ES2 SPINAL SYSTEM 482802650Recalls & reported problems
Stryker Corporation · SHORT BLADE CANNULATED SCREW · Class 2 · Implant
All models of ES2 SPINAL SYSTEM (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327001952), received between Sep 16, 2016 and Jun 5, 2023 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture4
- Break2
- Device Dislodged or Dislocated2
- Loose or Intermittent Connection2
- Mechanical Problem2
- Migration or Expulsion of Device2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Injury5
- No Code Available2
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2023Fracture; Fracture
IT WAS REPORTED THAT APPROXIMATELY 6 YEARS AFTER IMPLANTATION DURING A PLANNED REVISION SURGERY TO ADDRESS LUMBAR STENOSIS OF L4/5, IT WAS OBSERVED THAT AN ES2 SHORT BLADE CANNULATED SCREW AT RIGHT S1 HAD FRACTURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17065319 ↗InjuryAug 14, 2019Fracture; Appropriate Device Problem Term/Code Not Available; Fracture
POST-OPERATIVE X-RAY INDICATED THAT A PIECE OF THE REMOVABLE BLADE OF AN ES2 INTEGRATED BLADE SCREW REMAINED IN THE PATIENT. THE PATIENT WAS REVISED UNDER LOCAL ANESTHESIA ON (B)(6) 2019 TO EXTRACT THE PORTION OF THE REMOVABLE BLADE.
Patient problems: Injury
MAUDE report 8893562 ↗InjuryMay 19, 2017Loose or Intermittent Connection; Mechanical Problem; Migration or Expulsion of Device; Device Dislodged or Dislocated; Mechanical Problem; Migration or Expulsion of Device
IT WAS REPORTED THAT ON (B)(6) 2017 IMPLANTED THE PROCEDURE. (B)(6) 2017 LOOSED FOUR SCREWS AND DISLODGED THE GAGE. (B)(6) 2017, IT WAS REMOVED TO NEW SCREW (OTHER COMPANY) FROM FOUR OSING SCREWS.
Patient problems: Injury; No Code Available
MAUDE report 6578640 ↗MalfunctionMay 19, 2017Loose or Intermittent Connection; Mechanical Problem; Migration or Expulsion of Device; Device Dislodged or Dislocated; Mechanical Problem; Migration or Expulsion of Device
IT WAS REPORTED THAT ON (B)(6) 2017 IMPLANTED THE PROCEDURE. ON (B)(6) 2017 LOOSED FOUR SCREWS AND DISLODGED THE GAGE. ON (B)(6) 2017, IT WAS REMOVED TO NEW SCREW(OTHER COMPANY) FROM FOUR OSING SCREWS.
Patient problems: Injury; No Code Available
MAUDE report 6578635 ↗MalfunctionSep 19, 2016Break; Fracture; Fracture
IT WAS REPORTED THAT; ON (B)(6 )2016, ES2 SURGERY FOR L2/3 BURST FRACTURE WAS PERFORMED. T12-L5 WERE FIXED, L2 WERE SKIPPED. AFTER THE SURGERY, THE BREAKAGE OF BOTH SIDE L5 SCREW WAS FOUND. ALL THE IMPLANT WERE EXTRACTED ON (B)(6) 2016 BECAUSE THE FRACTURE HAS HEALED. THE TIPS OF THE BROKEN SCREWS WERE REMAINED TO THE PATIENT. UPDATE: PATIENT WAS FUSED.
Patient problems: Injury; No Known Impact Or Consequence To Patient
MAUDE report 5959444 ↗MalfunctionSep 16, 2016Break; Fracture; Fracture
IT WAS REPORTED THAT; ON (B)(6) 2016, ES2 SURGERY FOR L2/3 BURST FRACTURE WAS PERFORMED. T12-L5 WERE FIXED, L2 WERE SKIPPED. AFTER THE SURGERY, THE BREAKAGE OF BOTH SIDE L5 SCREW WAS FOUND. ALL THE IMPLANT WERE EXTRACTED ON (B)(6) 2016 BECAUSE THE FRACTURE HAS HEALED. THE TIPS OF THE BROKEN SCREWS WERE REMAINED TO THE PATIENT. UPDATE: PATIENT WAS FUSED.
Patient problems: Foreign Body In Patient; Injury
MAUDE report 5957731 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page