REUNION 5572-4524Recalls & reported problems
Stryker GmbH · 4.5mm PERIPHERAL SCREW · Class 2 · Implant
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 07613327098730), received between Jul 30, 2018 and Jan 27, 2026 — classified as 20 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
- Break2
- Device Dislodged or Dislocated2
- Fracture2
- Loose or Intermittent Connection2
- Loosening of Implant Not Related to Bone-Ingrowth2
- Loss of Osseointegration2
- Detachment of Device or Device Component1
Patient problems
- Injury4
- Failure of Implant3
- Bone Fracture(s)2
- Extraskeletal ossification2
- Fall2
- Hematoma2
- Implant Pain2
- No Code Available2
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
InjuryJan 27, 2026Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT REVISION SURGERY INVOLVING REMOVAL OF THE REUNION GLENOID COMPONENTS AND IMPLANTATION OF A HUMERAL HEMIARTHROPLASTY HEAD.
Patient problems: Bone Fracture(s); Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 24191006 ↗InjuryAug 19, 2025Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
REVISION OF LEFT SHOULDER. TRAY, LINER, BASEPLATE, GLENOSPHERE AND SCREWS WERE REMOVED, AND ANATOMIC HEAD WAS PLACED ON EXISTING STEM.
Patient problems: Pain
MAUDE report 22843528 ↗InjuryAug 19, 2025Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
TRAY, LINER, BASEPLATE, GLENOSPHERE AND SCREWS WERE REMOVED, AND ANATOMIC HEAD WAS PLACED ON EXISTING STEM. REVISION OF LEFT SHOULDER.
Patient problems: Pain
MAUDE report 22843271 ↗InjuryApr 12, 2024Break; Fracture; Break; Fracture
THERE WAS A REVISION SURGERY DUE TO A FAILED GLENOID IN PREVIOUS YEARS, CAUSED BY A BROKEN CENTER SCREW AND PERIPHERAL SCREWS. THE STRYKER REUNION GLENOSPHERE AND BASEPLATE WITH BROKEN SCREWS WERE REMOVED, AND TORNIER GLENOID COMPONENTS WERE SUCCESSFULLY IMPLANTED. THE REVISION SURGERY WAS SUCCESSFUL WITHOUT ANY ADDITIONAL ADVERSE CONSEQUENCES, AND THERE WAS NO SURGICAL DELAY REPORTED.
Patient problems: Failure of Implant
MAUDE report 19098173 ↗InjuryApr 11, 2024Break; Fracture; Break; Fracture
THERE WAS A REVISION SURGERY DUE TO A FAILED GLENOID IN PREVIOUS YEARS, CAUSED BY A BROKEN CENTER SCREW AND PERIPHERAL SCREWS. THE STRYKER REUNION GLENOSPHERE AND BASEPLATE WITH BROKEN SCREWS WERE REMOVED, AND TORNIER GLENOID COMPONENTS WERE SUCCESSFULLY IMPLANTED. THE REVISION SURGERY WAS SUCCESSFUL WITHOUT ANY ADDITIONAL ADVERSE CONSEQUENCES, AND THERE WAS NO SURGICAL DELAY REPORTED.
Patient problems: Failure of Implant
MAUDE report 19093676 ↗InjuryMar 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT THEY ARE EXPERIENCING SORENESS AND PAIN. ADDITIONALLY, THE PATIENT WANTS TO KNOW IF THEIR IMPLANTS WERE PART OF ANY RECALL.
Patient problems: Discomfort; Implant Pain
MAUDE report 18970005 ↗InjuryJan 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "A POST-MARKET CLINICAL EVALUATION OF THE REUNION REVERSE. SHOULDER ARTHROPLASTY (RSA) SYSTEM: SUBJECT (B)(6). POSTOPERATIVE HETEROTOPIC OSSIFICATION, SYMPTOMATIC. HO OF SUPERIOR/POSTERIOR ASPECTS OF SHOULDER AND INFERIOR GLENOID NECK. SUBJECT COMPLAINS OF OCCASIONAL PAIN."
Patient problems: Extraskeletal ossification; Implant Pain
MAUDE report 18602377 ↗InjuryJan 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "A POST-MARKET CLINICAL EVALUATION OF THE REUNION REVERSE. SHOULDER ARTHROPLASTY (RSA) SYSTEM: SUBJECT (B)(6). POSTOPERATIVE HETEROTOPIC OSSIFICATION, ASYMPTOMATIC. INFERIOR SCAPULAR NECK OSTEOPHYTE, ASYMPTOMATIC"
Patient problems: Extraskeletal ossification
MAUDE report 18601589 ↗InjuryNov 8, 2023Osseointegration Problem; Osseointegration Problem
THE SURGEON SCHEDULED A REVISION SURGERY WITH A STRYKER REPRESENTATIVE. DURING THE SURGERY, IT WAS FOUND THAT THE BASEPLATE/SCREW/GLENOSPHERE CONSTRUCT WAS LOOSE, WITH TWO OF THE FOUR SCREWS NOT CONTAINED WITHIN THE CONSTRUCT. THE GLENOID BASEPLATE/GLENOSPHERE/SCREWS AND THE CUP/POLY CONSTRUCT WERE REMOVED. A 40MM HUMERAL HEAD WAS ATTACHED TO THE EXISTING HUMERAL STEM. THE INDICATION FOR THE PRIMARY PROCEDURE WAS A FRACTURE. THE REVISION SURGERY WAS SUCCESSFUL.
Patient problems: Inadequate Osseointegration; Osteolysis
MAUDE report 18096711 ↗InjuryJul 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
POST-MARKET CLINICAL EVALUATION OF THE REUNION REVERSE SHOULDER ARTHROPLASTY (RSA) SYSTEM: (B)(4). DESCRIPTION OF THE EVENT: ACROMIAL STRESS REACTION ACTIONS TAKEN: SLING FOR TWO WEEKS RESOLVED WITH SEQUELAE: (B)(6) 2021.
Patient problems: Bone Fracture(s)
MAUDE report 17263123 ↗InjuryFeb 23, 2021Detachment of Device or Device Component; Migration; Detachment of Device or Device Component; Migration
AS REPORTED: "THE PATIENT RETURNED TO DR.'S OFFICE FOR A FOLLOW-UP. UPON EXAMINING X-RAYS IT WAS OBVIOUS THAT PERIPHERAL SCREWS HAD MIGRATED THROUGH THE BASEPLATE REQUIRING REVISION SURGERY."
Patient problems: Failure of Implant
MAUDE report 11368863 ↗InjuryDec 1, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "PATIENT EXPERIENCED EXCESS DRAINAGE FROM INCISION SITE ON (B)(6) 2020 THAT PERSISTED OVER 2 DAYS. PATIENT STARTED TAKING CEPHALEXIN 250 MG 4X/DAY ON (B)(6) 2020. PATIENT REPORTED IMPROVEMENT ON (B)(6) 2020. PATIENT WILL CONTINUE TO TAKE MEDICATION FOR TWO WEEKS AND WILL FOLLOW-UP FOR WOUND CHECK. (B)(6) 2020: INCISION BEGAN BLEEDING AGAIN ON (B)(6) 2020. PATIENT SEEN ON (B)(6) 2020, LABS ORDERED ON (B)(6) 2020. DUE TO EXCESS DRAINAGE, PLAN WILL BE REVISION SURGERY."
Patient problems: Hemorrhage/Blood Loss/Bleeding; Post Operative Wound Infection
MAUDE report 10931102 ↗InjuryMar 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "PATIENT RECEIVED A PRIMARY RSA ON (B)(6)2016 , WITH NO ISSUES. PATIENT CAME BACK AND HAD A SHOULDER SCOPE ON(B)(6)2019 , WHICH PRESENTED WHAT APPEARED TO BE VISIBLE METALLOSIS. PATIENT THEN CAME BACK ON (B)(6)2019 TO HAVE HIS RSA REMOVED AND HAVE A SHOULDER SPACER IMPLANTED." ADDITIONALLY REPORTED: "TISSUE WAS SENT TO THE LAB AND METALLOSIS, AS WELL AS P ACNES WERE CONFIRMED. PATIENT CURRENTLY AN ANTIBIOTIC SPACER IMPLANTED WITH GENT/VANC CEMENT AROUND THE HEAD/NECK JUNCTION."
Patient problems: Injury; Reaction
MAUDE report 9862621 ↗InjuryMar 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "PATIENT RECEIVED A PRIMARY RSA ON (B)(6) 2016, WITH NO ISSUES. PATIENT CAME BACK AND HAD A SHOULDER SCOPE ON (B)(6) 2019, WHICH PRESENTED WHAT APPEARED TO BE VISIBLE METALLOSIS. PATIENT THEN CAME BACK ON (B)(6) 2019 TO HAVE HIS RSA REMOVED AND HAVE A SHOULDER SPACER IMPLANTED." ADDITIONALLY REPORTED: "TISSUE WAS SENT TO THE LAB AND METALLOSIS, AS WELL AS P ACNES WERE CONFIRMED. PATIENT CURRENTLY AN ANTIBIOTIC SPACER IMPLANTED WITH GENT/VANC CEMENT AROUND THE HEAD/NECK JUNCTION."
Patient problems: Injury; Reaction
MAUDE report 9862618 ↗InjuryFeb 19, 2020Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT'S LEFT SHOULDER WAS REVISED DUE TO IMPLANT LOOSENING. DUE TO THE PRESENCE OF A HEMATOMA, THE SURGEON REPORTED THAT TRAUMA HAD TO BE A FACTOR. THE PATIENT'S ENTIRE SHOULDER CONSTRUCT EXCEPT FOR THE HUMERAL STEM WAS REVISED. REP REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Hematoma; Injury
MAUDE report 9727814 ↗InjuryFeb 19, 2020Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT'S LEFT SHOULDER WAS REVISED DUE TO IMPLANT LOOSENING. DUE TO THE PRESENCE OF A HEMATOMA, THE SURGEON REPORTED THAT TRAUMA HAD TO BE A FACTOR. THE PATIENT'S ENTIRE SHOULDER CONSTRUCT EXCEPT FOR THE HUMERAL STEM WAS REVISED. REP REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Hematoma; Injury
MAUDE report 9727706 ↗InjuryOct 20, 2019Loose or Intermittent Connection
SUSPECTED GLENOID LOOSENING OF REUNION RSA IMPLANTS LEADING TO REVISION SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9212571 ↗InjuryOct 20, 2019Loose or Intermittent Connection
SUSPECTED GLENOID LOOSENING OF REUNION RSA IMPLANTS LEADING TO REVISION SURGERY.
Patient problems: No Code Available
MAUDE report 9212566 ↗InjuryJul 30, 2018Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT'S RIGHT SHOULDER WAS REVISED DUE TO THE GLENOSPHERE DISENGAGING FROM BASEPLATE. SURGEON SUSPECTED THE GLENOSPHERE MAY HAVE BEEN MALPOSITIONED AT TIME OF IMPLANT. ADDITIONALLY, PATIENT REPORTED HAD EXPERIENCED A FALL IN THE PAST YEAR.
Patient problems: Fall
MAUDE report 7732925 ↗InjuryJul 30, 2018Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT'S RIGHT SHOULDER WAS REVISED DUE TO THE GLENOSPHERE DISENGAGING FROM BASEPLATE. SURGEON SUSPECTED THE GLENOSPHERE MAY HAVE BEEN MALPOSITIONED AT TIME OF IMPLANT. ADDITIONALLY, PATIENT REPORTED HAD EXPERIENCED A FALL IN THE PAST YEAR.
Patient problems: No Code Available
MAUDE report 7732919 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.