Triathlon 5550-G-319Recalls & reported problems
Howmedica Osteonics Corp. · Symmetric Patella · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07613327014761), received between Nov 12, 2018 and Aug 2, 2022 — classified as 10 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information3
- Unstable3
- Loss of Osseointegration2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
- Noise, Audible1
- Osseointegration Problem1
Patient problems
- Inadequate Osseointegration3
- Pain3
- Injury2
- Ambulation or Postural Difficulties1
- Fall1
- Fibrosis1
- Synovitis1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryAug 2, 2022Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO THE POST OF THE INSERT SHEARING OFF (SURGEON REPORTED THAT THE PATIENT WAS PRONE TO MULTIPLE FALLS). INTRA-OPERATIVELY, THE SURGEON NOTICED A HAIRLINE FRACTURE OF THE PATELLAR COMPONENT. WHILE ADDRESSING IT, THE PATELLAR COMPONENT BROKE IN HALF. A 6X9 PS INSERT WAS REVISED TO A 6X11 CS INSERT AND THE PATELLA WAS REVISED.
Patient problems: Fall
MAUDE report 15143480 ↗InjuryMay 9, 2022Unstable; Unstable
PATIENT CALLED AND IS INQUIRING ABOUT RECALL FOR HER RIGHT KNEE SURGERY ON (B)(6) 2016. UPDATE (B)(6) 2022: PATIENT IS EXPERIENCING INSTABILITY. SHE HAS SEEN THREE DIFFERENT SURGEONS AND HAS UPCOMING REVISION SURGERY IN (B)(6). PATIENT WOULD LIKE TO KNOW WHAT IMPLANTS SHE CAN USE IN HER REVISION FOR A BETTER OUTCOME.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 14342922 ↗InjuryNov 15, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
PATIENT HAD BILATERAL SIMULTANEOUS INDEX TKA ON (B)(6) 2015. PATIENT COMPLAINS OF RIGHT ANTERIOR PAIN WITH PATELLAR LYSIS (SUSPICION OF PJI OR TUMOR). UNDERGOES REVISION ON (B)(6), 2021. NO INFECTION, OR TUMOR. EXCHANGED PATELLA AND LINER ONLY.
Patient problems: Pain
MAUDE report 12813473 ↗InjurySep 22, 2021Noise, Audible; Noise, Audible
THIS PR WAS RAISED FROM THE MEDICAL REVIEW. THE DATE OF THE PROCEDURE WAS (B)(6) 2012. IN THE DISCHARGE SUMMARY IT NOTES THAT THE PATIENT WAS HAVING A POPPING SENSATION AROUND PATELLA AND WAS ADMITTED FOR EXPLORATION OF THE JOINT. SHE WAS NOTED TO HAVE EXTENSIVE SYNOVIAL HYPERTROPHY AND FIBROSIS AROUND THE PATELLA.
Patient problems: Fibrosis; Synovitis
MAUDE report 12514165 ↗InjuryNov 9, 2020Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED BY THE PATIENT'S ATTORNEY AS A RESULT OF A LEGAL CLAIM THAT ALLEGEDLY ON (B)(6) 2017 THE PATIENT UNDERWENT A LEFT TOTAL KNEE ARTHROPLASTY USING SIMPLEX HV BONE CEMENT. IT IS FURTHER ALLEGED THAT THE PATIENT WAS REVISED ON (B)(6) 2018 DUE TO ALLEGED LOOSENING OF HIS KNEE.
Patient problems: Inadequate Osseointegration
MAUDE report 10814573 ↗InjuryNov 9, 2020Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED BY THE PATIENT'S ATTORNEY THROUGH A LEGAL CLAIM THAT ALLEGEDLY ON (B)(6) 2018 THE PATIENT UNDERWENT A LEFT TOTAL KNEE ARTHROPLASTY WITH SIMPLEX HV BONE CEMENT WITH GENTAMICIN. IT IS FURTHER ALLEGED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT REVISION SURGERY DUE TO ALLEGED LOOSENING.
Patient problems: Inadequate Osseointegration
MAUDE report 10807873 ↗InjuryJun 11, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO " FAILED RIGHT TOTAL KNEE ARTHROPLASTY FOR INSTABILITY".
Patient problems: Injury
MAUDE report 8687090 ↗InjuryMar 18, 2019Unstable; Insufficient Device Problem Information; Unstable; Insufficient Device Problem Information
AS REPORTED: "DOCTOR REVISED INSERT AND PATELLA DUE TO INSTABILITY. NO MORE INFORMATION AVAILABLE". SPOKE TO REP. HOSPITAL WILL NOT ALLOW THE RELEASE OF ANY FURTHER INFORMATION.
Patient problems: Injury
MAUDE report 8429167 ↗InjuryFeb 21, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED: "THE PATIENT EXPERIENCED A LOT OF PAIN WHEN STANDING UP, SITTING DOWN, AND WHEN WALKING. THE PATELLA IMPLANT FAILED TO TRACK PROPERLY ON THE AVON PATELLOFEMORAL COMPONENT. WE CONVERTED INTO A PRIMARY PS TRIATHLON KNEE". REP REPORTED THAT HOSPITAL WOULD NOT RELEASE X-RAYS TO STRYKER.
Patient problems: Pain
MAUDE report 8357502 ↗InjuryNov 12, 2018Loss of Osseointegration; Loss of Osseointegration
PATIENT REPORTED HE IS EXPERIENCING LOOSENING AND PAIN AFTER HIS RIGHT KNEE SURGERY.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 8060113 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page