Triathlon 5531-G-213Recalls & reported problems
Howmedica Osteonics Corp. · Tibial Bearing Insert · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327045673), received between Oct 16, 2018 and Jan 8, 2025 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Unstable” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable4
- Degraded1
- Fracture1
- Insufficient Device Problem Information1
- Material Integrity Problem1
Patient problems
- Injury2
- Joint Laxity2
- Ambulation or Postural Difficulties1
- Fall1
- Hematoma1
- No Consequences Or Impact To Patient1
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJan 8, 2025Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO FLEXION INSTABILITY. A 2X13 CS INSERT WAS REVISED TO A 2X19 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Laxity
MAUDE report 21108358 ↗InjuryFeb 27, 2023Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY. INTRA-OPERATIVELY, WEAR WAS NOTED ON THE POSTERIOR MEDIAL ASPECT OF THE INSERT. A 2X13 CS INSERT WAS REVISED TO A 2X19 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Laxity
MAUDE report 16441558 ↗InjuryApr 13, 2021Degraded; Fracture; Degraded; Fracture
THE CUSTOMER REPORTED THAT HE REVISED A TRIATHLON FOR INSTABILITY AND OPENED THE KNEE TO DISCOVER THAT THE POLY INSERT HAD BROKEN WHICH HAS CAUSED THE INSTABILITY. THE PATIENT HAD A FALL IN (B)(6) 2020. THE SURGEON ALSO REPORTED THAT THE PATELLA HAD A WEAR PATTERN WHICH THE SURGEON BELIEVES IS NOT RELATED TO THE IMPLANT BEING BROKEN. THE TKR WAS INITIALLY DONE ON (B)(6) 2010.
Patient problems: Fall
MAUDE report 11654687 ↗InjuryDec 22, 2020Unstable; Unstable
PROCEDURE: THE PATIENT UNDERWENT BILATERAL KNEE ARTHROPLASTY (B)(6) 2015. PATIENT HAD RIGHT KNEE PAIN AND INSTABILITY FOR MORE THAN TWO YEARS. THEN PATIENT UNDERWENT RIGHT TOTAL KNEE REVISION ON (B)(6) 2020. ALL COMPONENTS EXCHANGED.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 11055290 ↗MalfunctionJun 12, 2019Material Integrity Problem; Material Integrity Problem
OPENED INSERT TO THE FIELD AND SURGEON NOTICED SOME LINES ON THE TOP SIDE OF THE POLY. UPDATE MAY 31, 2019: AS PER SALES REP PHONE CALL, "THE MARKS LOOKED LIKE UNFINISHED MACHINING GOUGES NOT MATERIAL TRANSFER." UPDATE: "RIGHT SIDE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8691299 ↗InjuryMay 22, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "FAILED RIGHT TOTAL KNEE REPLACEMENT, TIBIAL LOOSENING AND INSTABILITY".
Patient problems: Injury
MAUDE report 8631554 ↗InjuryOct 16, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED: PATIENT UNDERWENT LEFT TOTAL KNEE REPLACEMENT (B)(6) 2018. THEN PATIENT HAD A LEFT REVISION KNEE ARTHROPLASTY, POLYETHYLENE LINER EXCHANGE AND IRRIGATION AND DEBRIDEMENT OF THE HEMATOMA (B)(6) 2018.
Patient problems: Hematoma; Injury
MAUDE report 7971080 ↗
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- ATTUNE 1512-16-080Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page