NA 4405000000Recalls & reported problems
Stryker Corporation · Cordless Driver 4 Handpiece · Class 1
Quick summary
This Stryker Cordless Driver 4 handpiece is a cordless surgical tool used in orthopedic surgery. Its 105 reports are all malfunction reports, often describing the handpiece continuing to run after it was switched off, along with fluid leaks, parts coming loose and material breaking down. There were no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 105
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 105 adverse event reports that list this exact device (DI 07613252481355), received between Jul 24, 2018 and Oct 23, 2025 — classified as 105 malfunctions.
The most frequently reported specific device problem is “Device Remains Activated” (29 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Remains Activated29
- Fluid/Blood Leak24
- Detachment of Device or Device Component16
- Material Disintegration11
- Fail-Safe Problem10
- Physical Resistance/Sticking6
- Material Fragmentation5
- Unintended Power Up4
Patient problems
- No Clinical Signs, Symptoms or Conditions62
- No Patient Involvement33
- No Known Impact Or Consequence To Patient14
- No Consequences Or Impact To Patient7
- Insufficient Information4
Recent adverse event reports
The 25 most recent of 105 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 23, 2025Material Disintegration; Material Disintegration
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: MATERIAL DISINTEGRATION. - 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368686 ↗MalfunctionJan 21, 2025Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21200921 ↗MalfunctionOct 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501576 ↗MalfunctionOct 21, 2024Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20500816 ↗MalfunctionOct 21, 2024Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED REGARDING HEALTH IMPACT.
Patient problems: Insufficient Information
MAUDE report 20500642 ↗MalfunctionJul 24, 2024Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON.3 EVENTS HAD NO PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19815964 ↗MalfunctionApr 15, 2024Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113367 ↗MalfunctionJan 22, 2024Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18556932 ↗MalfunctionOct 23, 2023Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17993164 ↗MalfunctionJul 26, 2023Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401606 ↗MalfunctionJul 26, 2023Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17398448 ↗MalfunctionApr 25, 2023Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. 8 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807821 ↗MalfunctionApr 25, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807579 ↗MalfunctionApr 25, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807578 ↗MalfunctionJan 29, 2023Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. 9 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16262050 ↗MalfunctionJan 29, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261939 ↗MalfunctionJan 29, 2023Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261864 ↗MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261784 ↗MalfunctionOct 25, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15666873 ↗MalfunctionOct 25, 2022Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 7 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15666749 ↗MalfunctionOct 25, 2022Material Fragmentation; Material Fragmentation
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD SMALL PIECES BREAKING OFF. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15666447 ↗MalfunctionOct 24, 2022Unintended Power Up; Unintended Power Up
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663784 ↗MalfunctionOct 24, 2022Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 6 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663744 ↗MalfunctionJul 26, 2022Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 16 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 16 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102365 ↗MalfunctionJul 26, 2022Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. - 4 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102173 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.