Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613252274278), received between Feb 18, 2020 and Jan 7, 2026 — classified as 8 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture8
  • Break1
  • Device Damaged by Another Device1
  • Positioning Failure1

Patient problems

  • Failure of Implant7
  • Nonunion/Delayed-union Bone Fracture4
  • Implant Pain3
  • Injury3
  • Pain2
  • Fall1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2026Positioning Failure

    AS REPORTED: "THE SET SCREWS FROM THE GAMMA NAIL KIT DID NOT HOLD IN THE NAIL. AFTERWARDS, A SEPARATELY PACKAGED SET SCREW WAS OPENED, BUT THIS ONE ALSO DID NOT HOLD IN THE NAIL. SUBSEQUENTLY, A NEW NAIL WAS IMPLANTED, AND THE SET SCREW THAT CAME WITH THIS NAIL WORKED AS INTENDED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23989677 ↗
  • InjuryDec 22, 2025Fracture; Fracture

    AS REPORTED: "ON (B)(6) 2025, EMERGENCY ADMISSION DUE TO PAIN IN LEFT HIP; MULTIPLE FRAGMENT FEMUR FRACTURE AND OSTEOSYNTHESIS WITH GAMMA NAIL OUTWARD ON (B)(6) 2025. DIAGNOSIS: PSEUDARTHROSIS WITH IMPLANT FRACTURE IN LEFT HIP."

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 23871852 ↗
  • InjuryNov 27, 2024Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device

    AS REPORTED: "NAIL BREAKAGE THROUGH THE HOLE FOR THE FEMORAL NECK SCREW. NO FURTHER ACCIDENT. THE PATIENT HAD TO BE OPERATED ON AGAIN AND FITTED WITH A HIP TP."

    Patient problems: Failure of Implant

    MAUDE report 20797361 ↗
  • InjuryJul 22, 2021Fracture; Fracture

    IT WAS REPORTED THAT A GAMMA NAIL BROKE AND A REVISION SURGERY (TEP) WAS REQUIRED. UPDATE: WHILE REVIEWING THE X-RAYS, WE CAN NOTICE THAT, THERE IS BREAKAGE OF DISTAL LOCKING SCREW AS WELL.

    Patient problems: Failure of Implant

    MAUDE report 12212982 ↗
  • InjuryMar 5, 2021Fracture; Fracture

    THE CUSTOMER REPORTED THAT THE 13MM GAMMA NAIL WHICH WAS IMPLANTED APPROXIMATELY 2 MONTHS AGO WAS REVISED ON (B)(6) 2021. ADDITIONAL INFORMATION: SUDDEN ONSET OF PAIN, XRAYS SHOWED THE NAIL TO BE BROKEN THROUGH THE LAG-SCREW HOLE.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 11426256 ↗
  • InjuryDec 8, 2020Fracture; Fracture

    A BROKEN NAIL IS REPORTED. THE PRIMARY SURGERY TOOK PLACE ON (B)(6) 2020 AT THE HOSPITAL IN (B)(6). THE CUSTOMER REPORTS THAT THE BREAK MUST HAVE OCCURRED BETWEEN 1ST AUGUST AND 1ST OCTOBER. THE PATIENT CAME IN WITH PAIN.

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 10972461 ↗
  • InjuryAug 19, 2020Fracture; Fracture

    THE NAIL BROKE. REVISION HAD TO BE PERFORMED. ADDITIONAL INFORMATION FROM THE SURGEON: "AFTER 6 WEEKS OF RELIEF AFTER IMPLANTATION OF THE NAIL, MOBILIZATION WAS STARTED. AFTER INITIALLY GOOD MOBILIZATION, INCREASING PAIN OCCURRED IN THE OPERATED HIP UP TO IMMOBILIZATION. THE FRACTURE OF THE NAIL WAS THEN IDENTIFIED RADIOLOGICALLY. THERE WAS A CHANGE TO A CEMENTED LONG-STEM PROSTHESIS."

    Patient problems: Failure of Implant; Implant Pain; Injury; Nonunion/Delayed-union Bone Fracture

    MAUDE report 10427806 ↗
  • InjuryApr 6, 2020Fracture; Fracture

    AS REPORTED, GAMMA 3 NAIL BROKE THROUGH THE LAG SCREW HOLE ON THE (B)(6) 2020. THE NAIL WAS REMOVED ON THE (B)(6) 2020 AND A SYNTHES BLADE PLATE INSERTED. SURGEON THOUGHT THE TIME FROM IMPLANTATION UNTIL FAILURE FAIRLY SHORT. HE SAID THE ORIGINAL FRACTURE WAS A VERY CHALLENGING TYPE BUT IT HAD BEEN REDUCED WELL AND THE NAIL APPEARED TO BE IMPLANTED ALL IN ACCORDANCE WITH THE SURGICAL TECHNIQUE. THE ORIGINAL OPERATION WAS PERFORMED AT (B)(6) HOSPITAL.

    Patient problems: Failure of Implant; Injury; Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 9928477 ↗
  • InjuryFeb 18, 2020Fracture; Fracture

    IT WAS REPORTED, A PATIENT FELL OVER & FRACTURED LEFT NECK OF FEMUR. ALREADY HAD A LONG GAMMA NAIL IN HIP WHICH FRACTURED AT SITE OF LAG SCREW. OPERATION FOR REMOVAL OF BROKEN NAIL & INSERTION OF 13 MM LONG GAMMA NAIL.

    Patient problems: Failure of Implant; Fall; Injury; Nonunion/Delayed-union Bone Fracture

    MAUDE report 9721683 ↗