FlowPort CAT00776Recalls & reported problems
Stryker Corporation · FLOWPORT II CANNULA WITH OBTUR · Class 1
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613252632887), received between Apr 1, 2019 and Jan 27, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Break2
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient2
- Insufficient Information2
- No Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 27, 2026Fracture
IT WAS REPORTED THAT THE TIP OF FLOWPORT BROKE OFF IN JOINT OF MALE PATIENT. IT WAS RETRIEVED BEFORE CLOSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24195857 ↗MalfunctionSep 29, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PLASTIC TIP OF CANNULA BROKE INTO THE PATIENT. THIS WAS RETRIEVED BY THE SURGEON HOWEVER THIS PROCEDURE WAS PROLONGED BY 15 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23175671 ↗InjuryJan 30, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING THE PROCESS OF ENTERING THE HIP JOINT THE STRYKER BLUNT PLASTIC CANNULA WAS PINCHED BETWEEN THE LATERAL RIM OF THE ACETABULUM END FEMORAL HEAD AND APPROXIMATELY 1.6CM PLASTIC PIECE BROKE OFF. X-RAYS REVEALED THAT IT WAS PINCHED BETWEEN THE LATERAL ACETABULUM AND FEMORAL HEAD. THE FINAL X RAYS REVEALED A WELL REDUCED HIP JOINT WITH RETAINED PLASTIC PIECE UNCHANGED FROM INTRAOPERATIVE X-RAYS.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 21277214 ↗InjuryNov 11, 2024Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF THE CANNULA BROKE OFF IN THE PATIENT JOINT SPACE. IT WAS ABLE TO BE LOCATED BUT COULD NOT BE REMOVED. A SECOND SURGERY WILL BE REQUIRED.
Patient problems: Foreign Body In Patient
MAUDE report 20655776 ↗MalfunctionApr 3, 2024Fracture; Fracture
IT WAS REPORTED THAT THE TIP BREAKING DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19038634 ↗MalfunctionFeb 17, 2022Break
IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13549292 ↗MalfunctionFeb 10, 2022Break
IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 13507540 ↗InjuryApr 1, 2019Insufficient Device Problem Information
IT WAS REPORTED THAT THE PROCEDURE WAS CONVERTED TO AN OPEN PROCEDURE.
Patient problems: No Information
MAUDE report 8471349 ↗
Other ACCESSORIES,ARTHROSCOPIC devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.