Skip to content

C-Taper Head 06-2800Recalls & reported problems

Howmedica Osteonics Corp. · Low Friction Ion Treatment · Class 2 · Implant

All models of C-Taper Head (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327016260), received between Jul 19, 2017 and Jul 28, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K910988, K993601); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Material Deformation1

Patient problems

  • Ambulation or Postural Difficulties1
  • Inflammation1
  • Metal Related Pathology1
  • No Known Impact Or Consequence To Patient1
  • Reaction1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 28, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    REVISION LEFT THA. PER SURGEON, PATIENT REQUIRED LENGTHENING OF THE LEFT SIDE. NO ADDITIONAL DETAILS PROVIDED BY THE FACILITY.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 22636273 ↗
  • InjuryOct 4, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS RIGHT HIP ON OR ABOUT (B)(6) 2015. IT IS FURTHER ALLEGED THAT HE SUFFERED INJURIES AS A RESULT OF IMPLANTATION OF THE DEVICE AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HIS BLOOD. PATIENT HAS NOT YET SCHEDULED A SURGERY FOR EXPLANTATION OF THE FEMORAL HEAD AT ISSUE.

    Patient problems: Inflammation; Metal Related Pathology; Reaction

    MAUDE report 9156020 ↗
  • MalfunctionJul 19, 2017Device Dislodged or Dislocated; Material Deformation; Device Dislodged or Dislocated; Material Deformation

    AFTER IMPACTING THE FEMORAL HEAD, THE SURGEON WENT TO REDUCE THE HIP AND THE INNER C-TAPER CAME OFF THE TRUNNION OF THE STEM. THAT LEFT THE HEAD/BIPOLAR COMPONENT IN THE SOCKET. A NEW HEAD/BIPOLAR COMPONENT WERE USED SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6724724 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K910988, K993601. One submission often covers many models, so check the product and lot details in each record.

  • Z-0390-2010 · Sep 4, 2009 · Terminated
    Stryker Howmedica Osteonics, Skirted 5 40' Taper V40 Femoral Head (26 mm +8mm offset), hip prosthesis component.
  • Z-0391-2010 · Sep 4, 2009 · Terminated
    Stryker Howmedica Osteonics, Skirtless 5 40' LFIT V40 Femoral Head (32 mm +0 mm offset); hip prosthesis component.
  • Z-1822-2009 · Jun 18, 2009 · Terminated
    V 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended f

Other Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page