Skip to content

Triathlon 5537-G-619Recalls & reported problems

Howmedica Osteonics Corp. · Total Stabilizer+ Tibial Insert · Class 2 · Implant

All models of Triathlon (608)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327027426), received between Feb 16, 2018 and Jan 29, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Operates Differently Than Expected1
  • Fracture1
  • Insufficient Device Problem Information1

Patient problems

  • Injury2
  • Unspecified Infection2
  • No Known Impact Or Consequence To Patient1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS PI IS FOR REVISION OF THE PATIENT'S RIGHT KNEE ON (B)(6) 2024. AS REPORTED: "PATIENT UNDERWENT TWO POLY EXCHANGES. ONE WAS LAST MONDAY ONE WAS TODAY. POLY EXCHANGES WERE DUE TO INFECTION. X-RAYS ATTACHED. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY."

    Patient problems: Unspecified Infection

    MAUDE report 18594303 ↗
  • InjuryNov 13, 2019Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO PAIN. INTRA-OPERATIVELY, THE POLYETHYLENE COVERING THE POST OF THE INSERT WAS SHEARED, WORN, OR FRACTURED OFF THE METAL POST OF THE INSERT. SOME WEAR WAS NOTED WHERE THE INSERT'S METAL POST ARTICULATED AGAINST THE FEMORAL COMPONENT. THE PATIENT'S ENTIRE KNEE CONSTRUCT (SIZE 6 TS FEMUR WITH 2 AUGMENTS AND STEM, 6X19 TS INSERT, SIZE 6 UNIVERSAL BASEPLATE WITH STEM) WAS REVISED TO A TS KNEE. REP REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: Injury; Pain

    MAUDE report 9314676 ↗
  • InjuryJun 25, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    AS REPORTED: "RIGHT KNEE REVISION DUE TO SEPTIC KNEE". REP PROVIDED PRE-REVISION X-RAYS AND REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 8731666 ↗
  • MalfunctionFeb 16, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    SURGEON REPORTED THAT THE TRIATHLON TS 19 MM INSERT WOULDN¿T SIT CORRECTLY IN UNIVERSAL BASE PLATE. HE FURTHER REPORTED THAT HE THEN INSERTED THE TS 16 MM INSERT INTO THE UNIVERSAL BASE TIBIAL BASE PLATE, AGAIN THIS DIDN¿T SEAT PROPERLY, HE IMPACTED THE LINER AGAIN AND INSERTED THE POST AND SAID THAT HE WOULD BRING PATIENT BACK TO THEATRE NEXT DAY FOR A LINER EXCHANGE. THERE WAS A 10 MINUTE DELAY TO SURGERY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7277902 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates