The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613252600367), received between Jan 9, 2018 and Sep 13, 2019 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Premature Activation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Premature Activation1
  • Premature Separation1
  • Unintended Movement1

Patient problems

  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2019Premature Separation; Premature Separation

    DURING THE ANALYSIS OF THE RETURNED DEVICE, IT WAS REVEALED THAT THE MAIN COIL WAS PREMATURELY DETACHED INSIDE THE PATIENT. NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9042026 ↗
  • MalfunctionOct 2, 2018Unintended Movement; Unintended Movement

    A (B)(6) FEMALE (HYPERTENSION, PREVIOUS SMOKER) UNDERWENT STENT ASSISTED COIL EMBOLIZATION OF ANEURYSM IN RIGHT MIDDLE CEREBRAL ARTERY (MCA) (B)(6) 2017. WHEN THE PATIENT WAS SEEN FOR FOLLOW-UP APPROXIMATELY ONE YEAR LATER (B)(6) 2018, IT WAS REPORTED THE COIL MIGRATED THROUGH THE STENT STRUTS AND WAS PROTRUDING SLIGHTLY INTO THE ARTERY, AS OBSERVED ON DSA (DIGITAL SUBTRACTION ANGIOGRAPHY) IMAGING. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT AS A RESULT; THE EVENT WAS NOTED TO RESOLVE WITHOUT SEQUELAE THE SAME DAY ((B)(6) 2018). ADDITIONAL INFORMATION FROM THE PHYSICIAN INDICATED THE STENT WAS USED SUCCESSFULLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7928878 ↗
  • MalfunctionJan 9, 2018Premature Activation; Premature Activation

    BASED ON THE ANALYSIS OF THE DEVICE, IT WAS FOUND THAT THE MAIN COIL WAS PHYSICALLY DETACHED FROM THE DELIVERY WIRE. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7177422 ↗