Trident 623-00-36HRecalls & reported problems
Howmedica Osteonics Corp. · 0 Degree Polyethylene Insert · Class 2 · Implant
All models of Trident (160)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327039559), received between Oct 19, 2018 and Apr 29, 2022 — classified as 6 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Adverse Event Without Identified Device or Use Problem1
- Degraded1
- Insufficient Device Problem Information1
- Material Deformation1
- Noise, Audible1
Patient problems
- Injury5
- Joint Dislocation3
- Ambulation or Postural Difficulties1
- Insufficient Information1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryApr 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IN PROVIDING INFORMATION FOR A REVISION OF THE PATIENT'S LEFT HIP ON (B)(6) 2022, REP PROVIDED USAGE SHEETS FOR (B)(6) 2020. COMPARING THEM, A TRIDENT LINER WAS REVISED TO AN MDM METAL LINER AND ADM/ MDM POLY INSERT, 4 SCREWS ADDED TO THE IN SITU SHELL, AND A RESTORATION MODULAR STEM CONSTRUCT WITH A CERAMIC FEMORAL HEAD WERE IMPLANTED. THIS REPORT IS FOR THE (B)(6) 2020 REVISION.
Patient problems: Insufficient Information
MAUDE report 14250498 ↗InjuryAug 7, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. A 36 MM 'H' TRIDENT X3 0° POLY INSERT, V-40/C-TAPER ADAPTER SLEEVE, AND 36 +5 MM BIOLOX DELTA CERAMIC C-TAPER HEAD WERE REVISED TO A 44 MM 'H' TRIDENT X3 0° POLY INSERT, +4 MM UNIVERSAL V40 TAPER ADAPTER SLEEVE, AND A 44 MM BIOLOX DELTA ANATOMIC UNIVERSAL TAPER HEAD. "PATIENT HAS A CHRONIC INFECTION. LOW-ACTIVITY LEVEL, SEVERE BONE LOSS, COMPONENTS WERE EXCHANGED TO BEST TREAT THE INFECTION."
Patient problems: Injury; Unspecified Infection
MAUDE report 10378312 ↗InjuryJun 16, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2014: RIGHT HIP REPLACEMENT, ON (B)(6) 2014: DISLOCATION RIGHT HIP, ON (B)(6) 2019: DISLOCATION RIGHT HIP, ON (B)(6) 2020: DISLOCATION RIGHT HIP; REVISION SURGERY IS PLANNED.
Patient problems: Injury; Joint Dislocation
MAUDE report 10157130 ↗InjuryNov 27, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2014 : TOTAL HIP REPLACEMENT IN HS038 REGISTRY, GOOD OUTCOME.(B)(6) 2019 : DISLOCATION OF THE HIP, REPOSITIONED THE SAME DAY. DEVICE NOT EXPLANTED.
Patient problems: Ambulation or Postural Difficulties; Injury; Joint Dislocation
MAUDE report 9385510 ↗InjuryNov 8, 2019Degraded; Noise, Audible; Degraded; Noise, Audible
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. PRE-REVISION PATIENT COMPLAINT INCLUDED HEARING A POPPING SOUND. INTRA-OPERATIVELY, THE FOLLOWING WERE NOTED: TRUNNION WEAR, TRUNNION BROKE OFF THE STEM, GRAYISH BLACK TISSUE, WEAR OF THE LINER'S RIM DUE TO IMPINGEMENT OF THE STEM ON THE LINER. THE SHELL HAD NO ALLEGATIONS AGAINST IT AND WAS NOT REVISED. PATIENT WAS REVISED TO A RESTORATION MODULAR STEM CONSTRUCT WITH A CERAMIC HEAD AND NEW LINER. REP PROVIDED INTRA- OP PHOTOS AND AN EXPLANT PICTURE AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury
MAUDE report 9295776 ↗InjuryOct 19, 2018Material Deformation; Material Deformation
PATIENT CALLED STATING HE HAD A LEFT HIP IMPLANT ON (B)(6) 2005. PATIENT NOTED THAT 2 MONTHS AGO AS HE GOT UP FROM USING THE BATHROOM HE NOTICED PAIN. PATIENT STATED HIS PELVIS AND FEMUR DISCONNECTED. PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2018 TO STRYKER DEVICES.
Patient problems: Injury; Joint Dislocation; Pain
MAUDE report 7984813 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, uncemented devices
- LFIT 6260-9-136Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented166 reports
- Trident 623-00-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented112 reports
- Accolade 6021-0435Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented88 reports
- EXETER 0580-1-044Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall22 reports
- Endo-Model SL Knee Prosthesis System 16-2840/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall17 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall12 reports
- Trident 623-10-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented57 reports
- LFIT 6260-9-336Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page