Skip to content

Formula 0385950008Recalls & reported problems

Stryker Corporation · XL, Hip 8-Flute Round Bur. Arthroscopic Shaver Blade. For use with: REF 0375-701-500, 0375-704-500, 0375-708-500. Do Not Use if Package is Damaged, Keep Dry, Keep Away from Sunlight · Class 1

All models of Formula (43)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327055177), received between Aug 19, 2025 and Aug 19, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE TIP BROKE OFF. THE BROKEN TIP WAS REMOVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22844390 ↗

Other BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices

All BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page