Skip to content

Triathlon 5531-G-213Recalls & reported problems

Howmedica Osteonics Corp. · Tibial Bearing Insert · Class 2 · Implant

All models of Triathlon (608)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327045673), received between Oct 16, 2018 and Jan 8, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Unstable” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable4
  • Degraded1
  • Fracture1
  • Insufficient Device Problem Information1
  • Material Integrity Problem1

Patient problems

  • Injury2
  • Joint Laxity2
  • Ambulation or Postural Difficulties1
  • Fall1
  • Hematoma1
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2025Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO FLEXION INSTABILITY. A 2X13 CS INSERT WAS REVISED TO A 2X19 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Laxity

    MAUDE report 21108358 ↗
  • InjuryFeb 27, 2023Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY. INTRA-OPERATIVELY, WEAR WAS NOTED ON THE POSTERIOR MEDIAL ASPECT OF THE INSERT. A 2X13 CS INSERT WAS REVISED TO A 2X19 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Laxity

    MAUDE report 16441558 ↗
  • InjuryApr 13, 2021Degraded; Fracture; Degraded; Fracture

    THE CUSTOMER REPORTED THAT HE REVISED A TRIATHLON FOR INSTABILITY AND OPENED THE KNEE TO DISCOVER THAT THE POLY INSERT HAD BROKEN WHICH HAS CAUSED THE INSTABILITY. THE PATIENT HAD A FALL IN (B)(6) 2020. THE SURGEON ALSO REPORTED THAT THE PATELLA HAD A WEAR PATTERN WHICH THE SURGEON BELIEVES IS NOT RELATED TO THE IMPLANT BEING BROKEN. THE TKR WAS INITIALLY DONE ON (B)(6) 2010.

    Patient problems: Fall

    MAUDE report 11654687 ↗
  • InjuryDec 22, 2020Unstable; Unstable

    PROCEDURE: THE PATIENT UNDERWENT BILATERAL KNEE ARTHROPLASTY (B)(6) 2015. PATIENT HAD RIGHT KNEE PAIN AND INSTABILITY FOR MORE THAN TWO YEARS. THEN PATIENT UNDERWENT RIGHT TOTAL KNEE REVISION ON (B)(6) 2020. ALL COMPONENTS EXCHANGED.

    Patient problems: Ambulation or Postural Difficulties; Pain

    MAUDE report 11055290 ↗
  • MalfunctionJun 12, 2019Material Integrity Problem; Material Integrity Problem

    OPENED INSERT TO THE FIELD AND SURGEON NOTICED SOME LINES ON THE TOP SIDE OF THE POLY. UPDATE MAY 31, 2019: AS PER SALES REP PHONE CALL, "THE MARKS LOOKED LIKE UNFINISHED MACHINING GOUGES NOT MATERIAL TRANSFER." UPDATE: "RIGHT SIDE."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8691299 ↗
  • InjuryMay 22, 2019Unstable; Unstable

    IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "FAILED RIGHT TOTAL KNEE REPLACEMENT, TIBIAL LOOSENING AND INSTABILITY".

    Patient problems: Injury

    MAUDE report 8631554 ↗
  • InjuryOct 16, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    AS REPORTED: PATIENT UNDERWENT LEFT TOTAL KNEE REPLACEMENT (B)(6) 2018. THEN PATIENT HAD A LEFT REVISION KNEE ARTHROPLASTY, POLYETHYLENE LINER EXCHANGE AND IRRIGATION AND DEBRIDEMENT OF THE HEMATOMA (B)(6) 2018.

    Patient problems: Hematoma; Injury

    MAUDE report 7971080 ↗

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates