIXATION DEVICE HAD BROKEN. THEY WENT TO OUR HOSPITAL FOR THE SURGICAL REMOVAL AND NEW OSTEOSYNTHESIS OF THE LEFT FEMUR."
Patient problems: Failure of Implant
MAUDE report 18473738 ↗InjuryOct 25, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE PATIENT HAD A LONG GAMMA NAIL IN HER RIGHT FEMUR, WHICH BROKE SPONTANEOUSLY DURING REHABILITATION, NECESSITATING A RETURN TO THE OPERATING THEATRE. ACTIONS TAKEN IN THE CARE FACILITY TO MANAGE THE PATIENT: RETURN TO THE OPERATING ROOM TO REMOVE THE MATERIAL, THEN FITTING OF A NEW OSTEOSYNTHESIS.
Patient problems: Failure of Implant
MAUDE report 18001962 ↗InjuryJan 3, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THERE WAS A BROKEN GAMMA NAIL 6 MONTHS AFTER IMPLANTATION. THE PATIENT WAS REVISED TO A RECON NAIL.
Patient problems: Failure of Implant
MAUDE report 16086446 ↗InjuryNov 24, 2021Break; Break
AS REPORTED: "THE PATIENT CAME IN WITH PAIN. FRACTURE OF GAMMA -NAIL LONG AT FEMORAL NECK SCREW, NO NEW FALL. THE PATIENT REQUIRED A REVISION SURGERY."
Patient problems: Failure of Implant; Implant Pain
MAUDE report 12867646 ↗InjurySep 24, 2021Fracture; Fracture
THE CUSTOMER REPORTED THERE WAS A SPONTANEOUS FRACTURE OF THE IMPLANTABLE MEDICAL DEVICE. REVISION SURGERY WAS REQUIRED TO PUT THE NAIL BACK IN PLACE.
Patient problems: Failure of Implant
MAUDE report 12530129 ↗InjuryDec 12, 2020Fracture; Fracture
THE CUSTOMER REPORTED THAT THE GAMMA 3 FEMORAL NAIL (RIGHT) BROKE AT PROXIMAL END IN-SITU, JUST ABOVE LAG SCREW HOLE. THE NAIL HAS SUBSEQUENTLY SUNK AND CAUSED THE DISTAL LOCKING SCREWS TO SNAP. THIS NAIL HAS BEEN SUCCESSFULLY EXTRACTED AND REVISED.
Patient problems: Failure of Implant; Fall; Implant Pain
MAUDE report 10996126 ↗InjurySep 2, 2020Fracture; Fracture
AS REPORTED: "ON (B)(6) 2020, A PROPHYLACTIC NAILING WAS REALIZED IN CASE OF IMMINENT FRACTURE. A NAIL FRACTURE WAS CONFIRMED AFTER METAL REMOVAL ON (B)(6) 2020."
Patient problems: Failure of Implant; Injury
MAUDE report 10484701 ↗InjuryJun 7, 2019Fracture; Material Deformation; Break; Fracture; Material Deformation
THE CUSTOMER REPORTED THAT A BROKEN GAMM NAIL IS BEING REMOVED. THIS WAS IMPLANTED DURING JANUARY 2019 AND THE SURGEON REPORTED THAT IT HAS BROKEN AROUND THE LAG SCREW. THE PATIENT WAS WALKING A GOT A SUDDEN SHARP PAIN AND WAS UNABLE TO WALK.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 8680476 ↗InjuryMar 18, 2019Break; Fracture; Fracture
A NAIL WAS PLACED IN A PATIENT AT (B)(6) 2018. THE ELONGATED NAIL WAS PLACED IN A (B)(6) YEAR OLD FEMALE PATIENT. SINCE THE FIRST WEEK OF JANUARY SHE DEVELOPED SEVERE PAIN IN HER RIGHT HIP. ON (B)(6) 2019, IT SEEMED VIA X-RAY THAT THE NAIL HAS BEEN BROKEN. ON (B)(6) 2019, SURGEON PERFORMED EXPLANT OF BROKEN NAIL AND NEW NAIL WAS IMPLANTED.
Patient problems: Failure of Implant; Pain
MAUDE report 8430074 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page