Skip to content

Triathlon 5565-S-012Recalls & reported problems

Howmedica Osteonics Corp. · Fluted Stem · Class 2 · Implant

All models of Triathlon (608)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327026771), received between Aug 14, 2019 and Jun 29, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration2
  • Degraded1
  • Detachment of Device or Device Component1
  • Osseointegration Problem1

Patient problems

  • Inadequate Osseointegration2
  • Ambulation or Postural Difficulties1
  • Injury1
  • Metal Related Pathology1
  • Osteolysis1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 29, 2026Degraded; Loss of Osseointegration; Detachment of Device or Device Component

    THE FOLLOWING INFORMATION WAS PROVIDED: THE CUSTOMER REPORTED THAT: "FRACTURE (B)(6) OF THE TIBIAL AND FEMORAL STEMS LESS THAN TWO YEARS AFTER PROSTHESIS PLACEMENT. DURING REVISION SURGERY, FRACTURE OF THE IMPLANTS AND VERY SIGNIFICANT METALLOSIS (E1618) WITHIN THE KNEE WERE OBSERVED. IMPLANTATION DATE: (B)(6), 2024; DATE OF EXPLANTATION: (B)(6) 2026. IMPLANTED MEDICAL DEVICES: - 1 PG DEMI CALE TRIATHLON TS T2 5MM DM/GL (STRYKER FRANCE): REF. 5545-A-202, LOT NO. ERJVK2A, EXPIRY DATE 07/12/2026 - 1 PG REVIS EMBASE TIBIA TRIATHLON TS T2 (STRYKER FRANCE): REF. 5521-B-200, BATCH NO. O4G9RB, EXPIRY DATE 11/19/2028 - 1 PG QUILLE ANCR TRIATHLON TS 12X100MM (STRYKER FRANCE): REF. 5565-S-012, BATCH NO. 0131966D, EXPIRY DATE 12/20/2027 - 1 PG QUILLE ANCR TRIATHLON TS 15X100MM (STRYKER FRANCE): REF. 5565-S-015, BATCH NO. 0086223A, EXPIRY DATE 03/24/2024 - 1 PG REVIS FEMOR 2COND CIM TRIATHLON TS DTE T2 STEM (STRYKER FRANCE): REF. 5512F202, BATCH NO. L7T4U, EXPIRY DATE 07/19/2028 1 PG CALE FEMOR DIST TRIATHLON TS DTE T2 10MM (STRYKER FRANCE): REF. 5541-A-202, BATCH NO. EYS9G, EXPIRY DATE 09/23/2024 - 1 PG CALE FEMOR DIST TRIATHLON TS DTE T2 15MM (STRYKER FRANCE): REF. 5542-A-202, BATCH NO. EYR9

    Patient problems: Ambulation or Postural Difficulties; Inadequate Osseointegration; Metal Related Pathology; Osteolysis

    MAUDE report 25651634 ↗
  • InjuryAug 14, 2019Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    REVISION LEFT TOTAL KNEE REPLACEMENT. REMOVED TRIATHLON UNIVERSAL BASEPLATE & FLUTED STEM & PS INSERT FOR LOOSENING. REVISED WITH TRIATHLON UNIVERSAL BASEPLATE, CEMENTED TIBIAL STEM & TS INSERT. PREVIOUS REVISION SURGERY CONDUCTED AT (B)(6) IN (B)(6) 2014.

    Patient problems: Inadequate Osseointegration; Injury

    MAUDE report 8892301 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page