Skip to content

Triathlon 5531-G-313Recalls & reported problems

Howmedica Osteonics Corp. · Tibial Bearing Insert · Class 2 · Implant

All models of Triathlon (608)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 07613327045741), received between Jul 10, 2018 and Mar 3, 2026 — classified as 13 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information7
  • Unstable3
  • Degraded2
  • Naturally Worn2
  • Device Dislodged or Dislocated1
  • Fracture1

Patient problems

  • Injury6
  • Unspecified Infection3
  • Aspiration/Inhalation2
  • Joint Laxity2
  • Pain2
  • Arthritis1
  • Discomfort1
  • Fever1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2026Degraded

    THE FOLLOWING INFORMATION WAS PROVIDED: IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO PAIN AND FLEXION INSTABILITY. INTRA-OPERATIVELY, SIGNIFICANT INSERT WEAR WAS NOTED POSTERIORLY/ MEDIALLY, AND A PORTION OF THE BASEPLATE WAS BROKEN OFF (SURGEON REPORTED POSSIBLE CAUSE/ CONTRIBUTOR WAS THAT THE BASEPLATE WAS IMPLANTED WITH TOO MUCH POSTERIOR SLOPE). A CR/ CS KNEE WAS CONVERTED TO A TS KNEE (PATELLAR COMPONENT WAS RETAINED). REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Pain

    MAUDE report 24494747 ↗
  • InjuryJul 9, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT WAS BROUGHT TO OR FOR REVISION TKR FOR SUBLUXATION OF TOTAL KNEE COMPONENTS.

    Patient problems: Joint Dislocation

    MAUDE report 19700162 ↗
  • InjuryApr 17, 2023Degraded; Unstable; Naturally Worn; Degraded; Unstable; Naturally Worn

    KNEE REVISION DUE TO INSTABILITY. POLY WEAR.

    Patient problems: Insufficient Information; Joint Laxity

    MAUDE report 16758735 ↗
  • InjuryMar 24, 2023Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO FLEXION INSTABILITY. INTRA-OPERATIVELY, THE FOLLOWING WERE NOTED FOR THE INSERT: POSTERIOR MEDIAL WEAR WITH A PIECE OF THE LIP CHIPPED OFF, POSTERIOR LATERAL WEAR WITH THE LIP WORN DOWN. A 3X13 CS INSERT WAS REVISED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Laxity

    MAUDE report 16610995 ↗
  • InjuryJan 31, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO LOOSENING OF THE TIBIAL BASEPLATE. INTRA-OPERATIVELY, THE LOCKING WIRE OF THE INSERT WAS OBSERVED TO HAVE BROKEN. THOUGH THE FEMORAL COMPONENT WAS WELL-FIXED, IT WAS REVISED IN ORDER TO ACCESS THE BASEPLATE. THE PATELLA WAS RETAINED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13405287 ↗
  • InjuryJun 17, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IN RESPONSE TO FOLLOW-UP FOR PI 2364534, MEDICAL RECORDS WERE PROVIDED WHICH INDICATE THE PATIENT HAD A LEFT KNEE ASPIRATION ON (B)(6) 2019. AS NOTED: "...SHE DEVELOPED FEVER, CHILLS, BILATERAL HIP AND LOW BACK DISCOMFORT...NO INFECTION WAS FOUND...WHEN SHE WALKS UP AND DOWN THE STAIRS SHE FEELS A POPPING IN HER LEFT KNEE ON THE INSIDE OF IT THAT ONLY HAPPENS WITH STAIRS...AN 18-GAUGE NEEDLE WAS USED TO ASPIRATE 50CC OF SEROSANGUINEOUS FLUID FROM THE POPLITEAL SPACE..."

    Patient problems: Aspiration/Inhalation; Discomfort; Fever

    MAUDE report 10162192 ↗
  • InjuryJun 17, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IN RESPONSE TO FOLLOW-UP, MEDICAL RECORDS WERE PROVIDED WHICH INDICATE THE PATIENT HAD A LEFT KNEE ASPIRATION ON (B)(6) 2019. AS NOTED: "...SHE HAS NOTICED THAT THE AREA IN THE BACK OF HER KNEE HAS 'SWELLED UP AGAIN.'...AN 18 GAUGE NEEDLE WAS USED TO ASPIRATE 90CC OF SANGUINEOUS FLUID FROM THE POPLITEAL SPACE."

    Patient problems: Aspiration/Inhalation; Loss of Range of Motion; Swelling/ Edema

    MAUDE report 10161523 ↗
  • InjuryMar 4, 2020Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO GLOBAL INSTABILITY. THERE ARE NO ALLEGATIONS AGAINST THE STRYKER IMPLANTS. A 3X13 CS INSERT WAS REVISED TO A 3X19 CS INSERT. REP REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: Injury

    MAUDE report 9785404 ↗
  • InjuryOct 18, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    THE PATIENT UNDERWENT INDEX LEFT TOTAL KNEE REPLACEMENT OUTSIDE OF CORS SCOPE. PATIENT UNDERWENT LEFT KNEE REVISION ON (B)(6) 2017 FOR PAIN AND DYSFUCTION ONLY POLYETHELINE WAS EXCHANGED. PATIENT CONTINUED WITH PAIN AND DYSFUNCTION, SO UNDERWENT ANOTHER REVISION, WITH OFF-LABEL USE ON (B)(6) 2019.

    Patient problems: Injury; Pain

    MAUDE report 9211043 ↗
  • InjuryJul 2, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "PATELLOFEMORAL ARTHRITIS, POST PARTIAL TOTAL RIGHT KNEE REPLACEMENT."

    Patient problems: Arthritis; Injury

    MAUDE report 8753403 ↗
  • InjuryMay 14, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    COVERED REVISION OF A TRIATHLON LT TKA ORIGINALLY IMPLANTED (B)(6) 2014. KNEE WAS FOUND TO BE INFECTED. ALL IMPLANTS WERE REMOVED AND ANTIBIOTIC SPACER WAS IMPLANTED. PATIENT REQUESTED TO KEEP REMOVED IMPLANTS AND DID.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 8608817 ↗
  • InjuryMay 1, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    POSTOPERATIVE DIAGNOSIS: THE PATIENT UNDERWENT LEFT TOTAL KNEE REPLACEMENT (B)(6) 2018. PATIENT HAD REVISION FOR THE INFECTED LEFT TOTAL KNEE ARTHROPLASTY. THE MODULAR POLYETHYLENE WAS EXCHANGED (B)(6) 2018.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 8568812 ↗
  • InjuryJul 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    INFECTED TOTAL KNEE. "REMOVAL OF IMPLANTS."

    Patient problems: Injury; Unspecified Infection

    MAUDE report 7674616 ↗

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page