Triathlon 5536-B-200Recalls & reported problems
Howmedica Osteonics Corp. · Tibial Component · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Howmedica Osteonics Triathlon knee implant's tibial baseplate is the part being discussed in these reports. Most describe the baseplate loosening or sinking into the bone, so the knee had to be revised or replaced, along with reports of pain, bone fractures and difficulty walking. There are 28 reports, with28 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 28
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 07613327041590), received between Mar 2, 2018 and Aug 6, 2026 — classified as 28 injuries.
The most frequently reported specific device problem is “Loss of Osseointegration” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration11
- Migration5
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Osseointegration Problem2
- Unintended Movement2
- Unstable2
- Collapse1
Patient problems
- Inadequate Osseointegration12
- Injury9
- Pain7
- Bone Fracture(s)5
- Ambulation or Postural Difficulties3
- Insufficient Information2
- Cyst(s)1
- Deformity/ Disfigurement1
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
InjuryAug 6, 2026Loss of Osseointegration
Patient problems: Inadequate Osseointegration; Loss of Range of Motion; Pain
MAUDE report 50159828 ↗InjuryFeb 25, 2026Use of Device Problem
InjurySep 3, 2025Failure to Osseointegrate; Failure to Osseointegrate
REVISED A RIGHT KNEE TRIATHLON CEMENTLESS TIBIAL BASEPLATE AND POLY DUE TO LOOSENING OF THE TIBIAL BASEPLATE COMPONENT. REVISED TO UNIVERSAL BASEPLATE.
Patient problems: Inadequate Osseointegration
MAUDE report 22952137 ↗InjuryAug 22, 2025Loss of Osseointegration; Loss of Osseointegration
LOOSENING OF THE TIBIAL BASEPLATE OF A MAKO TKA PERFORMED IN MARCH WAS CONFIRMED. ON (B)(6) 2025 EXTRACTION SURGERY WAS PERFORMED AND A CEMENT MOLD CONTAINING ANTIBIOTICS WAS PLACED THINKING ABOUT THE POSSIBILITY OF INFECTION. A REVISION TKA IS PLANNED AT A LATER DATE. IN THE SURGEON'S OPINION, THE INITIAL SURGERY WAS PERFORMED WITH A UKA SKIN INCISION, WHICH RESULTED IN POOR VISIBILITY DURING SURGERY, CAUSING THE TIBIA TO OVERHANG MEDIALLY, AND THE LATERAL SIDE OF THE TIBIA TO SINK AFTER SURGERY.
Patient problems: Inadequate Osseointegration
MAUDE report 22873754 ↗InjuryMay 13, 2025Migration; Migration
RIGHT-KNEE REVISION. PT. PRESENTED WITH INDICATORS OF "SUBSIDED" TRITANIUM BASEPLATE, FOLLOWING A MAKO-TKA. DR. SUSPECTED A MINOR FRACTURE OF THE TIBIA BENEATH THE BASEPLATE. HE REMOVED THE TRITANIUM BASEPLATE AND REVISED TO A CEMENTED UNIVERSAL BASEPLATE WITH BILATERAL AUGMENTS. FEMUR COMPONENT WAS LEFT IN-SITU. THE KNEE "FELT BALANCED, CLINICALLY" AND HE WENT BACK TO A CS-INSERT. NO COMPLAINTS FILED AGAINST THE COMPONENTS THEMSELVES, NO FURTHER INFO. AVAILABLE.
Patient problems: Ambulation or Postural Difficulties; Bone Fracture(s)
MAUDE report 22010922 ↗InjuryOct 30, 2024Loss of Osseointegration; Loss of Osseointegration
LEFT TOTAL KNEE. THE REVISION TODAY WAS SECONDARY TO TIBIAL LOOSENING. THE USE OF MAKO IN THE PRIMARY PROCEDURE WAS NOT THOUGHT TO HAVE CONTRIBUTED TO THIS FAILURE. THE PATELLA WAS NOT RESURFACED DURING THE INDEX PROCEDURE SO IT WAS RESURFACED TODAY FOR THE FIRST TIME. THE TIBIA WAS REVISED AND FEMORAL COMPONENT WAS LEFT IN PLACE. NO FURTHER INFORMATION AVAILABLE FROM THE DOCTOR OR HOSPITAL.
Patient problems: Inadequate Osseointegration
MAUDE report 20569829 ↗InjuryJul 9, 2024Loss of Osseointegration; Loss of Osseointegration
AS REPORTED: "REVISION RIGHT TOTAL KNEE DUE TO TIBIAL COMPONENT LOOSENING. COMPONENTS REMOVED TODAY INCLUDED FEMUR, TIBIAL INSERT AND TIBIAL BASEPLATE." REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Inadequate Osseointegration
MAUDE report 19700271 ↗InjuryApr 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RIGHT TOTAL KNEE. THE REVISION TODAY WAS SECONDARY TO FALL AND FRACTURE OF THE PROXIMAL TIBIA. THE USE OF MAKO ON THE PRIMARY KNEE WAS NOT THOUGHT TO HAVE ANYTHING TO DO WITH THE REVISION. NO FURTHER INFORMATION AVAILABLE FROM THE DOCTOR OR HOSPITAL.
Patient problems: Bone Fracture(s)
MAUDE report 19104246 ↗InjuryApr 2, 2024Degraded; Degraded
DR. REVISED A LEFT TKA DONE ORIGINALLY MANUALLY DUE TO POSTERIOR WEAR OF THE TIBIAL INSERT AND METAL WEAR OF THE BASEPLATE FROM FEMORAL COMPONENT CONTACT. REVISED TO A UNIVERSAL BASEPLATE AND CEMENTED CR FEMUR. NO ALLEGATIONS MADE AGAINST THE IMPLANT.
Patient problems: Insufficient Information
MAUDE report 19017670 ↗InjuryDec 11, 2023Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED. AS REPORTED: "PATIENT REVISE DUE TO METAL ALLERGY. PATELLA WAS RETAINED. USED COMPETITOR IMPLANTS AND ONE OF OUR CONES. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY."
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 18307838 ↗InjuryAug 1, 2023Migration; Migration
PATIENT HAD A PRIMARY KNEE REPLACEMENT ON (B)(6) 2023 AND PRESENTED TO HOSPITAL WITH PAIN AND SWELLING X-RAY SHOWED THE TIBIAL BASEPLATE HAD MOVED AND SUBSIDED. PATIENT HAD HER TIBIA REVISED SUCCESSFULLY.
Patient problems: Pain; Swelling/ Edema
MAUDE report 17438305 ↗InjuryMar 30, 2023Fracture; Loss of Osseointegration; Osseointegration Problem; Fracture; Loss of Osseointegration; Osseointegration Problem
REVISED A RIGHT TRIATHLON TOTAL KNEE ORIGINALLY DONE ON (B)(6) 2022 DUE TO ASEPTIC LOOSENING, POSSIBLE INFECTION. INTRA-OP HE NOTICED WHEN REMOVING THE TIBIAL BASEPLATE, THE POSTERIOR-MEDIAL PEG WAS SHEARED OFF. DR. DIDN'T BELIEVE IT HAPPENED DUE TO HIS TECHNIQUE OF EXTRACTION AS THE BASEPLATE WAS ALREADY LOOSE. HE IS NOT SURE HOW OR WHEN IT COULD HAVE HAPPENED. HE REVISED TO A TRIATHLON TS KNEE. HE DID NOT REVISE THE PATELLA COMPONENT.
Patient problems: Inadequate Osseointegration; Unspecified Infection
MAUDE report 16642403 ↗InjuryMar 27, 2023Loss of Osseointegration; Noise, Audible; Loss of Osseointegration; Noise, Audible
PATIENT REPORTED JOINT LOOSENING AND AUDIBLE NOISE FOLLOWING PRIMARY TRIATHLON LEFT KNEE SURGERY. PATIENT STATED THAT SYMPTOMS STARTED "IMMEDIATELY" FOLLOWING THE (B)(6) 2020 SURGERY AND THAT THE KNEE WAS"NEVER GOOD". PATIENT STATED SHE FEELS THE LOOSENING AND DID NOT STATE WHETHER HER SURGEON CONFIRMED THE LOOSENING. PATIENT STATED THAT THE JOINT NOISE IS HEARD WHEN SHE MOVES HER "TOES UP AND DOWN".
Patient problems: Inadequate Osseointegration
MAUDE report 16619645 ↗InjuryFeb 7, 2023Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO BASEPLATE LOOSENING. A TRIATHLON TRITANIUM BASEPLATE AND 2X9 CS INSERT WERE REVISED TO A UNIVERSAL BASEPLATE AND 2X13 CS INSERT.
Patient problems: Inadequate Osseointegration
MAUDE report 16312816 ↗InjuryJan 13, 2023Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem; Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem
PATIENT'S LEFT KNEE WAS REVISED. AS REPORTED: "PATIENT WAS HAVING PAIN. X-RAY SHOWED A MEDIAL TIBIAL CYST. FEMUR, TIBIA AND INSERT WERE REVISED."
Patient problems: Cyst(s); Pain
MAUDE report 16163635 ↗InjuryJan 11, 2023Loss of Osseointegration; Unintended Movement; Loss of Osseointegration; Unintended Movement
PATIENT'S TRITANIUM TIBIAL BASEPLATE WAS REVISED FOR SUBSIDENCE AND LOOSENING. TIBIAL WAS REPLACED WITH A CEMENTED UNIVERSAL BASEPLATE WITH 5MM MEDIAL AUGMENT AND SHORT CEMENTED STEM.
Patient problems: Inadequate Osseointegration
MAUDE report 16135693 ↗InjuryDec 21, 2022Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED. [,,,] "A PATELLA COMPONENT WHICH HAS DISPLACED FROM THE BONE AND A TIBIAL COMPONENT WHICH APPEARS SUBSIDED AND IN EXTENSION". A BASEPLATE, INSERT, AND PATELLAR COMPONENT WERE REVISED.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 16021518 ↗InjuryJun 29, 2022Fracture; Unstable; Fracture; Unstable
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED. AS REPORTED: "PAIN AND DEFORMITY." A BASEPLATE, INSERT, AND PATELLAR COMPONENT WERE REVISED.
Patient problems: Deformity/ Disfigurement; Joint Laxity; Pain
MAUDE report 14865141 ↗InjuryApr 11, 2022Migration; Migration
IT WAS REPORTED THAT THE SIZE #2 TRITANIUM BASEPLATE SUBSIDED.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 14073714 ↗InjuryJun 15, 2020Collapse; Collapse
REP LEFT A VOICEMAIL STATING THAT THE PATIENT'S TRITANIUM BASEPLATE WAS REVISED DUE TO ANTERIOR COLLAPSE. UPDATE: "NOTES: ANTERIOR COLLAPSE OF TRI BP AT 3MO POST OP FROM IDOS. PREVIOUSLY ASYMPTOMATIC FOR PAIN/INSTABILITY. EXPLANTS UNAVAILABLE FOR RETURN, PER HOSPITAL POLICY."
Patient problems: Injury
MAUDE report 10152699 ↗InjuryMay 19, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT HAD 2 WEEK POST OP EXAM WITH X-RAYS. THE X-RAY LOOKED NORMAL. 2 MONTHS LATER KNEE STARTED HURTING AND PATIENT NOTICED HER KNEE DID NOT LOOK STRAIGHT. PATIENT CAME IN FOR MORE X-RAYS, TIBIAL PLATEAU HAD FRACTURED AND WAS IN VARUS.
Patient problems: Bone Fracture(s); Injury; Pain
MAUDE report 10070027 ↗InjuryJul 2, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "INSTABILITY, LEFT KNEE."
Patient problems: Injury
MAUDE report 8754453 ↗InjuryApr 1, 2019Migration; Migration
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO SUBSIDENCE OF THE TIBIAL BASEPLATE (POSSIBLE CAUSE OR CONTRIBUTOR FOR SUBSIDENCE NOT REPORTED TO REP). SIZE 2 BASEPLATE AND INSERT WERE REVISED TO A UNIVERSAL BASEPLATE WITH CEMENTED STEM, AND A NEW INSERT. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE INSERT.
Patient problems: Injury
MAUDE report 8470303 ↗InjuryJan 15, 2019Osseointegration Problem; Osseointegration Problem
AS REPORTED: "PATIENT...HAD A LEFT PRIMARY TOTAL KNEE DONE...ON (B)(6) 2017. OVER THE PAST YEAR PATIENT...ENDED UP [AT SURGEON AND HOSPITAL] WITH A CHIEF COMPLAINT OF PAIN AND INSTABILITY. [SURGEON] TOLD ME THERE WAS POSITIVE UPTAKE ON A BONE SCAN AS WELL AS MRI FOR TIBIAL BASEPLATE LOOSENING. THE PATIENT WENT INTO SURGERY ON (B)(6) 2018 AND HAD THE SIZE 2 TRITANIUM BASEPLATE AND SIZE 2X9MM POLY REPLACED WITH A CEMENTED UNIVERSAL BASEPLATE AND 12X50 STEM AND 2X16MM POLY.
Patient problems: Inadequate Osseointegration; Injury; Pain
MAUDE report 8250294 ↗InjuryJan 10, 2019Loss of Osseointegration; Loss of Osseointegration
ON (B)(6) 2018 INITIAL PROCEDURE DONE. (B)(6) 2018 REVISION DONE. UPDATE 26-DECEMBER-2018: REVISION SURGERY DUE TO LEFT TIBIA LOOSENING.
Patient problems: Bone Fracture(s); Inadequate Osseointegration; Injury
MAUDE report 8235833 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.