Dall-Miles 6704-8-240Recalls & reported problems

Howmedica Osteonics Corp. · Cable · Class 2 · Implant

All models of Dall-Miles (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327036497), received between Feb 15, 2022 and Jan 29, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Insufficient Information2
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2025Fracture; Fracture

    REVISION PROCEDURE OF PROXIMAL FEMUR GMRS. MULTIPLE PREVIOUS REVISIONS; CABLES IMPLANTED DURING PROCEDURE (B)(6) 2023. NO DETAILS AVAILABLE OF PROCEDURES PRIOR TO THIS DATE. DALL MILES CABLE HAD SNAPPED AND FRAYED IN THE PATIENT. CABLES AND CABLE GRIP REMOVED AND FRAYED METAL REMOVED UNDER X RAY GUIDANCE. NO NEW IMPLANTS WERE PUT IN. ALL OF THE ITEMS ARE STILL IN PATIENT EXCEPT FOR THE CABLES AND CABLE GRIP WHICH WERE REMOVED DURING THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21263920 ↗
  • InjuryJan 29, 2025Fracture; Fracture

    REVISION PROCEDURE OF PROXIMAL FEMUR GMRS. MULTIPLE PREVIOUS REVISIONS: CABLES IMPLANTED DURING PROCEDURE (B)(6) 2023. NO DETAILS AVAILABLE OF PROCEDURES PRIOR TO THIS DATE. DALL MILES CABLE HAD SNAPPED AND FRAYED IN THE PATIENT. CABLES AND CABLE GRIP REMOVED AND FRAYED METAL REMOVED UNDER X RAY GUIDANCE. NO NEW IMPLANTS WERE PUT IN. ALL OF THE ITEMS ARE STILL IN PATIENT EXCEPT FOR THE CABLES AND CABLE GRIP WHICH WERE REMOVED DURING THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21263918 ↗
  • InjuryFeb 15, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    PATIENT STATED SHE HAD A LEFT THA ON (B)(6) 2016. TWO HOURS AFTER HER SURGERY THE PATIENT WAS BROUGHT BACK TO THE OPERATING ROOM AND WAS TOLD THE PIECE FAILED AND WAS REVISED. PATIENT STATED SHE WAS EXPERIENCING PAIN AND HAVE SEEN HER SURGEON BUT WAS TOLD THAT EVERYTHING WAS FINE. THE PATIENT SOUGHT A SECOND OPINION ON (B)(6) 2018 AND WAS REVISED DUE TO HER X-RAY FINDINGS. PATIENT CONTINUED TO EXPERIENCE PAIN AND WAS REVISED ON (B)(6)2019. PATIENT IS STILL EXPERIENCING EXCRUCIATING PAIN. PATIENT HAS BEEN IN AND OUT OF THE HOSPITAL FOR THE PAST 5 YEARS AND IS SEEKING COMPENSATION FOR ALL HER SUFFERING. THIS PI IS FOR REVISION1. UPDATE ON 21-JAN-2022 BASED ON MED REVIEW: AN OPERATION NOTE ON (B)(6) 2018. THE POSTOPERATIVE DIAGNOSIS WAS "PAINFUL HARDWARE LEFT HIP, REMOVAL OF HARDWARE LEFT HIP, WE ARE REMOVING A CABLE OFF THE FEMUR". NO SIGNS OF INFECTION WERE NOTED A NORMAL AMOUNT OF JOINT FLUID WAS SEEN WITH NO SIGNS OF PURULENCE. FROZEN SECTION CAME BACK NOT CONSISTENT WITH INFECTION. THE CABLE WAS LOOSE AND REMOVED. A REPAIR OF THE ABDUCTOR MECHANISM WAS CARRIED OUT.

    Patient problems: Pain

    MAUDE report 13539225 ↗

Other CERCLAGE, FIXATION devices

All CERCLAGE, FIXATION devices →

Sources and updates