Skip to content

GAMMA 3525-1320SRecalls & reported problems

Stryker GmbH · Long Nail Kit R1.5, Ti, LEFT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613252274261), received between Apr 22, 2020 and Sep 17, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Break4
  • Device Damaged by Another Device2

Patient problems

  • Failure of Implant3
  • Implant Pain2
  • Injury1
  • Insufficient Information1
  • Nonunion/Delayed-union Bone Fracture1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2025Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device

    AS REPORTED: "THE GAMMA3 NAIL BROKE."

    Patient problems: Insufficient Information

    MAUDE report 23076717 ↗
  • InjuryJan 28, 2025Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device

    THE CUSTOMER REPORTED THAT: "THE PATIENT UNDERWENT SURGERY ON (B)(6) 2024 FOR A GAMMA NAIL OF THE LEFT FEMUR (STRYKER). SINCE THE AFTERNOON OF (B)(6) 2024, SHE HAD BEEN EXPERIENCING PROGRESSIVE PAIN IN HER LEFT HIP. ON THE NIGHT OF (B)(6) 2024, DURING A TRANSFER TO THE TOILET, THE PATIENT DESCRIBED A ROTATING MOVEMENT OF THE HIP WITH AN AUDIBLE AN AUDIBLE CRACK. THE PAIN MADE IT IMPOSSIBLE TO STAND ON THE FLOOR. DURING SURGERY ON (B)(6) 2024, THE FEMORAL NAIL WAS FOUND TO HAVE RUPTURED."

    Patient problems: Pain

    MAUDE report 21256688 ↗
  • InjuryJan 4, 2022Break; Fracture; Break; Fracture

    AS REPORTED: "BROKEN GAMMA NAIL, 2 MONTHS POST-SURGERY. NEED A SECOND MEDICAL PROCEDURE. REVISION OF GAMMA NAIL + U-BLADE COLUMN SCREW." UPDATE: WHEN WALKING FROM TOILET TO THE KITCHEN THE PATIENT SUFFERED ACUTE SEVERE PAIN. SHE HAD TO SIT DOWN AND COULD NO LONGER LOAD THE LEG.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 13156511 ↗
  • InjurySep 3, 2021Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT 3 MONTHS AFTER IMPLANTATION OF A PROXIMAL FEMUR NAIL, THE MATERIAL BROKE. ADDITIONAL INFORMATION RECEIVED: "THE PATIENT PRESENTED TO OUR ORTHOPEDIC TRAUMA SURGERY DEPARTMENT ON (B)(6) 2021 WITH INCREASING SYMPTOMS AND STATUS AFTER IMPLANTATION OF THE ABOVE PRODUCT. BY MEANS OF CT ((B)(6) 2021) A FRACTURE OF THE NAIL AT THE LEVEL OF THE THIGH SCREW PENETRATION COULD BE DETECTED. ACCORDING TO THE PATIENT, SHE ALREADY NOTICED A CRACKING DURING THE REHABILITATION STAY WITH AN INCREASE IN PAIN. IN THE COURSE OF THE SUBSEQUENT X-RAY CONTROL, NO COMPLICATIONS COULD BE DETECTED. CONSEQUENCES: FAILURE OF THE BONE FRACTURE TO HEAL SUBTROCHANTERICALLY, REVISION WITH COMPLETE REMOVAL OF THE MARROW AND RE-IMPLANTATION OF A LONG INTRAMEDULLARY NAIL OSTEOSYNTHESIS ON (B)(6) 2021. ANOTHER REHABILITATION MEASURES."

    Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture

    MAUDE report 12421434 ↗
  • InjuryApr 22, 2020Fracture; Fracture

    IT WAS REPORTED, AFTER X-RAYS, IT WAS DISCOVERED, 4 MONTHS AFTER THE IMPLANTATION, THE BREAKAGE OF THE NAIL, AT LAG SCREW INTERSECTION AND A TOTAL HIP REPLACEMENT HAS BEEN PERFORMED.

    Patient problems: Failure of Implant; Injury

    MAUDE report 9988705 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates