Formula 0375941000Recalls & reported problems

Stryker Corporation · [Aggressive 6-Flute Barrel Bur, Arthroscopic Shaver Blade, Do Not Resterilize, Do Not Use if Package is Damaged, Keep Dry, Keep Away from Sunlight] · Class 1

Adverse event reports
22
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 07613327054743), received between Jul 30, 2018 and Mar 25, 2026 — classified as 4 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Break” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break11
  • Material Fragmentation7
  • Component Misassembled1
  • Delivered as Unsterile Product1
  • Device Contaminated During Manufacture or Shipping1
  • Packaging Problem1
  • Patient Device Interaction Problem1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Foreign Body In Patient3
  • No Known Impact Or Consequence To Patient3
  • Insufficient Information1
  • Muscle/Tendon Damage1
  • No Patient Involvement1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2026Smoking

    IT WAS REPORTED THAT THERE WAS EXCESSIVE FRICTION AND SMOKE FROM THE MOMENT IT WAS SWITCHED ON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691305 ↗
  • MalfunctionJul 2, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE BUR CRACKED AND FILINGS WERE PRESENT IN THE JOINT. THE INTRA-ARTICULAR METAL DEBRIS WAS ASPIRATED AT END OF PROCEDURE. NO CONSEQUENCES TO DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22384266 ↗
  • MalfunctionJul 2, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE BUR CRACKED AND FILINGS WERE PRESENT IN THE JOINT. THE INTRA-ARTICULAR METAL DEBRIS WAS ASPIRATED AT END OF PROCEDURE. NO CONSEQUENCES TO DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22384263 ↗
  • MalfunctionJul 2, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE BUR CRACKED AND FILINGS WERE PRESENT IN THE JOINT. THE INTRA-ARTICULAR METAL DEBRIS WAS ASPIRATED AT END OF PROCEDURE. NO CONSEQUENCES TO DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22384260 ↗
  • InjuryJan 10, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT A SMALL PIECE OF THE TIP OF A BUR BROKE OFF INSIDE A PATIENT. IT WAS NOTED THAT THE BROKEN PIECE MIGHT HAVE BEEN SUCKED OUT, BUT THAT IS NOT CONFIRMED. LOOKED FOR THE PIECE FOR A WHILE AND THEN OPENED UP A NEW BUR. THERE ARE PRESENTLY NO ADVERSE CONSEQUENCES TO THE PATIENT, BUT HARD TO SAY FOR THE LONG TERM NOW.

    Patient problems: Foreign Body In Patient

    MAUDE report 21130578 ↗
  • MalfunctionOct 18, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT THERE WAS A BREACH IN THE STERILE BARRIER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20488097 ↗
  • MalfunctionOct 9, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20412305 ↗
  • InjurySep 13, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE SHAVER BLADE BROKE IN THE PATIENT'S SHOULDER. THE PIECE OF METAL WAS NEVER RECOVERED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20212378 ↗
  • MalfunctionApr 12, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. PLEASE NOTE, THE DEVICE DID NOT BREAK WITHIN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19097422 ↗
  • MalfunctionJun 6, 2023Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17074237 ↗
  • MalfunctionMay 25, 2023Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT THERE WAS A BREACH IN THE STERILE BARRIER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17005510 ↗
  • MalfunctionMay 10, 2023Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16907212 ↗
  • MalfunctionMay 10, 2023Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16907202 ↗
  • MalfunctionMar 21, 2023Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16584721 ↗
  • MalfunctionJan 6, 2023Component Misassembled; Break; Component Misassembled

    IT WAS REPORTED THAT THE DEVICE BROKE DURING PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16111057 ↗
  • MalfunctionSep 28, 2022Patient Device Interaction Problem; Break; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE DEVICE BROKE LEADING TO AN INJURY TO THE MUSCLE OF THE PATIENT.

    Patient problems: Muscle/Tendon Damage; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15496270 ↗
  • MalfunctionJun 21, 2022Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14754600 ↗
  • InjuryJun 29, 2021Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE AND A PIECE POTENTIALLY REMAINED IN THE PATIENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12086488 ↗
  • MalfunctionJan 8, 2020Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THERE WAS FOREIGN MATERIAL INSIDE THE STERILE PACKAGING.

    Patient problems: No Patient Involvement

    MAUDE report 9564716 ↗
  • InjuryMar 26, 2019Break

    PER ADDITIONAL INFORMATION RECEIVED, THE BURS BROKE AT THE PROXIMAL END. THE BURS BREAKING AT THE HUB WOULD CAUSE THE BUR TO BE UNUSABLE BUT WILL NOT CAUSE HARM TO THE PATIENT. THIS SHOULD NOT RESULT IN ANY ADVERSE CONSEQUENCES IF IT WERE TO RECUR. THEREFORE MAKING THIS EVENT NOT REPORTABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8453038 ↗
  • MalfunctionOct 9, 2018Break

    IT WAS REPORTED THAT THE SHEATH BROKE OFF. NO PIECES FELL INTO THE JOINT SPACE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7949809 ↗
  • MalfunctionJul 30, 2018Break

    IT WAS REPORTED THAT THE TIPS BROKE DURING THE PROCEDURE. THE PIECES FELL OFF OUTSIDE OF THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7733065 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.