Skip to content

ES2 SPINAL SYSTEM 482802645Recalls & reported problems

Stryker Corporation · SHORT BLADE CANNULATED SCREW · Class 2 · Implant

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327001945), received between Aug 22, 2016 and Nov 21, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Fracture6
  • Material Integrity Problem4
  • Detachment Of Device Component1
  • Device Dislodged or Dislocated1
  • Device Reprocessing Problem1
  • Insufficient Device Problem Information1
  • Material Separation1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • Injury3
  • Failure of Implant2
  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • Pain1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2025Device Reprocessing Problem; Insufficient Device Problem Information; Device Reprocessing Problem; Insufficient Device Problem Information

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23623162 ↗
  • InjuryMar 26, 2025Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration

    IT WAS REPORTED THAT A PATIENT WITH PAIN WAS REVISED TO ADDRESS A MIGRATED ES2 INTEGRATED BLADE SCREW.

    Patient problems: Pain

    MAUDE report 21700643 ↗
  • MalfunctionApr 25, 2017Break; Fracture; Fracture

    IT WAS REPORTED THAT; THE EXTRACTION SURGERY WAS PERFORMED DUE TO HEALING OF THE DISEASE. DURING THE EXTRACTION SURGERY, THE BREAKAGE OF L5 RIGHT SIDE SCREW WAS FOUND. THE TIP OF THE SCREW WAS REMAINED TO THE PATIENT AND THE SURGERY WAS FINISHED.

    Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient

    MAUDE report 6518666 ↗
  • MalfunctionApr 18, 2017Break; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2015 IMPLANTED FROM L4 TO L5 BY PLIF USING SOME SCREWS. IMPLANTED 4 SCREW AND L4 LEFT, L5 RIGHT. ON (B)(6) 2016, IMPLANTED TO EXTEND BY PEEK TO S FROM L5 BY PLIF. IMPLANTED 2 SCREWS FOR SI AND FIXED FOR L4-S BY SCREW(S1RIGHT). UNK DATE, IT WAS CONFIRMED 3 SCREW WERE BROKEN(L4LEFT,L5RIGHT,S1RIGHT). ON (B)(6) 2017 REMOVED BROKEN 3 SCREW AND IMPLANTED NEW SCREWS.

    Patient problems: Failure of Implant; Injury

    MAUDE report 6501560 ↗
  • MalfunctionApr 18, 2017Break; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2015 IMPLANTED FROM L4 TO L5 BY PLIF USING SOME SCREWS. IMPLANTED 4 SCREW AND L4 LEFT, L5 RIGHT. ON (B)(6) 2016, IMPLANTED TO EXTEND BY PEEK TO S FROM L5 BY PLIF. IMPLANTED 2 SCREWS FOR SI AND FIXED FOR L4-S BY SCREW (S1RIGHT). UNK DATE, IT WAS CONFIRMED 3 SCREW WERE BROKEN (L4LEFT,L5RIGHT,S1RIGHT). ON (B)(6) 2017 REMOVED BROKEN 3 SCREW AND IMPLANTED NEW SCREWS.

    Patient problems: Failure of Implant; Injury

    MAUDE report 6501558 ↗
  • InjuryApr 18, 2017Break; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2015 IMPLANTED FROM L4 TO L5 BY PLIF USING SOME SCREWS. IMPLANTED 4 SCREW AND L4 LEFT, L5 RIGHT. ON (B)(6) 2016, IMPLANTED TO EXTEND BY PEEK TO S FROM L5 BY PLIF. IMPLANTED 2 SCREWS FOR SI AND FIXED FOR L4-S BY SCREW(S1RIGHT). UNK DATE, IT WAS CONFIRMED 3 SCREW WERE BROKEN(L4LEFT,L5RIGHT,S1RIGHT). ON (B)(6) 2017 REMOVED BROKEN 3 SCREW AND IMPLANTED NEW SCREWS.

    Patient problems: Injury; No Code Available

    MAUDE report 6501557 ↗
  • MalfunctionApr 5, 2017Detachment Of Device Component; Material Separation; Material Integrity Problem; Material Separation; Material Integrity Problem

    IT WAS REPORTED THAT; THIS EVENT OCCURRED AT THE FIXATION OF THE L4/5 AND THE ILIAC BONE OF THE PELVIC FRACTURE. WHEN THE FINAL TIGHTENING OF THE BLOCKER AT RIGHT L4 SCREW, THE SURGEON CONFIRMED THAT THE SCREW HEAD WAS CAME OFF FROM THE SCREW SHAFT. AFTER THAT, THE SURGEON PULLED OUT THE ROD AND CONFIRMED THAT THE SCREW OF THE L5 AND ILIAC BONE WHICH WAS ALREADY IMPLANTED. THE TWO SCREW HEADS WERE CAME OFF FROM THE SCREW SHAFT TOO. THEREFORE, THE THREE SCREWS WERE EXCHANGED THE NEW OTHER SIZE SCREWS AND THE OPERATION WAS SUCCEEDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6460585 ↗
  • MalfunctionFeb 7, 2017Break; Fracture; Fracture

    IT WAS REPORTED THAT DURING A PERCUTANEOUS ARTHRODESIS, INSERTION OF 4 SCREWS INTO THE PEDICLES OF THE PATIENT. THEN, TO PROCEED THE DISCECTOMY TO INSERT A ROTATIVE LUMBAR CAGE INTO THE INTRADISCAL SPACE, THE SURGEON INSTALLED THE PBR SPACER. HE INTRODUCED A SHORT PIN INTO EACH SUPPORTING SCREWS OF THE SPACER. TO INSERT THE PIN INTO THE THREAD OF THE SCREW, THE SURGEON USED A COUNTER TORQUE KEY TO HOLD THE SCREW. WHEN INSERTING THE PIN INTO THE THREAD, THE SHORT FINS OF THE SCREW BROKE. THEY FRACTURED AT THE LEVEL OF THE SPACE PROVIDED TO ALLOW BREAKING AT THE END OF THE INTERVENTION. THE SURGEON REPLACED THE SCREW BY A NEW ONE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6307068 ↗
  • MalfunctionAug 22, 2016Break; Fracture; Fracture

    IT WAS REPORTED THAT; PLIF (L4-5) WAS PERFORMED ON (B)(6) 2015. ADDITIONAL FIXATION (L5-S1) WAS PERFORMED ON (B)(6) 2016. 1 MONTH AFTER THE SECOND SURGERY, THE BREAKAGE OF S1 SCREW WAS FOUND. THE REVISION SURGERY IS NOT PLANNED AT PRESENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5894734 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.