C-Taper Head 06-3205Recalls & reported problems
Howmedica Osteonics Corp. · Low Friction Ion Treatment · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327016321), received between Oct 4, 2018 and Nov 1, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Degraded” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K910988, K993601); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded2
- Adverse Event Without Identified Device or Use Problem1
- Corroded1
- Detachment of Device or Device Component1
- Device Appears to Trigger Rejection1
- Device Dislodged or Dislocated1
Patient problems
- Cancer1
- Foreign Body Reaction1
- Inflammation1
- Injury1
- Insufficient Information1
- Joint Dislocation1
- Metal Related Pathology1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO ELEVATED COBALT AND CHROMIUM LEVELS. A METAL HEAD WAS REVISED TO A CERAMIC HEAD. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Metal Related Pathology
MAUDE report 18045626 ↗InjuryJul 19, 2022Corroded; Degraded; Corroded; Degraded
IT WAS REPORTED THAT THE PATIENT UNDERWENT A PRIMARY TOTAL HIP ARTHROPLASTY WITH SUREFIT STEM AND 32 MM +5 HEAD AND TRIDENT CUP. THE PATIENT LATER PRESENTED WITH A PAINFUL RIGHT HIP AND REQUIRED A REVISION DUE TO POSSIBLE TRUNNIONOSIS AND POSSIBLE PSEUDOTUMOR. INITIAL REPORTER CONFIRMED TRUNNIONOSIS WAS PRESENT & PSEUDOTUMOR WAS SENT TO PATHOLOGY FOR ANALYSIS.
Patient problems: Cancer; Foreign Body Reaction; Pain
MAUDE report 15050781 ↗InjuryJan 31, 2022Degraded; Detachment of Device or Device Component; Degraded; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO TRUNNIONOSIS, GROSS TRUNNION FAILURE, AND DISSOCIATION OF THE HEAD FROM THE STEM. THE STEM, HEAD, AND COMPETITOR LINER WERE REVISED TO COMPETITOR IMPLANTS.
Patient problems: Insufficient Information
MAUDE report 13392142 ↗InjuryOct 4, 2018Device Appears to Trigger Rejection; Device Dislodged or Dislocated; Device Appears to Trigger Rejection; Device Dislodged or Dislocated
AS REPORTED: "PATIENT DISLOCATED HIS PRIMARY IMPLANT DUE TO ASEPTIC LOOSENING CAUSED BY METALLOSIS CREATED FROM THE TRUNNION AND THE HEADBALL. THE HEADBALL EVENTUALLY DISASSOCIATED FROM THE STEM IN VIVO. PATIENT REQUIRED REVISION SURGERY".
Patient problems: Inflammation; Injury; Joint Dislocation; Reaction
MAUDE report 7934370 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K910988, K993601. One submission often covers many models, so check the product and lot details in each record.
- Z-0390-2010 · Sep 4, 2009 · TerminatedStryker Howmedica Osteonics, Skirted 5 40' Taper V40 Femoral Head (26 mm +8mm offset), hip prosthesis component.
- Z-0391-2010 · Sep 4, 2009 · TerminatedStryker Howmedica Osteonics, Skirtless 5 40' LFIT V40 Femoral Head (32 mm +0 mm offset); hip prosthesis component.
- Z-1822-2009 · Jun 18, 2009 · TerminatedV 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended f
Other Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices
- LFIT 6260-9-136Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented166 reports
- Trident 623-00-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented112 reports
- Accolade 6021-0435Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented88 reports
- EXETER 0580-1-044Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall22 reports
- Endo-Model SL Knee Prosthesis System 16-2840/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall17 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall12 reports
- Trident 623-10-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented57 reports
- LFIT 6260-9-336Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page