Skip to content

L10 0220220300Recalls & reported problems

Stryker Corporation · LED Light Source with AIM Technology · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker LED light source used for confocal optical imaging during surgery. Most of the 76 reports describe the light staying on or getting too hot — burning through surgical drapes and causing thermal burns — plus loss of light output that sometimes cancelled procedures, and electronics or sensor problems. There were 8 injuries and no deaths, and no recalls name this device. Reports in the last 12 months dropped to none from 7 in the prior12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 07613327051377), received between Oct 15, 2018 and Jun 12, 2026 — classified as 8 injuries, 69 malfunctions.

The most frequently reported specific device problem is “Output Problem” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 4 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem32
  • Temperature Problem18
  • Adverse Event Without Identified Device or Use Problem10
  • Device Sensing Problem8
  • Electrical /Electronic Property Problem8
  • Overheating of Device6
  • Thermal Decomposition of Device6
  • Insufficient Device Problem Information4

Patient problems

  • No Clinical Signs, Symptoms or Conditions51
  • Insufficient Information20
  • No Known Impact Or Consequence To Patient20
  • Burn(s)5
  • Burn, Thermal5
  • No Patient Involvement2
  • Superficial (First Degree) Burn2

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2026Overheating of Device; Device Sensing Problem

    IT WAS REPORTED THAT TEST ALL ITEMS (L10 LIGHT SOURCE (REF 220-220-300/SN (B)(6)), SAFELIGHT LIGHT CABLE, AND SAFELIGHT ADAPTER) CONNECTED TO THE EVENT IMMEDIATELY UPON AVAILABILITY OF THE ROOM, CONFIRMING THAT THE ACTION CONTINUES TO OCCUR ON THIS DEVICE ONLY. THE LIGHT CABLE AND ADAPTER FUNCTIONED AS INTENDED WITH A SEPARATE DEVICE. ADDITIONALLY, WE TESTED A SEPARATE LIGHT CABLE AND ADAPTER WITH THE DEVICE IN QUESTION, WHICH AGAIN RESULTED IN A FAILED TRIGGERING OF THE SAFELIGHT SAFETY FUNCTION. THIS ONLY HAPPENED IN ONE PROCEDURE. NOT AWARE ABOUT ISSUE WITH THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25487289 ↗
  • MalfunctionAug 8, 2025Overheating of Device; Output Problem; Overheating of Device; Output Problem

    IT WAS REPORTED THAT THE LIGHT CABLE BURNED THE DRAPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22754798 ↗
  • MalfunctionJul 16, 2025Display or Visual Feedback Problem; Overheating of Device; Device Sensing Problem; Display or Visual Feedback Problem; Overheating of Device; Device Sensing Problem

    IT WAS REPORTED THAT LIGHT SOURCE THAT WOULD NOT IMMEDIATELY TURN OFF WHEN PRESSING THE CONSOLE TOUCH SCREEN. THEREFORE, THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22524450 ↗
  • MalfunctionMay 21, 2025Overheating of Device; Output Problem; Temperature Problem; Overheating of Device; Output Problem; Temperature Problem

    IT WAS REPORTED THAT WHEN DISCONNECTING LIGHT CORD FROM SAFELIGHT ADAPTER THE LIGHT STAYED ON AND DID NOT TURN OFF. THEREFORE, THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22066144 ↗
  • MalfunctionApr 16, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21852454 ↗
  • MalfunctionApr 11, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT THE LIGHT BOX WENT DOWN AND THAT THERE WAS A FULL LOSS OF LIGHT. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY AND WAS CANCELLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21826187 ↗
  • MalfunctionApr 9, 2025Computer Software Problem; Output Problem; Computer Software Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21809428 ↗
  • MalfunctionApr 9, 2025Material Frayed; Output Problem; Material Frayed; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21809409 ↗
  • MalfunctionMay 23, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19383633 ↗
  • MalfunctionMay 23, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19383620 ↗
  • MalfunctionMay 23, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19383613 ↗
  • MalfunctionApr 25, 2024Mechanical Problem; Output Problem; Mechanical Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19186360 ↗
  • MalfunctionFeb 26, 2024Device Sensing Problem; Output Problem; Device Sensing Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18778783 ↗
  • MalfunctionFeb 2, 2024Circuit Failure; Output Problem; Circuit Failure; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18632637 ↗
  • MalfunctionJan 17, 2024Device Sensing Problem; Temperature Problem; Device Sensing Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18530808 ↗
  • MalfunctionNov 28, 2023Device Sensing Problem; Output Problem; Device Sensing Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18220830 ↗
  • MalfunctionNov 28, 2023Device Sensing Problem; Output Problem; Device Sensing Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18220803 ↗
  • MalfunctionNov 21, 2023Sensing Intermittently; Output Problem; Sensing Intermittently; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18187547 ↗
  • MalfunctionOct 20, 2023Device Sensing Problem; Optical Problem; Device Sensing Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17978524 ↗
  • MalfunctionJun 1, 2023Device Sensing Problem; Output Problem; Device Sensing Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17039863 ↗
  • MalfunctionMay 18, 2023Energy Output Problem; Output Problem; Energy Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16953272 ↗
  • MalfunctionMay 4, 2023Energy Output Problem; Output Problem; Energy Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16867367 ↗
  • MalfunctionFeb 23, 2023Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16427900 ↗
  • MalfunctionFeb 16, 2023Thermal Decomposition of Device; Device Emits Odor; Device Emits Odor

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16391656 ↗
  • MalfunctionOct 31, 2022Output Problem; Temperature Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15700950 ↗

FDA clearance and classification

Cleared through Special 510(k) K191046 on May 10, 2019 (decision: Substantially Equivalent). Submitted by Stryker as “L10 LED Light Source with AIM, L11 LED Light Source with AIM”. The FDA review took 21 days. FDA record ↗

Cleared through 510(k) K142310 on Dec 2, 2014 (decision: Substantially Equivalent). Submitted by Stryker Endoscopy as “Stryker Infrared Fluorescence (IRF) Imaging System”. The FDA review took 105 days. FDA record ↗

The FDA regulates product code OWN (confocal optical imaging) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the L10 0220220300

Has the L10 0220220300 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613327051377), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the L10 0220220300?

The FDA's MAUDE database lists 77 adverse event reports for this device, received between Oct 15, 2018 and Jun 12, 2026. In those reports, the most common specific device problems are “Output Problem” (32), “Temperature Problem” (18) and “Device Sensing Problem” (8); the most common patient problems are “Burn(s)” (5), “Burn, Thermal” (5) and “Superficial (First Degree) Burn” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the L10 0220220300?

None of the reports are classified as deaths, 8 as injuries and 69 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the L10 0220220300 FDA approved?

It was cleared through 510(k) K142310 on Dec 2, 2014; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the L10 0220220300, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand L10, model 0220220300). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Confocal optical imaging devices

All Confocal optical imaging devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page