NA 7203000000Recalls & reported problems
Stryker Corporation · System 7 Single Trigger Rotary Handpiece · Class 1
Quick summary
This Stryker System 7 rotary handpiece is a powered surgical instrument attachment used during orthopedic procedures. The 28 reports mostly describe the tool staying activated after the trigger was released, parts coming loose and detaching, and the handpiece sticking or resisting during use. None of the reports describe a death or an injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 07613327064841), received between Jul 24, 2018 and Apr 25, 2023 — classified as 28 malfunctions.
The most frequently reported specific device problem is “Device Remains Activated” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Remains Activated8
- Detachment of Device or Device Component7
- Physical Resistance/Sticking4
- Unintended Movement3
- Fail-Safe Problem2
- Fluid/Blood Leak1
- Material Disintegration1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- No Patient Involvement8
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807596 ↗MalfunctionApr 24, 2023Unintended System Motion; Unintended System Motion
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD UNINTENDED SYSTEM MOTION. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16801683 ↗MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261944 ↗MalfunctionJan 29, 2023Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261900 ↗MalfunctionOct 24, 2022Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663607 ↗MalfunctionJul 26, 2022Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15107950 ↗MalfunctionJan 19, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13289831 ↗MalfunctionOct 28, 2021Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12715375 ↗MalfunctionOct 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12680292 ↗MalfunctionJul 21, 2021Unintended Movement; Unintended Movement
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OPERATED IN THE WRONG MODE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12203670 ↗MalfunctionJul 21, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12203663 ↗MalfunctionApr 26, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 5 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11728975 ↗MalfunctionJan 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 5 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11218315 ↗MalfunctionOct 28, 2020Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10749162 ↗MalfunctionJul 27, 2020Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES 0 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10329694 ↗MalfunctionJan 28, 2020Fail-Safe Problem; Fail-Safe Problem
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 9636621 ↗MalfunctionOct 28, 2019Fail-Safe Problem; Fail-Safe Problem
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 9244544 ↗MalfunctionJul 23, 2019Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8816412 ↗MalfunctionApr 24, 2019Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8546402 ↗MalfunctionJan 29, 2019Physical Resistance/Sticking; Physical Resistance/Sticking
THIS REPORT SUMMARIZES <NOE> 3 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A STICKY TRIGGER THAT COULD LEAD TO RUN-ON. 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8285065 ↗MalfunctionJan 29, 2019Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 8284735 ↗MalfunctionOct 29, 2018Unintended Movement; Unintended Movement
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE HAD UNINTENDED MOVEMENT. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8016228 ↗MalfunctionOct 29, 2018Device Remains Activated; Device Remains Activated
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 8015486 ↗MalfunctionOct 29, 2018Physical Resistance/Sticking; Physical Resistance/Sticking
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE TRIGGER WAS STICKING. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 8014683 ↗MalfunctionOct 29, 2018Physical Resistance/Sticking; Physical Resistance/Sticking
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD THE DEVICE HAD A STUCK SAFETY SWITCH, WHICH CAN INDICATE A POTENTIAL FOR UNINTENDED ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 8014418 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.