Adverse event reports
16
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 07613327045949), received between Mar 29, 2017 and Jul 28, 2025 — classified as 12 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information5
  • Adverse Event Without Identified Device or Use Problem4
  • Unstable3
  • Mechanical Problem2
  • Connection Problem1
  • Device Contamination with Chemical or Other Material1
  • Device Difficult to Setup or Prepare1
  • Difficult to Insert1

Patient problems

  • Unspecified Infection7
  • Injury6
  • No Known Impact Or Consequence To Patient4
  • Pain2
  • Ambulation or Postural Difficulties1
  • Bacterial Infection1
  • Limited Mobility Of The Implanted Joint1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE IMPLANTED OVER 10 YEARS AGO WAS REVISED DUE TO A SUSPECTED INFECTION. A 6X13 CS INSERT WAS EXCHANGED FOR ANOTHER. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Unspecified Infection

    MAUDE report 22637802 ↗
  • InjuryMar 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A POLY EXCHANGE DUE TO KNEE PAIN. PATIENT NOTED THAT THE PRIMARY WAS DONE IN ANOTHER LOCATION. NO ADDITIONAL INFORMATION AVAILABLE DUE TO HOSPITAL POLICY.

    Patient problems: Pain

    MAUDE report 21586597 ↗
  • InjuryDec 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT UNDERWENT PRIMARY RIGHT TKA OUTSIDE CORS SCOPE. ON (B)(6), 2021 THE PATIENT WAS REVISED FOR MECHANICAL LOOSENING. PATIENT PRESENTED WITH PAIN AND SIGNS OF PJI. WAS REVISED ON (B)(6), 2022. ALL COMPONENTS EXCHANGED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 15976780 ↗
  • InjuryOct 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED FOR SUSPECTED INFECTION. A POLY SWAP WAS DONE. OTHER THAN THE PROVIDED REVISION USAGE SHEET, NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Unspecified Infection

    MAUDE report 12671784 ↗
  • InjuryOct 2, 2020Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY. A 6X13 CS INSERT WAS REVISED TO A 6X19 CS INSERT. REP PROVIDED THE REVISION USAGE SHEET AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 10619036 ↗
  • InjurySep 15, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO INFECTION. A 6X13 CS POLY WAS EXCHANGED FOR ANOTHER 6X13 CS POLY.

    Patient problems: Bacterial Infection

    MAUDE report 10534489 ↗
  • MalfunctionMar 30, 2020Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    PRIMARY PROCEDURE, LEFT KNEE. IT WAS REPORTED THAT THE INSERT WOULD NOT SEAT INTO THE BASEPLATE. A 2ND IMPLANT OF THE SAME LOT CODE WAS OPENED AND SEATED, BUT ADDITIONAL EFFORT THAN NORMAL WAS REQUIRED. PROCEDURE COMPLETED SUCCESSFULLY WITH NO DELAY. THE WASTED INSERT IS ALLEGEDLY AVAILABLE FOR RETURN. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9897820 ↗
  • InjuryFeb 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    AS REPORTED: "LEFT KNEE REVISION. REVISION DONE DUE TO INFECTION. ALL IMPLANTS WERE EXPLANTED."

    Patient problems: Injury; Unspecified Infection

    MAUDE report 9702562 ↗
  • InjuryFeb 10, 2020Unstable; Unstable

    KNEE REVISION DUE TO INSTABILITY.

    Patient problems: Injury

    MAUDE report 9688502 ↗
  • InjuryOct 15, 2019Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY. A 6X13 CS INSERT WAS REVISED TO A 6X16 CS X3 INSERT. THE SURGEON NOTED THAT THE POLY HAD SOME YELLOWING AND WOULD LIKE TO ENSURE THAT IT IS NOT OXIDATION. THE EXPLANT IS AVAILABLE FOR RETURN, AND THE REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: Injury; Limited Mobility Of The Implanted Joint

    MAUDE report 9191852 ↗
  • MalfunctionSep 9, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    A TRIATHLON IMPLANT TRIATHLON CS X3 INSERT #6 13MM WHICH WAS EXPOSED TO WATER AND STEAM AT (B)(6) HOSPITAL, WAS REDISTRIBUTED BY STRYKER (B)(4) AND SUBSEQUENTLY IMPLANTED DURING A PROCEDURE AT (B)(6) HOSPITAL. A FULL LOAN TRIATHLON IMPLANT SET WAS ACCIDENTALLY RETRIEVED FROM A HOSPITAL BY A COURIER AND WAS RETURNED TO A STRYKER KIT ROOM. HOWEVER, THE COURIER WAS UNAWARE THAT THE ITEMS WERE AFFECTED AND THEREFORE DID NOT COMMUNICATE TO THE KIT ROOM THAT THESE PARTS WERE PREVIOUSLY QUARANTINED AT THE HOSPITAL. AS A RESULT THE KIT ROOM DID NOT PROCEED WITH ANY QUARANTINE ACTION AND 18 OF THE AFFECTED ITEMS WERE RE-DISTRIBUTED. UPDATE 19/AUGUST/2019: NONE OF THE PACKAGING WAS RETAINED FROM THE IMPLANTED DEVICES. SURGEONS EXPRESSED NO CONCERN, AND HAVE COMMUNICATED NO PLANS TO SURGICALLY ADDRESS THE IMPLANTATION OF THE DEVICES. TO DATE, NO ADVERSE CONSEQUENCES HAVE BEEN REPORTED FOR THESE PATIENTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8974456 ↗
  • MalfunctionJul 1, 2019Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THERE IS A GAP BETWEEN THE INSERT AND TIBIA TRAY AFTER IMPACTION. THIS ISSUE WAS FOUND OUT DURING THE SURGERY. UPDATE: DELAY OF SURGERY: 20 MIN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8748038 ↗
  • InjuryAug 6, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    THE SURGEON EXPLANTED TRIATHLON PRIMARY TKA IMPLANTS FROM A RIGHT INFECTED TOTAL KNEE AND PLACED A TEMPORARY SPACER. ORIGINAL SURGERY WAS (B)(6) 2018 AND AN I AND D WITH TIBIAL INSERT EXCHANGE WAS DONE ON (B)(6) 2018.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 7754316 ↗
  • InjuryApr 2, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    SURGEON DID AN I&D WITH INSERT EXCHANGE, DUE TO POSSIBLE INFECTION.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 7390711 ↗
  • InjuryMay 31, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    REVISION OF A PRIMARY KNEE WITH POLY SWAP DUE TO INFECTION.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 6600549 ↗
  • MalfunctionMar 29, 2017Difficult to Insert; Mechanical Problem; Fitting Problem; Connection Problem; Difficult to Insert; Mechanical Problem; Fitting Problem

    POLY INSERT DIDN'T LOCK CORRECTLY IN BASEPLATE. THEY USED ANOTHER INSERT WHICH LOCKED INTO THE BASEPLATE WITH NO ISSUES. THEY DID NOT NEED TO EXCHANGE THE BASEPLATE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 6444746 ↗

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →