Triathlon 5517-F-702Recalls & reported problems
Howmedica Osteonics Corp. · Cruciate Retaining Femoral · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327041019), received between Oct 9, 2018 and Feb 17, 2026 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Unstable” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable2
- Device Appears to Trigger Rejection1
- Device Dislodged or Dislocated1
- Fitting Problem1
- Inadequacy of Device Shape and/or Size1
- Loss of Osseointegration1
- Malposition of Device1
- Migration1
Patient problems
- Injury3
- Inadequate Osseointegration2
- Ambulation or Postural Difficulties1
- Discomfort1
- Fall1
- Joint Disorder1
- Joint Laxity1
- Loss of Range of Motion1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryFeb 17, 2026Loss of Osseointegration
THE FOLLOWING INFORMATION WAS PROVIDED: SURGEON DID REVISION OF RIGHT STRYKER PRESSFIT CR FEMUR AND BASEPLATE AFTER 1 YEAR OUT OF SURGERY - SURGEON DID NOT PERFORM THE PRIMARY SURGERY JUST THE REVISION AND WANTS TO SEND BACK FEMORAL COMPONENT. HE BELIEVES NOT ENOUGH BONE GROWTH WAS ON THE FEMUR. AS REPORTED BY HOSPITAL MANAGEMENT: I INCLUDED THE TIBIA SO YOU CAN SEE THERE IS BONE GROWTH. THE FEMUR HAS SOME TOO. THE PATIENT IS A YEAR OR SO OUT. THE SURGEON FELT IT DID NOT HAVE ENOUGH BONE GROWTH ON THE FEMUR.
Patient problems: Ambulation or Postural Difficulties; Inadequate Osseointegration
MAUDE report 24363026 ↗InjuryFeb 7, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS PI IS FOR THE SOFT TISSUE REALIGNMENT PROCEDURE. IN REPORTING A REVISION OF THE PATIENT'S RIGHT KNEE ON (B)(6) 2022, REP REPORTED THAT IN (B)(6) OF 2021 THE PATIENT FELL AND DISLOCATED THEIR RESURFACED PATELLA. A SOFT TISSUE REALIGNMENT WAS PERFORMED. REP CONFIRMED THAT NO STRYKER REP WAS PRESENT FOR THE PROCEDURE AS THE SURGEON DID NOT GO INTO THE PATIENT'S JOINT SPACE OR REVISE ANY IMPLANTS. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Fall
MAUDE report 13471335 ↗InjuryJan 4, 2022Unstable; Malposition of Device; Unstable; Malposition of Device
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO MALPOSITION AND INSTABILITY. INTRA-OPERATIVELY THE FOLLOWING WAS NOTED: SLIGHT MALPOSITION OF FEMORAL COMPONENT, WORN INSERT WITH BROKEN LOCKING WIRE, AND BROKEN TIBIAL BASEPLATE. THE BASEPLATE WAS BROKEN MEDIALLY 1/3 OF THE WAY THROUGH THE ANTERIOR MEDIAL SPIKE. THE PATIENT WAS REVISED TO A TS KNEE WITH A UNIVERSAL BASEPLATE AND STEMS.
Patient problems: Joint Laxity
MAUDE report 13150845 ↗MalfunctionMar 2, 2020Fitting Problem; Fitting Problem
FEMUR CUT WITH MAKO, TRIAL SEEMED TO FIT WELL, PLACED PA BEADED FEMUR SIZE 7 RIGHT IMPLANT ON AND DID NOT SEEM TO FIT PROPERLY. OFF ON ANTERIOR CHAMFER AND ANTERIOR CUT APPROX 2-3MM. HAD TO OPEN ANOTHER IMPLANT AND. CEMENT FEMUR. HAD TO CEMENT FEMUR MEANT CASE TOOK APPROX 15MIN LONGER. CASE TYPE: TKA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9775744 ↗InjuryDec 30, 2019Migration; Migration
KNEE REVISION POLY WEAR. COMMENT IN ADVERSE CONSEQUENCES DETAILS: PATIENT NEEDED KNEE REVISION DUE TO INSTABILITY. *UPDATE: BASED ON MEDICAL REVIEW, X-RAYS DATED (B)(6) 2019 ARE AN AP, FOUR LATERALS, AND ONE PATELLAR VIEW OF THE RIGHT KNEE, WHICH IS ESSENTIALLY UNCHANGED FROM THE PREVIOUS STUDY, BUT THE POSTERIOR SUBLUXATION OF THE FEMORAL COMPONENT IS NOTED ON THE LATERAL X-RAYS.
Patient problems: Injury; Joint Disorder
MAUDE report 9530775 ↗InjuryJul 3, 2019Inadequacy of Device Shape and/or Size; Unstable; Osseointegration Problem; Inadequacy of Device Shape and/or Size; Unstable; Osseointegration Problem
"MY KNEE SPACER IS SLIGHTLY LOOSE AND MY PHYSICIAN, A MAN I LIKE AND TRUST AS A QUALITY SURGEON, TOLD ME THAT THE SPACERS COME IN TWO MILLIMETER INCREMENTS AND HE GAVE ME A SIZE CLOSE AS POSSIBLE GIVEN MY ANATOMY BUT TO REPEAT, MINE IS SLIGHTLY LOOSE AND SLIPS LATERALLY, CREATING A LACK OF ABILITY TO EVEN STAND NORMALLY ON MY LEG, WHICH CAUSES ME TO FAVOR THE LEG UNNATURALLY, GIVING ME CHRONIC LOW BACK AND HIP DISCOMFORT, WHICH IS NOT HORRIBLE, BUT IT IS NAGGING." UPDATE 10-MAY-2018: AS PER OPT REPORT, RIGHT KNEE REPLACEMENT WAS DONE ON (B)(6) 2018. PATIENT HAD A MANIPULATION DONE ON (B)(6) 2018. UPDATE 11 JUNE 2018 (B)(6): AS PER MEDICAL REVIEW, PATIENT HAD A MANIPULATION DONE ON (B)(6) 2018 BECAUSE OF LIMITED KNEE FLEXION TO 90° WITH FULL EXTENSION. [...] SURGICAL ERROR IN CHOICE OF FEMORAL COMPONENT SIZE BETWEEN INTENDED AND BIGGER COMPONENT ACTUALLY IMPLANTED HAS CONTRIBUTED TO A STIFFNESS PROBLEM IN THE KNEE IN AN ARTHROPLASTY THAT IN GENERAL IS ALREADY AT RISK FOR SCAR TISSUE FORMATION WITH RESULTING ARTHROFIBROSIS. MANIPULATION OF THE KNEE UNDER ANAESTHESIA IMPROVED KNEE FUNCTIONALITY TO NORMAL ALTHOUGH THERE IS NO RECENT INFORMATION AVAILABLE ABOUT RETENTION OF ADEQUATE KNE
Patient problems: Discomfort; Inadequate Osseointegration; Injury; Loss of Range of Motion; Pain
MAUDE report 8758394 ↗InjuryOct 9, 2018Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
SURGEON REPORTED THAT PATIENT DEVELOPED RECURRENT EFFUSIONS FOLLOWING TOTAL KNEE REPLACEMENT WITH STRYKER TRIATHLON CEMENTLESS COMPONENTS. DUE TO MULTIPLE PATIENT ALLERGIES, PATIENT WAS THEN TESTED FOR METAL ALLERGIES, WHICH CAME BACK POSITIVE FOR ALLERGY TO NICKEL. REVISION WAS SCHEDULED AND PERFORMED ON (B)(6) 2018, REMOVING FEMUR, POLY INSERT, AND TIBIAL COMPONENT. TRITANIUM PATELLA WAS RETAINED. REVISION WITH NICKEL FREE IMPLANTS FROM [COMPETITOR] WAS PERFORMED.
Patient problems: Injury; Reaction
MAUDE report 7948873 ↗
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- ATTUNE 1512-16-080Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page