Skip to content

V40 Head 6260-5-232Recalls & reported problems

Howmedica Osteonics Corp. · V40 Femoral Head · Class 2 · Implant

All models of V40 Head (30)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613327012903), received between Jun 13, 2018 and Jan 14, 2025 — classified as 11 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated6
  • Material Erosion4
  • Degraded2
  • Detachment of Device or Device Component2
  • Corroded1
  • Insufficient Device Problem Information1
  • Loss of Osseointegration1
  • Noise, Audible1

Patient problems

  • Injury7
  • Pain4
  • Ambulation or Postural Difficulties2
  • Foreign Body Reaction2
  • Metal Related Pathology2
  • Reaction2
  • Insufficient Information1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 14, 2025Material Erosion; Material Erosion

    IT WAS REPORTED THAT ¿TRUNNIONOSIS OF HIP IMPLANT - RESULTING IN BROKEN IMPLANT NEEDING TO BE REVISED SURGICALLY¿. THIS EVENT WAS REPORTED TO STRYKER CANADA BY HEALTH CANADA VIA MANDATORY HOSPITAL REPORTING: HEALTH CANADA REFERENCE #: (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 21147435 ↗
  • InjuryAug 24, 2023Degraded; Material Erosion; Detachment of Device or Device Component; Degraded; Material Erosion; Detachment of Device or Device Component

    THE NECK OF THE ACCOLADE TMZF WAS DAMAGED DUE TO WEAR, WHICH CAUSED THE METAL HEAD TO COME OFF. METALOSIS WAS ALSO CONFIRMED DURING REVISION SURGERY.

    Patient problems: Ambulation or Postural Difficulties; Foreign Body Reaction; Metal Related Pathology

    MAUDE report 17609800 ↗
  • InjuryJan 21, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    RIGHT SIDE HIP REVISION, ALL COMPONENTS EXPLANTED WERE STRYKER COMPONENTS. THE PATIENT WAS RE-IMPLANTED WITH NEW REVISION STRYKER COMPONENTS. THE PATIENT RECENTLY WAS REPORTING PAIN IN HER RIGHT HIP, X-RAYS SHOWED A DISASSOCIATION AND INTRAPROSTHETIC DISLOCATION OF THE FEMORAL HEAD FROM THE TRUNNION OF THE STEM. THE PATIENT¿S COBALT CHROME LEVELS WERE ELEVATED AND THE STEM/FEMORAL HEAD JUNCTION WAS FAILED, SO THE SURGEON NEEDED TO REVISE THE FEMORAL COMPONENT. THERE WERE ADVERSE LOCAL TISSUE REACTIONS SURROUNDING THE COMPONENT, THE FEMORAL HEAD HAD CORROSION AND SEVERE WEAR WAS NOTED ON THE TRUNNION AS THE FEMORAL HEAD WAS FRACTURED OFF.

    Patient problems: Ambulation or Postural Difficulties; Foreign Body Reaction; Metal Related Pathology; Pain

    MAUDE report 13314414 ↗
  • InjuryDec 10, 2021Corroded; Material Erosion; Loss of Osseointegration; Corroded; Material Erosion; Loss of Osseointegration

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO PAIN AND SUPERIOR LUCENCY AROUND THE ACETABULAR SCREWS. INTRA-OPERATIVELY, THE FOLLOWING WERE NOTED: SOME TISSUE STAINING, CORROSION INSIDE OF THE FEMORAL HEAD (NO CORROSION OR WEAR WAS NOTED ON THE TRUNNION), AND ONCE THE SCREWS WERE REMOVED, THE SHELL CAME OUT EASILY. THE HEAD, LINER, SHELL, AND SCREWS WERE REVISED. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED LINER. REP CAN PROVIDE PICTURES AND OTHERWISE CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Pain

    MAUDE report 12980523 ↗
  • InjuryAug 6, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT PRESENTED WITH PERIPROSTHETIC HIP FRACTURE. REVISION WAS DONE. *UPDATE ON 10-JUL-2020: "(B)(6) 2017 OP NOTE - REVISION LEFT TOTAL HIP  PRE/POST-OP DIAGNOSIS "FAILURE LEFT TOTAL HIP REPLACEMENT" GEN ANESTHESIA, "EXTENDED TROCHANTERIC OSTEOTOMY. SIGNIFICANT TITANIUM DEBRIS. FEMORAL STEM WELL FIXED. TRUNNION ERODED. BEAKLIKE IN SHAPE. HEAD DISASSOCIATED. ACET. SHELL WELL FIXED. " PER CLINICIAN REVIEW COMMENT: "THE EVENT DESCRIPTION STATES "PERIPROSTHETIC FRACTURE" BUT THE REVISION OP REPORT DESCRIBES A PROSTHETIC DISASSOCIATION AND EROSION. BASED UPON THE INFORMATION AVAILABLE FOR REVIEW, NO CREATION OF A MEDICAL REPORT IS POSSIBLE FOR THIS CASE."

    Patient problems: Injury

    MAUDE report 10375815 ↗
  • InjuryApr 9, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THROUGH A MEDWATCH 5093292: "PATIENT ID:[...]; ON [...] 2006, HE UNDERWENT A RIGHT TOTAL HIP ARTHROPLASTY WITH STRYKER (HOWMEDICA) COMPONENTS, UNCEMENTED PSL 54 OD SOCKET, 32MM CROSSFIRE POLYETHYLENE INSERT. ACCOLADE HFX STEM SIZE 3 (REF #6077-03358), AND A 32MM +4 LFIT ANATOMIC V40 HEAD (REF #6077-0335) AND A 32MM +4 LFIT ANATOMIC V40 HEAD (REF #6260-5-232). IN [...] OF 2016, HE HAD A SUDDEN EPISODE OF WEAKNESS IN THE RIGHT LEG, WHICH CONCERNED HIM AND PROMPTED EVALUATION OF HIS RIGHT HIP REPLACEMENT. IN THE YEAR PRECEDING THAT EPISODE, HE HAD BEEN EXPERIENCING A SIGNIFICANT TREMOR OF HIS HANDS, FATIGUE AND REDUCED STAMINA. METAL SUPPRESSION MRI OF RIGHT HIP ON [...] 2016 SHOWED A SMALL PSEUDOTUMOR JUST SUPERIOR TO THE FEMORAL NECK WITH INVOLVEMENT OF THE GLUTEUS MEDIUS TENDON. FDG PET BRAIN SCAN AND NEURO Q ANALYSIS SHOWED ABNORMAL HYPOMETABOLISM IN FOCAL AND GENERALIZED DUSTER REGIONS IN A PATTERN COMPATIBLE WITH CHRONIC TOXIC ENCEPHALOPATHY. ON [...] 2016, SERUM/ PLASMA COBALT LEVEL WAS 5.8 MCG/L AND CHROMIUM WAS 1.7MCG/L. ON [...] 2016, RIGHT HIP WAS REVISED TO A ZIMMER WAGNER STEM 190 BY 14MM, TRIDENT X3 32ID LATERALLY OFFSET LINER, A 32MM DELTA CERAMIC +0 HEA

    Patient problems: Injury; Reaction

    MAUDE report 9944146 ↗
  • InjuryMar 2, 2020Material Erosion; Material Erosion

    AT (B)(6) 2012, THA WAS DONE WITH ACCOLADE TMZF, TRIDENT CUP, X3 LINNER AND METAL HEAD V40 TEPER. AT (B)(6) 2020, THE PATIENT FELT UNCOMFORTABLE AND WENT TO THE HOSPITAL. EXAMINATION AT THE HOSPITAL CONFIRMED TRUNNIONOSIS AND THE REVISION SURGERY OF STEM, LINNER, HEAD IS SCHEDULED FOR AROUND (B)(6) 2020.

    Patient problems: Injury

    MAUDE report 9775869 ↗
  • InjuryFeb 24, 2020Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information

    2019/DEC, IT WAS COMFIRMED A ARMD (ADVERSE REACTIONS TO METAL DEBRIS). ON (B)(6) 2020 REVISION SURGERY DID. IT WAS NO TRANIONOSIS.

    Patient problems: Injury

    MAUDE report 9746617 ↗
  • InjurySep 24, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AS REPORTED: "PAIN & SHORTENING LEG. X-RAY REVEALED HEAD [DISASSOCIATED] FROM STEM. ALL AVAILABLE INFORMATION SUBMITTED." PATIENT WAS REVISED TO "RESTORATION W/ MDM." REP PROVIDED EXPLANT PICTURES AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: Injury; Pain

    MAUDE report 9109041 ↗
  • InjuryAug 8, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT PRESENTED WITH PAIN IN RIGHT HIP AT THE EMERGENCY DEPARTMENT. THE PATIENT REPORTED THE PAIN STARTED APPROXIMATELY ONE YEAR AGO. IT WAS FURTHER REPORTED THAT A SUBSEQUENT X-RAY SHOWED HEAD DISSOCIATED FROM STEM. PATIENT REQUIRED REVISION HIP SURGERY TO REMOVE HIP STEM.

    Patient problems: Injury; Pain

    MAUDE report 8873362 ↗
  • InjuryJun 13, 2018Device Dislodged or Dislocated; Noise, Audible; Device Dislodged or Dislocated; Noise, Audible

    THE PATIENT UNDERWENT THA SURGERY WITH CENTPILLAR ON (B)(6) 2009. THE PATIENT CLAIMED ABNORMAL SOUND FROM THE LEFT HIP AND A DISLOCATION OF THE METAL HEAD WAS CONFIRMED. REVISION SURGERY WAS PERFORMED ON (B)(6) 2018, AND AN INSERT, THE METAL HEAD AND THE STEM WERE REPLACED. WEAR OF THE STEM NECK WAS NOTICED. ALSO, INSIDE OF THE PATIENT'S HIP JOINT WAS BLACK BY METAL ABRASION POWDER.

    Patient problems: Injury; Reaction

    MAUDE report 7597103 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices →

Sources and updates