Adverse event reports
13
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 07613252274315), received between Nov 14, 2019 and Sep 14, 2026 — classified as 13 injuries.

The most frequently reported specific device problem is “Fracture” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture12
  • Break5
  • Device Damaged by Another Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Failure of Implant7
  • Implant Pain5
  • Insufficient Information3
  • Pain3
  • Ambulation or Postural Difficulties2
  • Impaired Healing1
  • Injury1
  • Limb Fracture1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Fracture

    A broken Gamma3 nail as reported. The patient required revision surgery.

    Patient problems: Insufficient Information

    MAUDE report 50756193 ↗
  • InjuryDec 15, 2025Fracture; Device Damaged by Another Device; Fracture; Device Damaged by Another Device

    IT WAS REPORTED THAT A GAMMA 3 LONG NAIL BROKE. THIS LEAD TO A REVISION SURGERY TO IMPLANT A PLATE.

    Patient problems: Insufficient Information

    MAUDE report 23805310 ↗
  • InjuryJul 21, 2025Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device

    IT WAS REPORTED THAT A GAMMA3 NAIL, THAT WAS IMPLANTED 3 MONTHS AGO, FRACTURED AT THE LEVEL OF THE FEMORAL NECK SCREW RESULTING IN A REVISION SURGERY.

    Patient problems: Impaired Healing; Insufficient Information

    MAUDE report 22559881 ↗
  • InjuryMay 22, 2025Break; Fracture; Break; Fracture

    AS REPORTED: "THE PATIENT WAS IMPLANTED WITH THE LONG GAMMA3 NAIL DESCRIBED ABOVE ON (B)(6) 2024 FOR A SUBTROCHANTERIC FRACTURE ON THE LEFT. HE WAS ABLE TO BE MOBILIZED UNDER FULL WEIGHT-BEARING DURING THE PROCEDURE. A FRACTURE OF THE LONG GAMMA NAIL WAS DIAGNOSED DURING A FOLLOW-UP X-RAY EXAMINATION, WHICH WAS PERFORMED DUE TO PAIN. ON (B)(6) 2025, REVISION SURGERY WAS PERFORMED WITH EXPLANTATION OF THE BROKEN GAMMA 3 NAIL ON THE LEFT."

    Patient problems: Pain

    MAUDE report 22072283 ↗
  • InjuryAug 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PRESENT SWISS TRAUMA FRACTURE REGISTRY (SWTR) IS A PROSPECTIVE, MULTI-CENTER, OBSERVATIONAL STUDY TO COLLECT DATA ON ALL STRYKER TRAUMA AND EXTREMITIES PRODUCTS WITH THE AIM TO ADD TO THE EXISTING CLINICAL EVIDENCE TO SUPPORT THE SAFETY AND PERFORMANCE OF STRYKER IMPLANTS FOR THE APPROVED INDICATIONS FOR USE. THE SWTR CAPTURES DATA EVALUATING THE REAL-WORLD USAGE OF STRYKER IMPLANTS AT TRAUMA CENTERS WITH THE CAPACITY, TRAINING, AND EXPERIENCE TO GATHER CLINICALLY RELEVANT STANDARD OF CARE SAFETY AND PERFORMANCE ENDPOINT DATA. DURING THE REVIEW OF THE REPORT, PERSISTENT PAIN WAS NOTED FOR THIS PATIENT DURING THE FOLLOW-UP VISIT.

    Patient problems: Implant Pain

    MAUDE report 19954796 ↗
  • InjuryMar 26, 2024Break; Fracture; Break; Fracture

    THE CUSTOMER REPORTED: "I WOULD LIKE TO CONTACT YOU REGARDING THE ATTACHED REPORT ON A GAMMA NAIL. BREAKAGE OF THE NAIL AND DISTAL SCREW WHEN THE PATIENT MOVES (WALKING) NO NOTION OF TRAUMA. DEVICES INVOLVED: TWO. CLINICAL CONSEQUENCES OBSERVED: SECONDARY FRACTURE DISPLACEMENT FUNCTIONAL IMPOTENCE + PAIN PROTECTIVE MEASURES AND ACTIONS TAKEN: REMOVAL OF THE NAIL AND FITTING OF ANOTHER NAIL + GTR HOOK".

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Implant Pain; Limb Fracture

    MAUDE report 18981205 ↗
  • InjuryJul 25, 2023Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A GAMMA 3 LONG NAIL BROKE. NAIL WILL BE EXPLANTED.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Implant Pain; Pain

    MAUDE report 17393090 ↗
  • InjuryAug 19, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT GAMMA NAIL AND SCREW BROKE. IMPLANTATION TOOK PLACE IN (B)(6) 2022. PATIENT STATED TO THE PHYSICIAN THAT HE HAD FOLLOWED THE PHYSICIAN'S INSTRUCTIONS, BUT HAD NOTICED PAIN WHILST WALKING. EXPLANTATION AND SIMULTANEOUS IMPLANTATION OF A NEW GAMMA3 NAIL.

    Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture

    MAUDE report 15260081 ↗
  • InjuryNov 6, 2021Fracture; Fracture

    THE CUSTOMER REPORTED THAT THE GAMMA NAIL WHICH WAS IMPLANTED IN (B)(6) 2021 HAS BROKEN POST IMPLANTATION, REQUIRING REVISION SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12755552 ↗
  • InjuryMay 27, 2021Fracture; Fracture

    THE CUSTOMER REPORTED: "79-YEAR-OLD FEMALE PATIENT OPERATED ON IN (B)(6) 2019 FOR A SUBTROCHANTERIC FRACTURE OF THE LEFT FEMUR REDUCED AND OSTEOSYNTHESIZED WITH GAMMA LONG NAIL. ON (B)(6) 2020, SHE CONSULTED THE EMERGENCY ROOM FOR PAIN IN THE LEFT THIGH THAT HAD BEEN EVOLVING FOR 1 WEEK. THE RADIOGRAPHIC WORKUP REVEALED FAILURE OF THE OSTEOSYNTHESIS MATERIAL OF THE LONG GAMMA NAIL REQUIRING A SURGICAL REVISION. ON THE RADIOGRAPHIC ASSESSMENT PERFORMED ON (B)(6) 2021 THE MATERIAL IS WELL IN PLACE AND NOT BROKEN. THERE IS A CLEAR IMPROVEMENT OF THE FOCUS OF PSEUDOARTHROSIS. INTERNALLY, WE HAVE THE IMPRESSION TO HAVE A GOOD BONE BASE AND IN PROFILE IT LOOKS AS IF ALL THE IMPRESSION THAT ALL THE CORTICES ARE SOLID. EXPLANTATION OF THE DEFECTIVE DM AND REIMPLANTATION OF A LONG GAMMA NAIL."

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 11898820 ↗
  • InjuryMay 26, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE LEFT LONG GAMMA NAIL BROKE. IT WAS FURTHER REPORTED THAT AS A RESULT THE PATIENT HAD TO UNDERGO AN ADDITIONAL SURGERY TO HAVE A NEW FEMORAL NAIL INSERTED.

    Patient problems: Failure of Implant

    MAUDE report 11884886 ↗
  • InjuryNov 13, 2020Fracture; Fracture

    THE CUSTOMER REPORTED THAT THE GAMMA NAIL HAD BROKEN WHERE THE NECK OF THE NAIL INTERFACES WITH THE LAG SCREW. THE NAIL WAS IMPLANTED A FEW MONTHS AGO. THE BROKEN NAIL AND IMPLANTS WERE REMOVED AND REVISED WITH 13MM GAMM.

    Patient problems: Failure of Implant

    MAUDE report 10840210 ↗
  • InjuryNov 14, 2019Fracture; Fracture; Appropriate Device Problem Term/Code Not Available

    ON (B)(6) 2019 ,BIKE ACCIDENT. ON (B)(6) 2019, GAMMA3 NAIL IMPLANTED. ON (B)(6) 2019, REVISION SURGERY. PATIENT VISITED HOSPITAL AFTER PAIN.

    Patient problems: Injury; Pain

    MAUDE report 9324326 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →