Triathlon 5512-F-502Recalls & reported problems
Howmedica Osteonics Corp. · Total Stabilizer Femoral Component · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327026634), received between Sep 10, 2019 and Jan 16, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Loss of Osseointegration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration2
- Osseointegration Problem2
- Unstable1
Patient problems
- Inadequate Osseointegration4
- Injury3
- Pain3
- Joint Laxity1
- Periprosthetic fracture1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJan 16, 2026Unstable
UPDATE: UPDATE ON (b)(6)2026 BASED ON MEDICAL REVIEW: ""[...]" TIBIA WITH LOOSENING SEEN AROUND THE TIBIAL STEM. IMPLANT HAS FALLEN INTO VARUS, AND THE MEDIAL PLATEAU IS COLLAPSED. THE TIBIAL STEM IS IMPINGING ON THE LATERAL CORTEX. FEMORAL COMPONENT NOT CLEARLY LOOSE BUT IN SIGNIFICANT EXTENSION ON THE LATERAL VIEW. "[...]"
Patient problems: Inadequate Osseointegration; Joint Laxity; Pain; Periprosthetic fracture
MAUDE report 24097107 ↗InjuryJun 16, 2022Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED. AS REPORTED BY REP: "PT. PRESENTED WITH THIGH PAIN, IMAGING WAS INDICATIVE OF A LOOSE FEMORAL IMPLANT FOLLOWING A KNEE-REVISION "BACK IN (B)(6) OF 2017". SURGEON RULED OUT INFECTION AND DETERMINED THE FEMORAL CONSTRUCT WAS ASEPTICALLY LOOSE. HE PERFORMED A REVISION OF THE FEMORAL IMPLANTS, LEAVING THE TIBIAL BASEPLATE AND STEM IN-SITU. NO FURTHER INFORMATION AVAILABLE." SPOKE TO REP. NO WEAR OR DAMAGE WAS NOTED INTRA-OPERATIVELY.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 14723096 ↗InjuryJun 15, 2020Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO LOOSENING. REVISING SURGEON REPORTED THAT CAUSE OR CONTRIBUTOR TO LOOSENING WAS DUE TO THE PRIOR SURGEON NOT USING ENOUGH CEMENT. THE PATIENT'S ENTIRE KNEE CONSTRUCT WAS REVISED. REP PROVIDED EXPLANT PICTURES AND REPORTED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Injury
MAUDE report 10152174 ↗InjurySep 24, 2019Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO PROGRESSIVE RADIOLUCENCY AROUND THE TIBIAL AND FEMORAL STEMS. A KNEE CONSTRUCT WITH TIBIAL BASEPLATE AND STEM, FEMORAL COMPONENT AND STEM, AND A CS INSERT WERE REVISED TO A TS CONSTRUCT. REP PROVIDED A WEBOPS REPORT, AN EXCERPT FROM AN OPERATIVE REPORT, EXPLANT PICTURES AND X-RAYS FROM (B)(6) 2019 AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 9109427 ↗InjurySep 10, 2019Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO PAIN AND LOSS OF FUNCTION. INTRA-OPERATIVELY, THE FOLLOWING WAS NOTED: THE SIZE 5 CEMENTED FEMORAL COMPONENT WAS LOOSE; THERE WAS A FRACTURE AT THE JUNCTION OF THE FEMORAL STEM AND 12X100 STEM; THE CEMENTED TIBIAL COMPONENT WAS LOOSE; THE JUNCTION OF THE STEM EXTENDER AND 12X50 STEM ON THE TIBIAL SIDE WAS LOOSE. THE PATIENT'S ENTIRE TS KNEE CONSTRUCT (INCLUDING FIVE AUGMENTS) WAS REVISED TO A TRIATHLON TS KNEE CONSTRUCT WITH A CONE. REP PROVIDED ONE PAGE OF THE PRIMARY OPERATIVE REPORT, AND REPORTED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Inadequate Osseointegration; Injury; Pain
MAUDE report 8981540 ↗
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- ATTUNE 1512-16-080Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.