Quick summary

This Stryker SafeLight fiber optic light cord is used with endoscopy towers to illuminate the inside of the body during surgery. The 55 reports mostly describe the cable failing to light properly or shutting off mid-procedure, flickering, and getting hot enough to burn the light cord itself or the patient, including one report of a burn after the scope came off the cord and the safety shutoff failed. There are no deaths and no injuries reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 07613327462777), received between Jul 27, 2021 and Sep 10, 2026 — classified as 56 malfunctions.

The most frequently reported specific device problem is “Output Problem” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem37
  • Optical Problem23
  • Temperature Problem9
  • Overheating of Device7
  • Electrical Power Problem2
  • Intermittent Energy Output2
  • Adverse Event Without Identified Device or Use Problem1
  • Continuous Firing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • Insufficient Information6
  • Burn(s)5
  • Superficial (First Degree) Burn2

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Overheating of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50725009 ↗
  • MalfunctionAug 28, 2026Optical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50500585 ↗
  • MalfunctionAug 17, 2026Overheating of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50288688 ↗
  • MalfunctionAug 6, 2026Overheating of Device

    Patient problems: Burn(s)

    MAUDE report 50165965 ↗
  • MalfunctionMay 21, 2026Overheating of Device; Electrical Shorting

    IT WAS REPORTED THAT LIGHT CABLE DID NOT TURN OFF WHEN ADAPTER/SCOPE WAS REMOVED CAUSING A BURN ON THE PATIENT. FIRST DEGREE BURN TO PATIENT.

    Patient problems: Burn(s); Superficial (First Degree) Burn

    MAUDE report 25248513 ↗
  • MalfunctionNov 21, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT BURN OCCURRED ON LIGHT CORD, DURING PROCEDURE. POWER WAS NOT ON AND DEVICE HAD BEEN OFF FOR SOME TIME. AND THERE WAS SMOKE IN THE ROOM. PROCEDURE COMPLETED USING ROBOTIC PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23619840 ↗
  • MalfunctionOct 30, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THERE WAS AN OUT OF BOX FAILURE FOR QUANTITY 1 SAFELIGHT OPTIC CORD. THEY WENT LIVE WITH THE NEW EQUIPMENT A WEEK OR TWO AGO AND THE ATTACHED LIGHT CORD DOESN¿T TURN OFF WHEN DISCONNECTED FROM THE SAFELIGHT ADAPTER. THEY HAVE TRIED OTHER LIGHT CORDS / LIGHT SOURCES / ADAPTERS TO NARROW THE CHALLENGE TO THIS LIGHT CORD. THEREFORE, THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23425762 ↗
  • MalfunctionAug 4, 2025Optical Problem; Output Problem; Optical Problem; Output Problem

    IT WAS REPORTED THAT THE SAFELIGHT CABLE WAS FLICKERING AND THAT THERE WAS 5 MINUTES OF LOSS OF LIGHT. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22708418 ↗
  • MalfunctionJun 25, 2025Optical Problem; Output Problem; Optical Problem; Output Problem

    IT WAS REPORTED BY THE CUSTOMER THAT "DURING DISSECTION ON A LAPAROSCOPIC HYSTERECTOMY, THE STRYKER TOWER ANNOUNCED "LIGHT SOURCE DEACTIVATED". IT THEN CONTINUED TO STROBE BETWEEN ON AND OFF, ALTERNATING SAYING "LIGHT SOURCE ACTIVATED AND LIGHT SOURCE DEACTIVATED". GYNE CHARGE NURSE WAS QUICKLY RETRIEVED FOR THE SITUATION. A NEW TOWER AND CAMERA AND LAPAROSCOPE WERE QUICKLY ACQUIRED, BUT AFTER HOOKING THEM UP, THE SAME PROBLEM OCCURRED. THE TEAM THEN TRIED A NEW LIGHT CORD, AND THIS FIXED THE PROBLEM. THE LIGHT CORD WAS ISOLATED IN A ZIPLOC BAG WITH A NOTE TO RETURN IT TO THE CNL UPON CLEANING". THEREFORE, THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332862 ↗
  • MalfunctionApr 9, 2025Overheating of Device; Temperature Problem; Overheating of Device; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21809408 ↗
  • MalfunctionApr 3, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21767234 ↗
  • MalfunctionFeb 10, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT THAT RESULTED IN A 2ND DEGREE BURN.

    Patient problems: Burn(s)

    MAUDE report 21345918 ↗
  • MalfunctionJan 14, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: Insufficient Information

    MAUDE report 21154285 ↗
  • MalfunctionJan 14, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: Insufficient Information

    MAUDE report 21154133 ↗
  • MalfunctionDec 27, 2024Electrical Power Problem; Output Problem; Temperature Problem; Electrical Power Problem; Output Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT AND A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21028043 ↗
  • MalfunctionDec 18, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: Burn(s)

    MAUDE report 20972705 ↗
  • MalfunctionDec 16, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20939341 ↗
  • MalfunctionDec 4, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20846644 ↗
  • MalfunctionSep 17, 2024Electrical Power Problem; Output Problem; Electrical Power Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20241765 ↗
  • MalfunctionJun 5, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THE PATIENT WAS BURNED.

    Patient problems: Burn(s)

    MAUDE report 19472422 ↗
  • MalfunctionMay 16, 2024Continuous Firing; Temperature Problem; Continuous Firing; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A THERMAL EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19337563 ↗
  • MalfunctionApr 25, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181722 ↗
  • MalfunctionOct 6, 2023Optical Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17890119 ↗
  • MalfunctionOct 5, 2023Optical Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17876262 ↗
  • MalfunctionOct 5, 2023Optical Problem; Optical Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF LIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17876248 ↗