Pneumosure 0620040690Recalls & reported problems

Stryker Corporation · [High Flow Insufflator, Do not use if package is damaged, Keep dry, Keep away from sunlight, Protect from heat and radioactive sources.] · Class 2

All models of Pneumosure (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 07613327055658), received between Aug 30, 2018 and Nov 27, 2024 — classified as 22 malfunctions.

The most frequently reported specific device problem is “Disconnection” (8 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 7, 2019Class IITerminated

    The heated tube sets were consistently leaking and detaching from the cassette.

    Z-1728-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Disconnection8
  • Decrease in Pressure4
  • Output Problem4
  • Therapeutic or Diagnostic Output Failure3
  • Appropriate Device Problem Term/Code Not Available2
  • Leak/Splash2
  • Connection Problem1
  • Contamination /Decontamination Problem1

Patient problems

  • No Known Impact Or Consequence To Patient15
  • No Clinical Signs, Symptoms or Conditions6
  • Bradycardia1
  • Insufficient Information1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2024Decrease in Pressure; Output Problem; Decrease in Pressure; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20800965 ↗
  • MalfunctionNov 27, 2024Decrease in Pressure; Output Problem; Decrease in Pressure; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20800808 ↗
  • MalfunctionNov 27, 2024Decrease in Pressure; Output Problem; Decrease in Pressure; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20800781 ↗
  • MalfunctionNov 27, 2024Decrease in Pressure; Output Problem; Decrease in Pressure; Output Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20800773 ↗
  • MalfunctionJan 20, 2022Connection Problem

    ADULT FEMALE WITH HISTORY OF OBESITY AND PERSISTENT ATRIAL FIBRILLATION. PROCEDURE: HYBRID CONVERGENT MAZE PROCEDURE WITH POSTERIOR LEFT ATRIAL WALL ABLATION, LEFT ATRIAL APPENDAGE EXCLUSION WITH ATRICLIP. ISSUE: STRYKER CO2 HIGH FLOW TUBING WOULD NOT CONNECT TO THE STRYKER TOWER. NEW TUBING OBTAINED AND WORKED APPROPRIATELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13306598 ↗
  • MalfunctionNov 23, 2020Infusion or Flow Problem; Therapeutic or Diagnostic Output Failure; Infusion or Flow Problem

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING SURGERY DUE TO TUBING DISCONNECTION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10886494 ↗
  • MalfunctionAug 26, 2019Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TUBING BECAME DETACHED, WHICH COULD POTENTIALLY LEAD TO A LOSS OF INSUFFLATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8932398 ↗
  • MalfunctionMay 8, 2019Disconnection; Leak/Splash

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DUE TO INSUFFLATION TUBING LEAKING AND DISCONNECTING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8592661 ↗
  • MalfunctionMay 8, 2019Disconnection; Leak/Splash

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DUE TO INSUFFLATION TUBING LEAKING AND DISCONNECTING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8592645 ↗
  • MalfunctionApr 30, 2019Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8563343 ↗
  • MalfunctionMar 27, 2019Disconnection

    IT WAS REPORTED THAT DURING SURGERY THERE WAS A POTENTIAL LOSS OF INSUFFLATION DUE TO INSUFFLATION TUBING DISCONNECTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8456984 ↗
  • MalfunctionMar 25, 2019Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT POTENTIAL LOSS OF INSUFFLATION OCCURED DURING SURGERY, DUE TO INSUFFLATION TUBING DISCONNECTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8450108 ↗
  • MalfunctionJan 31, 2019Disconnection

    IT WAS REPORTED THAT THE INSUFFLATION TUBING DISCONNECTED DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8294484 ↗
  • MalfunctionJan 31, 2019Disconnection

    IT WAS REPORTED THAT THE INSUFFLATION TUBING DISCONNECTED DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8294481 ↗
  • MalfunctionJan 30, 2019Disconnection

    IT WAS REPORTED THAT THE INSUFFLATION TUBING DISCONNECTED DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8289674 ↗
  • MalfunctionJan 30, 2019Disconnection

    IT WAS REPORTED THAT THE INSUFFLATION TUBING DISCONNECTED DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8289673 ↗
  • MalfunctionJan 30, 2019Disconnection

    IT WAS REPORTED THAT THE INSUFFLATION TUBING DISCONNECTED DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8289668 ↗
  • MalfunctionDec 28, 2018Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THERE WAS LOSS OF INSUFFLATION DURING SURGERY DUE TO INSUFFLATION TUBING DISCONNECTING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8203140 ↗
  • MalfunctionDec 26, 2018Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THERE WAS CONTENTIONAL LOSS OF INSUFFLATION DUE TO TUBING DISCONNECTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8197424 ↗
  • MalfunctionDec 21, 2018Contamination /Decontamination Problem

    IT WAS REPORTED THAT THERE WAS FOREIGN MATERIAL INSIDE OF THE STERILE PACKAGING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8190544 ↗
  • MalfunctionDec 5, 2018Gas/Air Leak; Noise, Audible

    WHEN THE INSUFFLATION TUBING WAS INSERTED INTO THE STRYKER INSUFFLATION GENERATOR AN AUDIBLE AIR LEAKING WAS NOTED. THE CLEAR GAS TUBING WAS SLIGHTLY AJAR FROM THE GREY HOUSING; THIS RN WAS IN A CASE EARLIER WHERE THE EXACT NOISE/CIRCUMSTANCE OCCURRED. THE REMEDY FOR BOTH RNS WAS TO FIRMLY INSERT THE CLEAR TUBING FURTHER WITHIN THE GREY HOUSING AND THAT ELIMINATED THE NOISE. THE DOCTOR FELT THE INSUFFLATION WAS CONTINUING EVEN THOUGH THE GENERATOR AND ITS READINGS WERE ACCURATE AND THE MACHINE WAS GIVING TOO MUCH (CAUSING IT TO ERROR/ALARM). THE PATIENT BECAME BRADYCARDIC WITH A HEART RATE OF 22, GAS STOPPED, THE MDA WAS CALLED AND PHARMACOLOGICAL INTERVENTION WAS GIVEN WITH GOOD RESULT. ANOTHER INSUFFLATION GENERATOR WAS OBTAINED AND STAFF RESUMED THE CASE WITH THE NON-HEATED VERSION OF THE PNEUMOSURE. THE DOCTOR REUSED THE PRIOR USED HEATED TUBING WITH THE NEW GENERATOR, WITH A GOOD RESULT. NO LASTING HARM CAME TO THE PATIENT AND THE PROCEDURE WAS COMPLETED AS PLANNED. THE MACHINE WAS INSPECTED BY BIOMED WITH NO ISSUES FOUND AND THE TUBING WAS SENT TO CLINICAL ENGINEERING. THE TUBING WILL BE RETURNED TO THE MANUFACTURER FOR FAILURE ANALYSIS. MANUFACTURER RESPONSE FOR INSUFFLATOR,

    Patient problems: Bradycardia

    MAUDE report 8132451 ↗
  • MalfunctionAug 30, 2018Material Separation

    PATIENT SCHEDULED FOR LAPAROSCOPIC CHOLECYSTECTOMY. WHEN IT WAS TIME FOR THE ABDOMEN TO BE INSUFFLATED WITH CARBON DIOXIDE, THE TEAM NOTED THAT THE ABDOMEN WAS NOT INSUFFLATED. WHEN THEY CHECKED, THE TUBING COVERING THE WIRES WAS NOT IN ITS PLACE. THEY APPEAR TO BE SEPARATED FROM THE GREY COLORED PLASTIC DISC. THE DEVICE WAS CHANGED AND ADEQUATE PNEUMOPERITONEUM WAS OBTAINED. LAP CHOLECYSTECTOMY PROCEEDED WITHOUT ANY MORE UNTOWARD EVENT. PATIENT TOLERATED PROCEDURE WELL WITH NO COMPLICATIONS. MANUFACTURER RESPONSE FOR INSUFFLATOR, LAPAROSCOPIC, PNEUMOSURE (PER SITE REPORTER). SURGERY INVENTORY COORDINATOR TO CALL COMPANY AND REPORT DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7831992 ↗

Other INSUFFLATOR, LAPAROSCOPIC devices

All INSUFFLATOR, LAPAROSCOPIC devices →