GAMMA 3525-1320SRecalls & reported problems
Stryker GmbH · Long Nail Kit R1.5, Ti, LEFT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613252274261), received between Apr 22, 2020 and Sep 17, 2025 — classified as 5 injuries.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Break4
- Device Damaged by Another Device2
Patient problems
- Failure of Implant3
- Implant Pain2
- Injury1
- Insufficient Information1
- Nonunion/Delayed-union Bone Fracture1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2025Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device
AS REPORTED: "THE GAMMA3 NAIL BROKE."
Patient problems: Insufficient Information
MAUDE report 23076717 ↗InjuryJan 28, 2025Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device
THE CUSTOMER REPORTED THAT: "THE PATIENT UNDERWENT SURGERY ON (B)(6) 2024 FOR A GAMMA NAIL OF THE LEFT FEMUR (STRYKER). SINCE THE AFTERNOON OF (B)(6) 2024, SHE HAD BEEN EXPERIENCING PROGRESSIVE PAIN IN HER LEFT HIP. ON THE NIGHT OF (B)(6) 2024, DURING A TRANSFER TO THE TOILET, THE PATIENT DESCRIBED A ROTATING MOVEMENT OF THE HIP WITH AN AUDIBLE AN AUDIBLE CRACK. THE PAIN MADE IT IMPOSSIBLE TO STAND ON THE FLOOR. DURING SURGERY ON (B)(6) 2024, THE FEMORAL NAIL WAS FOUND TO HAVE RUPTURED."
Patient problems: Pain
MAUDE report 21256688 ↗InjuryJan 4, 2022Break; Fracture; Break; Fracture
AS REPORTED: "BROKEN GAMMA NAIL, 2 MONTHS POST-SURGERY. NEED A SECOND MEDICAL PROCEDURE. REVISION OF GAMMA NAIL + U-BLADE COLUMN SCREW." UPDATE: WHEN WALKING FROM TOILET TO THE KITCHEN THE PATIENT SUFFERED ACUTE SEVERE PAIN. SHE HAD TO SIT DOWN AND COULD NO LONGER LOAD THE LEG.
Patient problems: Failure of Implant; Implant Pain
MAUDE report 13156511 ↗InjurySep 3, 2021Break; Fracture; Break; Fracture
IT WAS REPORTED THAT 3 MONTHS AFTER IMPLANTATION OF A PROXIMAL FEMUR NAIL, THE MATERIAL BROKE. ADDITIONAL INFORMATION RECEIVED: "THE PATIENT PRESENTED TO OUR ORTHOPEDIC TRAUMA SURGERY DEPARTMENT ON (B)(6) 2021 WITH INCREASING SYMPTOMS AND STATUS AFTER IMPLANTATION OF THE ABOVE PRODUCT. BY MEANS OF CT ((B)(6) 2021) A FRACTURE OF THE NAIL AT THE LEVEL OF THE THIGH SCREW PENETRATION COULD BE DETECTED. ACCORDING TO THE PATIENT, SHE ALREADY NOTICED A CRACKING DURING THE REHABILITATION STAY WITH AN INCREASE IN PAIN. IN THE COURSE OF THE SUBSEQUENT X-RAY CONTROL, NO COMPLICATIONS COULD BE DETECTED. CONSEQUENCES: FAILURE OF THE BONE FRACTURE TO HEAL SUBTROCHANTERICALLY, REVISION WITH COMPLETE REMOVAL OF THE MARROW AND RE-IMPLANTATION OF A LONG INTRAMEDULLARY NAIL OSTEOSYNTHESIS ON (B)(6) 2021. ANOTHER REHABILITATION MEASURES."
Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture
MAUDE report 12421434 ↗InjuryApr 22, 2020Fracture; Fracture
IT WAS REPORTED, AFTER X-RAYS, IT WAS DISCOVERED, 4 MONTHS AFTER THE IMPLANTATION, THE BREAKAGE OF THE NAIL, AT LAG SCREW INTERSECTION AND A TOTAL HIP REPLACEMENT HAS BEEN PERFORMED.
Patient problems: Failure of Implant; Injury
MAUDE report 9988705 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page