Skip to content

Trident 663-10-36IRecalls & reported problems

Howmedica Osteonics Corp. · 10 Degree Eccentric Insert · Class 2 · Implant

All models of Trident (160)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327024944), received between Sep 28, 2018 and Sep 28, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Injury1
  • Joint Dislocation1
  • Loss of Range of Motion1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED. AN OPERATIVE REPORT SUPPLIED BY THE REP INDICATES "SEVERE PAIN DUE TO IMPINGEMENT WITH A RETROVERTED ACETABULUM...THE HIP IS EXTREMELY UNSTABLE WITH COMBINED ANTEVERSION OF 0...". PATIENT WAS REVISED TO AN ADM/MDM LINER CONSTRUCT AND A COMPETITOR HEAD (STRYKER SHELL REMAINED IMPLANTED BUT 1 SCREW WAS REMOVED). UPDATE: 07 SEP 2018: A REVIEW BY A CLINICAL CONSULTANT HAS CONFIRMED CUP MALPOSITION IN RETROVERSION AND EXCESSIVE MEDIALIZATION IN COMBINATION WITH LOSS OF HIP ABDUCTOR FUNCTIONALITY AFTER TROCHANTERIC AVULSION WITH CAPSULAR LAXITY AFTER MULTIPLE DISLOCATIONS HAS CONTRIBUTED TO HIP INSTABILITY WITH PAIN DUE TO IMPINGEMENT REQUIRING REVISION.

    Patient problems: Injury; Joint Dislocation; Loss of Range of Motion; Pain

    MAUDE report 7918713 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page