Triathlon 5517-F-601Recalls & reported problems
Howmedica Osteonics Corp. · Cruciate Retaining Femoral · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07613327040975), received between Jul 27, 2017 and Mar 19, 2025 — classified as 11 injuries, 1 malfunction.
The most frequently reported specific device problem is “Unstable” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable4
- Loss of Osseointegration3
- Insufficient Device Problem Information2
- Device Operates Differently Than Expected1
- Fitting Problem1
- Fracture1
- Mechanical Problem1
- Osseointegration Problem1
Patient problems
- Pain6
- Inadequate Osseointegration4
- Injury4
- Joint Laxity2
- Failure of Implant1
- Fibrosis1
- Loss of Range of Motion1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMar 19, 2025Mechanical Problem; Mechanical Problem
REVISED A LEFT TOTAL KNEE ORIGINALLY DONE (B)(6) 2024 DUE TO PAIN/LIMITED RANGE OF MOTION. DR. HAD REVISED THE PATIENT ON (B)(6) 2024 FROM AN MCK MEDIAL UNI DONE ORIGINALLY ON (B)(6) 2023 (UNDER ANOTHER PI). WHEN THE FEMUR WAS REMOVED, DR. OBSERVED THERE WAS MOSTLY FIBROUS INGROWTH.
Patient problems: Fibrosis; Loss of Range of Motion; Pain
MAUDE report 21642869 ↗InjuryNov 21, 2023Loss of Osseointegration; Loss of Osseointegration
PATIENT WOULD LIKE TO KNOW IF KNEE IMPLANTS ARE SUBJECT TO A RECALL. PATIENT NOTED LOOSENING. UPDATE PER CONVERSATION WITH PATIENT: PATIENT REPORTED AN ASPIRATION.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 18187244 ↗InjuryJan 24, 2022Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO PAIN AND FEMORAL LOOSENING. SURGEON REPORTED THAT A POSSIBLE CAUSE OR CONTRIBUTOR IS THAT THE ORIGINAL CUTS MADE WITH THE MAKO ROBOT SEEMED OFF. A CR FEMUR AND 7X9 CS INSERT WERE REVISED TO A TS FEMUR WITH STEM AND 7X9 PS INSERT. REP REPORTED THAT PICTURES CAN BE PROVIDED, AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 13333172 ↗InjuryAug 11, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED: "DR. REVISED A LEFT KNEE DUE TO PATELLA INSTABILITY. THE TIBIA BASEPLATE WAS NOT REVISED. ALL OTHER COMPONENTS CAME OUT." A TRIATHLON CR FEMORAL COMPONENT, TRIATHLON CR X3 INSERT, AND TRITANIUM PATELLA WERE REVISED TO A TS FEMUR, PS INSERT, AND PATELLA WITH A STEM, CONES, AND AUGMENTS. PER REP: THERE ARE NO ALLEGATIONS AGAINST THE REVISED LINER OR THE ROBOT USED TO COMPLETE THE PRIMARY SURGERY.
Patient problems: Joint Laxity
MAUDE report 12310041 ↗InjuryMar 31, 2021Fracture; Unstable; Fracture; Unstable
THE POSTERIOR MEDIAL ASPECT OF OUR CEMENTLESS TITANIUM KNEE REPLACEMENT BROKE OFF INSIDE THE PATIENT. THE ANTERIOR LOCKING WIRE ALSO FRAGMENTED INTO 3 PIECES, BUT THE POLY WAS STILL LOCKED AND ENGAGED INTO THE BROKEN TIBIA, AND NOT FRACTURED. PATIENT CAME IN WITH SWELLING, PAIN AND INSTABILITY, NOTICED ON XRAY, THEN CONFIRMED AND REVISED IN SURGERY. UPDATE: "THE REVISION WAS PERFORMED BECAUSE OF THE FRACTURE OF THE TIBIAL PLATE".
Patient problems: Failure of Implant; Pain
MAUDE report 11594862 ↗InjuryFeb 4, 2021Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO PAIN AND INSTABILITY. A CR FEMUR AND CS INSERT WERE REVISED TO A TS INSERT WITH AUGMENTS AND STEM, AND A PS INSERT. REP PROVIDED THE REVISION USAGE SHEET AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Pain
MAUDE report 11282982 ↗InjuryDec 8, 2020Unstable; Loss of Osseointegration; Unstable; Loss of Osseointegration
TRIATHLON TRITANIUM IMPLANTED AT INITIAL SURGERY. INSTABILITY LED TO FEMORAL AND INSERT EXCHANGE.
Patient problems: Inadequate Osseointegration; Joint Laxity
MAUDE report 10964782 ↗InjuryAug 6, 2019Patient-Device Incompatibility; Patient-Device Incompatibility
AS REPORTED: "...DURING THE LEFT TOTAL KNEE REPLACEMENT A CRUCIATE RETAINING FEMUR WAS MISTAKENLY OPENED AND IMPLANTED INSTEAD OF A POSTERIOR STABILIZED FEMUR. THIS MISTAKE WENT UNNOTICED UNTIL THE PATIENT WAS IN POST-OP RECOVERY. THE PATIENT WAS THEN RETURNED TO THE OPERATING ROOM AND THE CRUCIATE RETAINING FEMUR WAS REMOVED AND THE CORRECT POSTERIOR STABILIZED FEMUR WAS IMPLANTED".
Patient problems: Injury
MAUDE report 8864277 ↗InjuryJun 11, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED: "MAKO MPS OPENED A CR FEMUR AND IT WAS IMPLANTED ON A PS PREPPED FEMUR WITH A PS INSERT. I NOTICED WHEN I WAS LOOKING AT PAPERWORK THE NEXT DAY AND WENT STRAIGHT TO DOCTOR'S CLINIC TO NOTIFY HIM. REVISION WAS DONE TWO DAYS POST ORIGINAL SURGERY". UPDATE (B)(6) 2019 WG: AS REPORTED BY THE H.R. BUSINESS PARTNER: SPOKE TO MPS. MPS REPORTED HE PULLED THE IMPLANT BOX FROM THE CART AND HANDED IT TO THE SURGICAL STAFF WITHOUT OPENING IT.
Patient problems: Injury
MAUDE report 8687275 ↗InjuryMar 25, 2019Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "LOOSE LEFT TOTAL KNEE REPLACEMENT."
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8449151 ↗InjuryJul 27, 2018Unstable; Unstable
IT WAS REPORTED THAT PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY AND PAIN.
Patient problems: Injury; Pain
MAUDE report 7726549 ↗MalfunctionJul 27, 2017Fitting Problem; Device Operates Differently Than Expected; Scratched Material; Fitting Problem; Device Operates Differently Than Expected; Scratched Material
IT WAS REPORTED THAT A PRESS FIT LEFT FEMUR HAD AN IMPROPER FIT. SURGEON RE-CUT AND STILL COULD NOT GET IT SEATED. SURGEON IMPLANTED A CEMENTED FEMUR INSTEAD. THERE WAS A 45 MINUTE SURGICAL DELAY BUT NO ADVERSE CONSEQUENCE TO PATIENT. REP STATED NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6747051 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page