TARGET M0035443040Recalls & reported problems

Stryker Corporation · 360 Nano Detachable Coil · Class 2 · Implant

All models of TARGET (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613252600336), received between Apr 24, 2017 and Jun 30, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation3
  • Fracture2
  • Break1
  • Expulsion1
  • Migration or Expulsion of Device1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 30, 2026Fracture; Expulsion

    IT WAS REPORTED THAT DURING THE PROCEDURE, AN ATTEMPT WAS MADE TO DETACH THE SUBJECT COIL, BUT THE SUBJECT COIL FAILED TO DETACH. FORCEFUL TRACTION APPLIED TO SUBJECT COIL DELIVERY WIRE, RESULTED IN BREAKAGE AT THE DISTAL SEGMENT OF THE DELIVERY WIRE. THE DETACHED DISTAL SEGMENT COULD NOT BE RETRIEVED. FOLLOW-UP ANGIOGRAPHY REVEALED SUBJECT COIL PROTRUSION, AND A RESCUE STENT WAS PLACED TO MANAGE THIS COMPLICATION. THE PROCEDURE WAS THEN COMPLETED. THERE WAS A SURGICAL DELAY OF UNKNOWN MINUTES DUE TO THE REPORTED EVENT. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25669783 ↗
  • MalfunctionApr 25, 2023Premature Separation; Premature Separation

    THE SUBJECT COIL WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THE SUBJECT COIL PREMATURELY DETACHED/SEPARATED DURING USE. THERE WERE NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16805218 ↗
  • MalfunctionMay 12, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING COIL EMBOLIZATION FOR URGENT GASTROINTESTINAL BLEEDING. AFTER GUIDING THE MICROCATHETER TO THE TARGET SITE, THE COIL (SUBJECT DEVICE) WAS SELECTED AS THE FIRST COIL. AFTER PLACING IT, A SHORT SOUND WAS HEARD THREE TIMES. THE DETACH BUTTON WAS PRESSED AGAIN AND A LONG BEEP SOUNDED, THUS IT WAS DETERMINED THAT THE COIL (SUBJECT DEVICE) WAS DETACHED, AND THE DELIVERY WIRE WAS REMOVED. THE MICROCATHETER GOT DEVIATED WHEN THE COIL (SUBJECT DEVICE) CAME OUT AND DIDN'T DETACH. THE COIL (SUBJECT DEVICE) WAS RETRACTED. THE MICROCATHETER WAS INSERTED AGAIN AND THIS COIL (SUBJECT DEVICE) WAS REUSED, NEVERTHELESS THIS COIL WAS PREMATURELY DETACHED WHILE ADVANCEMENT IN THE MICROCATHETER SHAFT OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14379942 ↗
  • MalfunctionMar 29, 2021Premature Separation; Premature Separation

    IT WAS REPORTED THAT SUBJECT COIL COMES OFF ITSELF IN THE CATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11586725 ↗
  • MalfunctionMar 2, 2021Fracture; Fracture

    ANALYSIS OF THE RETURNED SUBJECT COIL REVEALED THAT THE COIL DELIVERY WIRE HAD BROKEN/FRACTURED DURING USE. THEREFORE, BASED ON THIS INFORMATION THE EVENT IS DEEMED REPORTABLE WITH AN AWARENESS DATE OF (B)(6) 2021. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11400066 ↗
  • MalfunctionDec 27, 2018Break; Break

    DURING THE COIL EMBOLIZATION PROCEDURE, IT WAS REPORTED THAT THE COIL GOT STUCK IN THE MICROCATHETER AND BROKE. NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE NOW.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8200439 ↗
  • InjuryApr 24, 2017Migration or Expulsion of Device; Premature Activation; Migration or Expulsion of Device; Premature Activation

    IT WAS REPORTED THAT AS THE COIL WAS BEING ADVANCED, IT PREMATURELY DETACHED AND MIGRATED. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE COIL BUT WAS UNSUCCESSFUL. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6517213 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices