TriathlonRecalls & reported problems across all models

Howmedica Osteonics Corp. · 608 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

15 open recalls name at least one Triathlon model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
3,524
Deaths reported
2
Injuries reported
3,223
Recalls
15

We found 608 models of Triathlon by Howmedica Osteonics Corp. in the FDA's device registry (GUDID). Together they are named in 3,524 adverse event reports in the FDA's MAUDE database, received between Jan 4, 2017 and Sep 25, 2026 — classified as 2 deaths, 3,223 injuries, 299 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (819 reports).

15 FDA recall records name at least one of these models by its device identifier, 15 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 24 reportsDec 2023: 27 reportsJan 2024: 29 reportsFeb 2024: 29 reportsMar 2024: 25 reportsApr 2024: 23 reportsMay 2024: 30 reportsJun 2024: 27 reportsJul 2024: 17 reportsAug 2024: 22 reportsSep 2024: 35 reportsOct 2024: 31 reportsNov 2024: 24 reportsDec 2024: 19 reportsJan 2025: 24 reportsFeb 2025: 21 reportsMar 2025: 16 reportsApr 2025: 19 reportsMay 2025: 22 reportsJun 2025: 13 reportsJul 2025: 46 reportsAug 2025: 22 reportsSep 2025: 24 reportsOct 2025: 22 reportsNov 2025: 19 reportsDec 2025: 21 reportsJan 2026: 14 reportsFeb 2026: 14 reportsMar 2026: 18 reportsApr 2026: 15 reportsMay 2026: 27 reportsJun 2026: 11 reportsJul 2026: 13 reportsAug 2026: 25 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202646
Last 36 months. The most recent month may be incomplete.

Triathlon recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 30, 2026Open, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3198-2026 · Root cause: Process change control · names 2 of these models · Recall details · FDA record ↗

  2. May 19, 2026Open, Classified

    Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).

    Z-2462-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. May 19, 2026Open, Classified

    Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).

    Z-2461-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Sep 4, 2024Open, Classified

    Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.

    Z-0023-2025 · Root cause: Process control · Recall details · FDA record ↗

  5. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1266-2024 · Root cause: Packaging · Recall details · FDA record ↗

  6. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1265-2024 · Root cause: Packaging · Recall details · FDA record ↗

  7. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1264-2024 · Root cause: Packaging · Recall details · FDA record ↗

  8. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1263-2024 · Root cause: Packaging · Recall details · FDA record ↗

  9. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1262-2024 · Root cause: Packaging · Recall details · FDA record ↗

  10. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1261-2024 · Root cause: Packaging · Recall details · FDA record ↗

  11. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1260-2024 · Root cause: Packaging · Recall details · FDA record ↗

  12. Feb 6, 2024Open, Classified

    Potential packaging breaches of inner blister and outer sterile blister.

    Z-1259-2024 · Root cause: Packaging · Recall details · FDA record ↗

  13. Apr 13, 2023Open, Classified

    Mislabeled:Labeled Catalog Number: 5531-G-312-E, TRIATHLON X3 TIBIAL BEARING INSERT CS #3 12MM, may contain Catalog Number: 5532-G-413-E, TRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM, and vice versa

    Z-1654-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  14. Apr 13, 2023Open, Classified

    Mislabeled:Labeled Catalog Number: 5531-G-312-E, TRIATHLON X3 TIBIAL BEARING INSERT CS #3 12MM, may contain Catalog Number: 5532-G-413-E, TRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM, and vice versa

    Z-1653-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  15. Mar 30, 2022Open, Classified

    Stryker has discovered a potential product mix where the size of the Triathlon Tritanium Tibial Component inside the package does not match the package labeling.

    Z-1237-2022 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

Most reported Triathlon problems

Device problems

  • Adverse Event Without Identified Device or Use Problem849
  • Insufficient Device Problem Information819
  • Unstable665
  • Loss of Osseointegration291
  • Fracture182
  • Osseointegration Problem164
  • Degraded104
  • Malposition of Device92

Patient problems

  • Injury1,300
  • Pain749
  • Unspecified Infection693
  • Inadequate Osseointegration441
  • Joint Laxity359
  • Ambulation or Postural Difficulties304
  • Insufficient Information265
  • Loss of Range of Motion183

Triathlon models

608 models, 100 per page, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Triathlon adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Howmedica Osteonics Corp. products

All Howmedica Osteonics Corp. devices →

Sources and updates