Skip to content

Triathlon 5531-G-409Recalls & reported problems

Howmedica Osteonics Corp. · Tibial Bearing Insert · Class 2 · Implant

All models of Triathlon (608)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Howmedica Osteonics Triathlon tibial bearing insert is the plastic cushion inside a knee replacement. The 45 reports mostly describe the insert wearing down or shifting, leaving the metal parts of the knee rubbing together, along with pain and infection that led to revision surgery, where surgeons swapped in a larger or new insert. No deaths were reported and 40 injuries were. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
40
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 07613327045802), received between Feb 9, 2017 and Dec 3, 2025 — classified as 40 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information22
  • Unstable10
  • Adverse Event Without Identified Device or Use Problem5
  • Degraded4
  • Device Contamination with Chemical or Other Material1
  • Device Operates Differently Than Expected1
  • Device Slipped1
  • Difficult to Insert1

Patient problems

  • Injury19
  • Pain14
  • Unspecified Infection10
  • Ambulation or Postural Difficulties6
  • Aspiration/Inhalation5
  • Insufficient Information5
  • Joint Laxity4
  • No Known Impact Or Consequence To Patient4

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO INFECTION. AN I&D WITH WASHOUT AND POLY SWAP OF A 4X9 CS INSERT WAS PERFORMED.

    Patient problems: Unspecified Infection

    MAUDE report 23708911 ↗
  • InjuryNov 19, 2025Unstable; Unstable

    AS REPORTED: "PATIENT IS S/P REV RTKA DUE TO INSTABILITY. NO ADDITIONAL DETAILS TO BE REPORTED BY THE FACILITY." A CR/ CS KNEE WAS CONVERTED TO A TS KNEE WITH STEMS AND AUGMENTS.

    Patient problems: Joint Laxity

    MAUDE report 23592867 ↗
  • InjuryOct 29, 2025Degraded; Fracture; Degraded; Fracture

    IT WAS REPORTED THE PATIENT'S LEFT KNEE WAS REVISED. PATIENT PRESENTED WITH COMPLAINT OF PAIN AND A SUPERFICIAL WOUND INFECTION, AND REPORTED A PRIOR FALL AND A BUG BITE. UPON OPENING THE JOINT, METAL-STAINED TISSUE WAS OBSERVED, AS WELL AS POSTERIOR LATERAL WEAR OF THE INSERT TO THE POINT WHERE THE FEMORAL COMPONENT WAS ARTICULATING AGAINST THE BASEPLATE. SURGEON SUSPECTED THE FALL MAY HAVE DISRUPTED THE SOFT TISSUES OR SHIFTED THE IMPLANTS, CAUSING THE UNEVEN WEAR. THOUGH NO LAB RESULTS CAME BACK TO INDICATE WHETHER THE JOINT WAS INFECTED OR NOT, ALL COMPONENTS WERE REMOVED AND AN ANTIBIOTIC SPACER WAS PLACED AS A PRECAUTIONARY MEASURE. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Fall; Metal Related Pathology; Pain; Unspecified Infection

    MAUDE report 23417534 ↗
  • InjurySep 3, 2024Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY. A 4X9 CS INSERT WAS REVISED TO A 4X12 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Laxity

    MAUDE report 20124268 ↗
  • InjuryAug 6, 2024Degraded; Degraded

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO PAIN AND INFLAMMATION. A 4X9 CS INSERT WAS EXCHANGED FOR ANOTHER 4X9 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 19916603 ↗
  • InjuryMay 20, 2024Unstable; Unstable

    REVISED THE TIBIAL INSERT OF A LEFT TRIATHLON DUE TO POSTERIOR INSERT WEAR/INSTABILITY FROM PCL RUPTURE. HE UPSIZED FROM A 4X9 TO 4X10.

    Patient problems: Insufficient Information

    MAUDE report 19349797 ↗
  • InjurySep 22, 2023Degraded; Unstable; Degraded; Unstable

    DAMAGED X3 INSERT RETRIEVED DURING POLYETHYLENE EXCHANGE FOR KNEE PAIN / INSTABILITY.

    Patient problems: Joint Laxity; Pain

    MAUDE report 17803136 ↗
  • InjuryApr 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO POSSIBLE INFECTION. A 4X9 CS INSERT WAS REMOVED AND A 4X10 CS INSERT WAS IMPLANTED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Unspecified Infection

    MAUDE report 16820531 ↗
  • InjuryOct 12, 2022Pitted; Pitted

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO TIBIAL SUBSIDENCE. INTRA-OPERATIVELY, THE SURGEON NOTED THAT THERE APPEARED TO BE PITTING IN THE POLY. THE BASEPLATE AND INSERT WERE REVISED.

    Patient problems: Insufficient Information

    MAUDE report 15585302 ↗
  • InjuryJun 7, 2022Unstable; Unstable

    THE FOLLOWING WAS REPORTED: 9MM TIBIAL INSERT WAS EXPLANTED AND AN 11MM INSERT WAS IMPLANTED TO ADDRESS INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 14617123 ↗
  • InjuryMay 9, 2022Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information

    THE FOLLOWING WAS REPORTED: KNEE REVISION.

    Patient problems: Insufficient Information

    MAUDE report 14342957 ↗
  • InjuryDec 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: PATIENTS LEFT KNEE POLY WAS SWAPPED DUE TO INFECTION. INITIAL OP DATE WAS (B)(6) 2018.

    Patient problems: Unspecified Infection

    MAUDE report 12907211 ↗
  • InjuryOct 27, 2021Unstable; Unstable

    AS REPORTED: "REVISION WAS PERFORMED FOR INSTABILITY OF THE JOINT IN EXTENSION. NO OTHER INFORMATION AVAILABLE FROM THE HOSPITAL." LEFT KNEE. PATIENT WAS REVISED FROM A CS TIBIAL INSERT AND CR FEMORAL COMPONENT TO A PS TIBIAL INSERT AND A TS FEMORAL COMPONENT WITH AUGMENTS AND A STEM.

    Patient problems: Insufficient Information

    MAUDE report 12704308 ↗
  • InjuryDec 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO SUSPECTED INFECTION. A LINER EXCHANGE AND WASHOUT WAS DONE. SURGEON ALSO WENT FROM A 4X9 CS INSERT TO A 4X16 CS INSERT. THE REP REPORTED THAT THE SURGEON THOUGHT A THICKER INSERT WOULD BE MORE STABLE, BUT NO PRIOR INSTABILITY WAS REPORTED. THE REP CONFIRMED THAT NO FURTHER INFORMATION WOULD BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Unspecified Infection

    MAUDE report 10975599 ↗
  • InjuryNov 16, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO RIGHT KNEE PAIN.

    Patient problems: Pain

    MAUDE report 10842936 ↗
  • InjuryNov 15, 2020Loss of Osseointegration; Insufficient Device Problem Information; Loss of Osseointegration; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO A FAILED LEFT TOTAL KNEE ARTHROPLASTY. 16 DEC 2020 , AS PER MEDICAL REVIEW "TIBIAL WAS GROSSLY LOOSE ", OTHER DEVICES WILL THEREFORE BE CONSIDERED CONCOMITANT

    Patient problems: Inadequate Osseointegration; Insufficient Information

    MAUDE report 10841558 ↗
  • InjuryJun 19, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION RECEIVED INCLUDES A REPORT DATED (B)(6) 2013: "UNFORTUNATELY, SHE HAD A FALL A WEEK BACK AND SINCE THEN HER KNEE HAS BEEN SWOLLEN...WITH REGARDS TO HER HEMARTHROSIS, WE PROCEEDED TO ASPIRATE OUT ABOUT 70 CC OF BLOOD FROM HER KNEE JOINT UNDER ALL ASEPTIC PRECAUTIONS..."

    Patient problems: Aspiration/Inhalation; Fall; Loss of Range of Motion

    MAUDE report 10173123 ↗
  • InjuryJun 19, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION RECEIVED INCLUDES A REPORT DATED (B)(6) 2013: "DIAGNOSIS: INFECTED RIGHT KNEE...THE KNEE IS RED AND SWOLLEN. THIS WAS PREPPED AND ASPIRATED OF 20 CC OF BLOODY FLUID. ALSO DURING THE PROCEDURE SHE HAD A LATERAL RELEASE..." PATIENT WAS PLACED ON ANTIBIOTICS.

    Patient problems: Ambulation or Postural Difficulties; Aspiration/Inhalation; Inflammation; Pain; Swelling; Unspecified Infection

    MAUDE report 10173122 ↗
  • InjuryJun 18, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION RECEIVED IN FOLLOW-UP INCLUDES A REPORT DATED SEPTEMBER 11, 2013: "HER MAIN CONCERN TODAY IS THE EFFUSION THAT HAS BEEN BEHIND HER RIGHT KNEE. SHE NOTES THAT IT IS MODERATELY PAINFUL. 20 CC OF SEROSANGUINEOUS FLUID WERE DRAINED FROM THE KNEE."

    Patient problems: Ambulation or Postural Difficulties; Aspiration/Inhalation; Inflammation; Pain

    MAUDE report 10167530 ↗
  • InjuryJun 18, 2020Unstable; Insufficient Device Problem Information; Unstable; Insufficient Device Problem Information

    INFORMATION RECEIVED IN FOLLOW-UP INCLUDES A REPORT DATED (B)(6) 2014: "STILL WITH A FEELING OF INSTABILITY AND GIVING WAY. A THROUGH SUPEROMEDIAL PORTAL OF KNEE ASPIRATION OF INTRA-ARTICULAR FLUID WAS DONE WITH ABOUT 20CC FOLLOWED BY LIGHT DRESSING.''

    Patient problems: Ambulation or Postural Difficulties; Aspiration/Inhalation; Fall

    MAUDE report 10167527 ↗
  • InjuryJun 18, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION RECEIVED IN FOLLOW-UP INCLUDES A REPORT DATED (B)(6) 2013: "...SHE HAS AN EFFUSION ABOUT THE RIGHT KNEE...SHE WAS ASPIRATED TODAY TAKING OUT APPROXIMATELY 20CC OF BLOOD-TINGED FLUID."

    Patient problems: Ambulation or Postural Difficulties; Aspiration/Inhalation; Inflammation; Pain

    MAUDE report 10167526 ↗
  • InjuryMay 19, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT'S KNEE WAS REVISED. AS REPORTED BY REP: "REVISED FOR INFECTION. NO OTHER INFORMATION IS AVAILABLE PER HOSPITAL POLICY."

    Patient problems: Injury; Unspecified Infection

    MAUDE report 10072448 ↗
  • InjuryMar 31, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IMPLANTS REMOVED DUE TO INFECTION. DOCTOR STATED THAT THE INFECTION WAS SECONDARY TO AN INFECTION FROM PRESSURE SORES ON THE PATIENT.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 9903783 ↗
  • InjuryMar 24, 2020Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    PATIENT HAD ORIGINAL KNEE DONE ON (B)(6) 2009 BY DR. AT HOSPITAL. PRESENTED TODAY FOR LEFT KNEE PAIN AND A POLY EXCHANGE BY DR.

    Patient problems: Injury; Pain

    MAUDE report 9872227 ↗
  • InjuryMar 23, 2020Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO INSTABILITY. A 4X9 CS INSERT AND CR FEMUR WERE REVISED TO A 4X11 PS INSERT AND PS FEMUR.

    Patient problems: Injury

    MAUDE report 9866445 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page