Triathlon 5551-G-320Recalls & reported problems
Howmedica Osteonics Corp. · Asymmetric Patella · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Howmedica Triathlon kneecap implant is used in knee replacement surgery. Reports mostly describe knee instability, loosening or wear of the kneecap, often leading to repeat surgery to replace parts. There are 26 reports naming its identifier, with no deaths and 25 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 07613327014815), received between Jan 29, 2018 and Dec 11, 2025 — classified as 24 injuries, 1 malfunction.
The most frequently reported specific device problem is “Unstable” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable10
- Insufficient Device Problem Information4
- Osseointegration Problem4
- Loss of Osseointegration3
- Degraded1
- Device Contamination with Chemical or Other Material1
- Fracture1
- Mechanical Problem1
Patient problems
- Injury19
- Inadequate Osseointegration6
- Pain5
- Ambulation or Postural Difficulties3
- Joint Laxity2
- Arthritis1
- Limited Mobility Of The Implanted Joint1
- Loss of Range of Motion1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
InjuryDec 11, 2025Degraded; Degraded
THE FOLLOWING INFORMATION WAS PROVIDED: LEFT TOTAL KNEE. THE REVISION TODAY WAS DUE TO HYPER EXTENSION AND INSTABILITY. THE 11 MM INSERT WAS REMOVED AND REPLACED WITH A 14 MM INSERT TO ADDRESS THE INSTABILITY AND HYPER EXTENSION. IT WAS ALSO DISCOVERED THAT THE PATELLA IMPLANT WAS SHOWING SIGNS OF WEAR. THE PATELLA WAS REVISED TO A COMPETITOR PATELLA AS THAT IS WHAT WAS AVAILABLE AT THE HOSPITAL. THIS SURGERY WAS ADDED ON LATE AND INSERTS WERE REQUESTED SO PATELLA COMPONENTS WERE NOT ORDERED IN. NO FURTHER INFORMATION AVAILABLE FROM THE DOCTOR OR HOSPITAL.
Patient problems: Ambulation or Postural Difficulties; Joint Laxity
MAUDE report 23783546 ↗InjuryMar 28, 2023Fracture; Osseointegration Problem; Fracture; Osseointegration Problem
PATIENT STATED THAT HIS RIGHT TRIATHLON KNEE IS "COMING APART", PRIMARY SURGERY (B)(6) 2011. STATES HIS KNEE CAP IS LOOSE AND EXPERIENCES SORENESS. PRIMARY SURGEON STATED TO HIM HE WOULD "NOT" DO A REVISION DUE TO HIS AGE AND HEART. PATIENT SOUGHT A SECOND OPINION, FEBRUARY 2023, DR. CONFIRMED THAT HIS KNEE CAP IS LOOSE. XRAYS CONFIRM THAT THE KNEE CAP IS IN "TWO PIECES". HE ALSO HAD A PROTRUDING BULGE ON THE KNEE AT THE TIME OF THE SECOND OPINION. PATIENT STATED THAT SINCE HIS SECOND OPINION IN FEBRUARY HE NOW HAS 2-3 KNEE BULGES/LUMPS THAT ARE PALPABLE AND PROTRUDING HOWEVER CANNOT CONFIRM THEY ARE ACTUAL KNEE CAP "PIECES".
Patient problems: Pain
MAUDE report 16626578 ↗InjuryApr 25, 2022Unstable; Unstable
FOLLOW-UP OF TRIATHLON TRITANIUM OUTCOMES STUDY, PATIENT REPORTED DISSATISFACTION WITH THEIR KNEE REPLACEMENT. ADDITIONAL COMMENTS: "SWOLLEN KNEE. UNSTABLE WHEN GETTING UP. DIFFICULTY WITH PUTTING ON SOCKS, TYING SHOES LACES; AND GETTING OUT OF DEEP CHAIRS."
Patient problems: Ambulation or Postural Difficulties; Swelling/ Edema
MAUDE report 14192812 ↗InjuryMar 12, 2021Unstable; Unstable
REMOVAL OF PRIMARY TRIATHLON IMPLANTS DUE TO INSTABILITY AND REPLACEMENT WITH TRIATHLON TS
Patient problems: Joint Laxity
MAUDE report 11474325 ↗InjurySep 22, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED: PATIENT HAD VISUAL PATELLA TILT WITH PAIN AND NOISE. THE FEMORAL COMPONENT, 4X9 CR INSERT, AND PATELLAR COMPONENT OF THE PATIENT'S LEFT KNEE WERE REVISED.
Patient problems: Pain
MAUDE report 10570002 ↗InjuryOct 22, 2019Unintended Movement; Migration; Unintended Movement; Migration
IT WAS NOTED THAT THE LEFT KNEE HAD AN ARTHROSCOPIC LATERAL RELEASE ON (B)(6) 2010.
Patient problems: Injury; Limited Mobility Of The Implanted Joint
MAUDE report 9219300 ↗InjurySep 4, 2019Naturally Worn; Naturally Worn
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO PAINFUL LEFT TOTAL KNEE WITH OSTEOLYSIS OF THE TIBIAL COMPONENT, POLYETHYLENE WEAR.
Patient problems: Injury; Pain
MAUDE report 8960141 ↗InjuryAug 9, 2019Unstable; Unstable
THIS PI IS FOR THE LEFT KNEE. IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO RIGHT AND LEFT KNEE POSTOPERATIVE TOTAL KNEE ARTHROPLASTY WITH INSTABILITY.
Patient problems: Injury
MAUDE report 8879983 ↗InjuryAug 9, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO RIGHT KNEE CORONAL PLANE INSTABILITY.
Patient problems: Injury
MAUDE report 8879888 ↗InjuryAug 9, 2019Unstable; Unstable
THIS PI IS FOR THE RIGHT KNEE. IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED . DUE TO RIGHT AND LEFT KNEE POSTOPERATIVE TOTAL KNEE ARTHROPLASTY WITH INSTABILITY.
Patient problems: Injury
MAUDE report 8878777 ↗InjuryJul 2, 2019Unstable; Unstable
THIS PI IS FOR THE LEFT KNEE IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "UNSTABLE LEFT TOTAL KNEE ARTHROPLASTY, ALL COMPONENTS." UPDATE (B)(6) 2019: AS PER MEDICAL REVIEW DATED (B)(6) 2019 "COMPONENT MALPOSITION OF TRIATHLON PS KNEE COMPONENTS WITH A ¿SIGNIFICANTLY ELEVATED JOINT LINE¿ HAS CONTRIBUTED TO GROSS INSTABILITY OF THE ARTHROPLASTY REQUIRING EXCHANGE OF ALL KNEE DEVICES AFTER 5-YEARS".
Patient problems: Injury
MAUDE report 8753918 ↗InjuryJun 20, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "GLOBAL INSTABILITY OF RIGHT KNEE FOLLOWING TOTAL KNEE ARTHROPLASTY OF RIGHT KNEE WITH CHRONIC EFFUSIONS".
Patient problems: Injury
MAUDE report 8716911 ↗InjuryJun 20, 2019Mechanical Problem; Insufficient Device Problem Information; Mechanical Problem; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "MECHANICAL FAILURE OF RIGHT TOTAL KNEE ARTHROPLASTY".
Patient problems: Injury
MAUDE report 8716836 ↗InjuryJun 20, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "RIGHT KNEE PROSTHESIS INSTABILITY".
Patient problems: Injury
MAUDE report 8716089 ↗InjuryJun 19, 2019Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO " LEFT TOTAL KNEE ARTHROPLASTY ASEPTIC LOOSENING".
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8712739 ↗InjuryJun 19, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO " LEFT TOTAL KNEE ARTHROPLASTY WITH LAXITY AND PAIN".
Patient problems: Injury; Loss of Range of Motion; Pain
MAUDE report 8712493 ↗InjuryMay 22, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "FAILED LEFT TOTAL KNEE ARTHROPLASTY WITH CATASTROPHIC POLYETHYLENE FAILURE".
Patient problems: Injury
MAUDE report 8633080 ↗InjuryMay 1, 2019Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "LOOSE TOTAL KNEE REPLACEMENT OF LEFT KNEE".
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8568439 ↗InjuryApr 11, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "INSTABILITY OF THE LEFT KNEE."
Patient problems: Ambulation or Postural Difficulties; Injury
MAUDE report 8506486 ↗InjuryApr 11, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "RIGHT TOTAL KNEE ARTHROPLASTY FAILURE WITH POLYETHYLENE WEAR, OSTEOLYSIS, AND ANTERIOR TO POSTERIOR INSTABILITY".
Patient problems: Injury; Osteolysis
MAUDE report 8505476 ↗InjuryApr 3, 2019Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "INSTABILITY, LEFT TOTAL KNEE WITH LOOSE PATELLAR COMPONENT."
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8479680 ↗InjuryMar 15, 2019Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "LOOSE IMPLANTS RIGHT TOTAL KNEE ARTHROPLASTY".
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8424761 ↗InjuryMar 15, 2019Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "RIGHT KNEE OSTEOARTHRITIS AND RIGHT KNEE TOTAL ARTHROPLASTY FAILURE OR LOOSENING."
Patient problems: Arthritis; Inadequate Osseointegration; Injury
MAUDE report 8424668 ↗InjuryJul 27, 2018Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT PATIENT'S RIGHT KNEE WAS REVISED DUE TO ASEPTIC LOOSENING AND PAIN.
Patient problems: Inadequate Osseointegration; Injury; Pain
MAUDE report 7725703 ↗MalfunctionJan 29, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
UPON OPENING THE PATELLA IMPLANT, A HAIR WAS FOUND ON THE DEVICE. REP WAS PRESENT IN THE OPERATING ROOM AT THE TIME OF DISCOVERY. DUE TO STERILITY CONCERNS, THE DEVICE WAS NOT USED AND REMOVED IMMEDIATELY FROM THE FIELD. A REPLACEMENT DEVICE WAS IMMEDIATELY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7225666 ↗
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- ATTUNE 1512-16-080Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page