Skip to content

FlowPort CAT02438Recalls & reported problems

Stryker Corporation · FLOWPORT II CANNULA WITH OBTURATOR STRYKER 165MM · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker FlowPort II cannula is a tube used in arthroscopic hip surgery. Most of the 46 reports describe the cannula or its obturator tip breaking or cracking during a procedure, in some cases leaving a fragment inside the patient's hip that required revision surgery. There were 8 injuries and no deaths. No recalls name this device. Reports in the last 12 months (2) are far fewer than the 11 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 07613252633181), received between Jul 2, 2018 and Apr 9, 2026 — classified as 8 injuries, 38 malfunctions.

The most frequently reported specific device problem is “Break” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break26
  • Fracture19
  • Crack2
  • Material Fragmentation2
  • Contamination /Decontamination Problem1
  • Device Contaminated During Manufacture or Shipping1
  • Material Integrity Problem1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • No Known Impact Or Consequence To Patient16
  • Foreign Body In Patient3
  • No Consequences Or Impact To Patient3
  • Insufficient Information2
  • Device Embedded In Tissue or Plaque1
  • Pain1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2026Fracture

    IT WAS REPORTED THAT WHILE INSERTING THE CANNULA INTO THE PATIENT'S HIP JOINT CAPSULE, THE CANNULA BROKE OFF INTO THE PATIENT. THE CANNULA WAS ABLE TO BE RETRIEVED BY THE SURGEON, AND IT WAS CONFIRMED NOTHING WAS STILL RETAINED IN THE PATIENT BY X-RAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24837603 ↗
  • MalfunctionMar 18, 2026Fracture

    IT WAS REPORTED THAT A PATIENT HAD SURGERY ON (B)(6) 2026 USING STRYKER PRODUCTS. ON (B)(6) 2026, THE DOCTOR REMOVED A DISPOSABLE PIECE OF FLOWPORT CANNULA FROM THE SAME PATIENT THAT HAD BROKEN OFF AND WAS LEFT LODGED IN THE PATIENT¿S HIP. REVISION SURGERY WAS REQUIRED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24631496 ↗
  • MalfunctionMay 23, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE PRODUCT BROKE OFF INTO THE HIP JOINT. THE BROKEN PIECE WAS NOT LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22080461 ↗
  • InjuryMay 23, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE PRODUCT BROKE OFF INTO THE HIP JOINT. ALL BUT ONE SMALL LOOSE BODY WAS REMOVED. THIS RETAINED PIECE WAS OUTSIDE THE JOINT.

    Patient problems: Foreign Body In Patient

    MAUDE report 22080111 ↗
  • MalfunctionMay 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE OBTURATOR BROKE OFF IN THE PATIENT. THE SURGEON WAS ABLE TO REMOVE IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22074584 ↗
  • MalfunctionMay 8, 2025Fracture; Fracture

    IT WAS REPORTED THAT PART OF THE CANNULA BROKE OFF DURING SURGERY. THE BROKEN PIECE WAS UNABLE TO BE LOCATED. HOWEVER, X-RAYS WERE NEGATIVE FOR RETAINED FOREIGN BODY. NO HARM TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21990174 ↗
  • MalfunctionMar 27, 2025Fracture; Product Quality Problem; Device Contaminated During Manufacture or Shipping; Fracture; Product Quality Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THERE WAS FOREIGN MATERIAL INSIDE THE STERILE PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21712769 ↗
  • MalfunctionMar 5, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21531973 ↗
  • MalfunctionMar 5, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21531878 ↗
  • InjuryFeb 26, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE FLOWPORT 165 BROKE OFF INTO JOINT DURING HIP SCOPE. A LARGER INCISION WAS NEEDED TO PULL THE BROKEN TIP OUT BUT OTHERWISE NO COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21472335 ↗
  • MalfunctionNov 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20799489 ↗
  • InjuryNov 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF FLOWPORT 165 BROKE OFF INSIDE HIP JOINT. WE HAD TO DO A MINI OPEN TO GET IT OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694915 ↗
  • MalfunctionNov 6, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20616172 ↗
  • MalfunctionSep 25, 2024Fracture; Fracture

    IT WAS REPORTED THAT FOR THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20306615 ↗
  • InjuryAug 29, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE SURGEON INTRODUCED THE FLOWPORT CANNULA THROUGH HIS INCISION AND INTO THE JOINT SPACE. AS THE SAMUARI BLADE WAS INTRODUCE INTO THE CANNULA, IT WAS RETRACTED BACK IN ORDER TO CUT THE CAPSULE AND THE SURGEON NOTICED THE DISTAL LEADING PLASTIC EDGE OF THE CANNULA HAD BROKEN OFF THE METAL PORTION OF THE CANNULA. THE ARTHROSCOPY WAS CANCELLED AND THE SURGEON HAD TO PERFORM AN ARTHROTOMY TO RETRIEVE THE BROKEN PLASTIC PIECE. THE POSTED PROCEDURE HAD TO BE STOPPED AND HAS TO BE CONDUCTED AT A LATER TIME DUE TO THE ARTHROTOMY OF THE JOINT CAPSULE. THE PROCEDURE TURNED FROM AN ARTHROSCOPY TO OPEN JOINT ARTHROTOMY TO SUCCESSFULLY RETRIEVE THE BROKEN PIECE.

    Patient problems: Insufficient Information

    MAUDE report 20101297 ↗
  • MalfunctionApr 10, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084361 ↗
  • MalfunctionFeb 21, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18753929 ↗
  • MalfunctionJan 5, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18450028 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18448367 ↗
  • InjurySep 8, 2023Break; Fracture; Break; Fracture

    IT WAS REPORTED DISTAL TIP OF THE CANNULA BROKE OFF, AND WAS LEFT IN THE PATIENT AFTER SURGERY, WITHOUT HIS KNOWLEDGE. THE PATIENT RETURNED TO THE SURGEON WITH PERSISTENT PAIN AND AN XRAY REVEALED THE BROKEN DISTAL TIP IN SITU. THE PATIENT HAD TO HAVE THE DISTAL TIP SURGICALLY REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 17715555 ↗
  • MalfunctionMay 22, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE WAS FOREIGN MATERIAL IN THE STERILE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16980231 ↗
  • InjuryApr 21, 2023Break; Break

    IT WAS REPORTED THERE THERE WAS A CONVERSION TO OPEN SURGERY DUE TO THE TIP BREAKING DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16792145 ↗
  • MalfunctionSep 7, 2022Break; Break

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE BROKEN TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15375337 ↗
  • MalfunctionApr 26, 2022Break

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14215027 ↗
  • MalfunctionMar 24, 2022Break; Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE DEVICE BROKE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13891767 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page