Triathlon 5551-G-381Recalls & reported problems
Howmedica Osteonics Corp. · Asymmetric Patella · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327014822), received between Aug 2, 2018 and Apr 2, 2021 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Unstable2
- Degraded1
- Device Contamination with Chemical or Other Material1
- Fracture1
- Loss of Osseointegration1
- Mechanical Problem1
- Osseointegration Problem1
Patient problems
- Injury5
- Inadequate Osseointegration2
- Insufficient Information2
- Ambulation or Postural Difficulties1
- Loss of Range of Motion1
- No Known Impact Or Consequence To Patient1
- Pain1
- Swelling1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryApr 2, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATELLA FAILED.
Patient problems: Insufficient Information
MAUDE report 11607480 ↗InjuryApr 2, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATELLA FAILED.
Patient problems: Insufficient Information
MAUDE report 11607479 ↗InjuryJan 31, 2020Fracture; Mechanical Problem; Loss of Osseointegration; Fracture; Mechanical Problem; Loss of Osseointegration
PROCEDURE: THE PATIENT UNDERWENT RIGHT TOTAL KNEE REPLACEMENT (B)(6) 2016. PATIENT HAD A LOT OF PAIN, RADIOGRAPHS REVEALED PATELLA FRACTURE WITH LOOSEN COMPONENT. THEN PATIENT UNDERWENT RIGHT TOTAL KNEE REVISION AND PATELLA WAS EXCHANGED (B)(6) 2019.
Patient problems: Ambulation or Postural Difficulties; Inadequate Osseointegration; Injury
MAUDE report 9652531 ↗MalfunctionSep 9, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
A TRIATHLON IMPLANT TRIATHLON ASYM X3 PAT 38X11MM WHICH WAS EXPOSED TO WATER AND STEAM AT (B)(6) HOSPITAL, WAS REDISTRIBUTED BY STRYKER AUSTRALIA AND SUBSEQUENTLY IMPLANTED DURING A PROCEDURE AT THE (B)(6) HOSPITAL. A FULL LOAN TRIATHLON IMPLANT SET WAS ACCIDENTALLY RETRIEVED FROM A HOSPITAL BY A COURIER AND WAS RETURNED TO A STRYKER KIT ROOM. HOWEVER, THE COURIER WAS UNAWARE THAT THE ITEMS WERE AFFECTED AND THEREFORE DID NOT COMMUNICATE TO THE KIT ROOM THAT THESE PARTS WERE PREVIOUSLY QUARANTINED AT THE HOSPITAL. AS A RESULT THE KIT ROOM DID NOT PROCEED WITH ANY QUARANTINE ACTION AND 18 OF THE AFFECTED ITEMS WERE RE-DISTRIBUTED. UPDATE 19/AUGUST/2019 WG: NONE OF THE PACKAGING WAS RETAINED FROM THE IMPLANTED DEVICES. SURGEONS EXPRESSED NO CONCERN, AND HAVE COMMUNICATED NO PLANS TO SURGICALLY ADDRESS THE IMPLANTATION OF THE DEVICES. TO DATE, NO ADVERSE CONSEQUENCES HAVE BEEN REPORTED FOR THESE PATIENTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8974364 ↗InjuryAug 13, 2019Degraded; Osseointegration Problem; Degraded; Osseointegration Problem
PATIENT'S WIFE CALLED REPORTING PATIENT HAD LEFT KNEE REPLACEMENT SURGERY (B)(6) 2013. PATIENT BEGAN EXPERIENCING RANGE OF MOTION ISSUES AND WAS REVISED ON (B)(6) 2018. PATIENT'S WIFE REPORTS THAT REVISING SURGEON SHOWED HER AND HER HUSBAND PICTURES OF THE EXPLANTED DEVICE, AND EXPLAINED THAT "THE PLASTIC WAS BUBBLED" AND THAT THE SURGEON REPORTED THAT THE DEVICE WAS REMOVED WITH VIRTUALLY NO RESISTANCE OR EFFORT, AND THAT THE DEVICE HAD SEVERED THE PATIENT'S MEDIAL LATERAL COLLATERAL LIGAMENT. SURGEON REPORTED TO PATIENT THE DEVICE WOULD BE GIVEN TO THE STRYKER SALES REP FOR RETURN TO STRYKER. SPOKE TO SALES REP. REP REPORTED THAT HOSPITAL'S POLICY WILL NOT ALLOW THE RELEASE OF EXPLANTED DEVICES, AND THAT HE HAS NO RECOLLECTION OF HAVING RECEIVED A DEVICE FROM THE SURGEON FOR RETURN. PATIENT IS INQUIRING AS TO WHETHER THE REVISED DEVICE(S) IS/ARE SUBJECT TO ANY RECALL, AND IS INQUIRING ABOUT COMPENSATION. PATIENT DOES NOT HAVE IMPLANT INFORMATION AT THE TIME OF REPORT. UPDATE AS PER MEDICAL REVIEW"...SUSPICION THIS WAS SEVERE CASE OF ASEPTIC LOOSENING¿FEMUR HAD SOME AREAS THAT HAD BONDED WITH THE CEMENT.. TIBIA GROSSLY LOOSE.. REMOVED BY HAND¿ FEMUR AND PATELLA REMOVED WITHOUT SIG
Patient problems: Inadequate Osseointegration; Injury; Loss of Range of Motion
MAUDE report 8887632 ↗InjuryMay 21, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "LEFT KNEE MID FLEXION INSTABILITY AND A CRUCIATE-RETAINING KNEE." UPDATE 18 OCTOBER, 2019: MEDICAL REVIEW NOTED THE FOLLOWING: REVISION DIAGNOSIS AFTER LESS THAN 5-YEARS OF IMPLANTATION WAS REPORTED AS ¿MID FLEXION INSTABILITY¿ PRIOR TO REVISION ALTHOUGH STABILITY TESTING DURING SURGERY IS MORE CONSISTENT WITH FULL FLEXION INSTABILITY. THIS MEANS THE KNEE IS (MORE OR LESS) STABLE IN EXTENSION BUT BECOMES PROGRESSIVELY MORE UNSTABLE IN AP (ANTEROPOSTERIOR) DIRECTION WITH INCREASING FLEXION. ALTHOUGH THERE ARE UNFORTUNATELY NO X-RAYS AVAILABLE AT THIS TIME TO DETAIL THE PROBLEM, THE CAUSE OF SUCH FLEXION INSTABILITY IS ALWAYS RELATED TO SUBOPTIMAL COMPONENT POSITION. CONCLUSION OF ASSESSMENT TRIATHLON COMPONENT MALPOSITION HAS CONTRIBUTED TO AN UNEQUAL FLEXION-EXTENSION GAP SPACE IN THE KNEE TO CONTRIBUTE TO SYMPTOMATIC FLEXION INSTABILITY OF THE IMPLANTED TRIATHLON DEVICES REQUIRING REVISION SURGERY WITH BEARING EXCHANGE WITHIN 5-YEARS OF PRIMARY IMPLANTATION.
Patient problems: Injury
MAUDE report 8629334 ↗InjuryMay 1, 2019Unstable; Unstable
IT WAS REPORTED THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "INSTABILITY RIGHT TOTAL KNEE REPLACEMENT ARTHROPLASTY."
Patient problems: Injury
MAUDE report 8568772 ↗InjuryDec 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THROUGH MEDWATCH (MW5080901) "HAVE STRYKER TOTAL KNEE REPLACEMENT IN 2015. STILL KNEE HURTS AND SWELLING. LUCENCY BETWEEN FEMORAL COMPONENT AND THE FEMORAL ON THE ANTERIOR ASPECT, RIGHT KNEE IS STILL GIVING OUT. "HALE" OSTEOPOROSIS, STILL WEARING KNEE BRACE, CANE AND WALKER. ALSO, THERE IS A SLIGHT LATERAL TILT TO THE PATELLA, STRYKER TOTAL KNEE REPLACEMENT WILL NOT STOP HURTING, WILL NOT STOP SWELLING, PAIN IS A LOT. DOCTOR SAID I NEED TOTAL KNEE REPLACEMENT."
Patient problems: Pain; Swelling
MAUDE report 8148710 ↗InjuryAug 2, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT'S RIGHT KNEE WAS REVISED DUE TO MECHANICAL COMPLICATIONS. SURGEON WOULD NOT ELABORATE ON THE MECHANICAL COMPLICATIONS BUT EMPHATICALLY RULED OUT THE DEVICES HAVING CAUSED OR CONTRIBUTED TO THE COMPLICATIONS. PATIENT'S ENTIRE KNEE CONSTRUCT WAS REVISED TO A TRIATHLON TS CONSTRUCT.
Patient problems: Injury
MAUDE report 7742567 ↗
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- ATTUNE 1512-16-080Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page