TARGET M0035453080Recalls & reported problems

Stryker Corporation · Helical Nano Detachable Coil · Class 2 · Implant

All models of TARGET (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613252600442), received between Dec 8, 2016 and Jan 13, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation2
  • Fracture1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2026Premature Separation

    ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE MAIN COIL WAS PREMATURELY DETACHED/SEPARATED DURING USE. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24064081 ↗
  • MalfunctionMay 3, 2023Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING AN ENDOVASCULAR PROCEDURE WHEN DELIVERING THE SUBJECT COIL TO ITS DESTINATION, THERE WAS RESISTANCE INSIDE THE MICROCATHETER. THE PHYSICIAN STARTED TO WITHDRAW THE SUBJECT COIL AND FELT THAT THERE WAS NO ANY RESISTANCE. THE SUBJECT COIL DETACHED FROM THE WIRE AND WAS PARTLY INSIDE THE MICROCATHETER. WITH A COIL PUSHER THE PHYSICIAN WAS ABLE TO GET THE SUBJECT COIL TO THE CORRECT DESTINATION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16861562 ↗
  • MalfunctionAug 4, 2021Fracture; Fracture

    IT WAS INITIALLY REPORTED THAT DURING THE EMBOLISM OF PCOMA ANEURYSM, A GUIDE WIRE AND MICROCATHETER WERE USED TO DELIVERED THE A COIL (SUBJECT DEVICE) TO TARGET SITE. HOWEVER, THE SUBJECT COIL EXPERIENCED FRICTION AT THE PROXIMAL OF THE MICROCATHETER AND COULD NOT BE ADVANCED OUT OF THE MICROCATHETER. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT. ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE COIL DELIVERY WIRE (SUBJECT DEVICE) WAS FOUND BROKEN/FRACTURE DURING USE. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12278018 ↗
  • MalfunctionDec 8, 2016Premature Activation; Premature Activation

    BASED ON INVESTIGATION OF THE RETURNED PRODUCT, THE MAIN COIL DETACHED PREMATURELY INSIDE THE PATIENT. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6156461 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →