REUNION 5901-1184Recalls & reported problems
Stryker GmbH · Glenosphere Jackscrew · Class 2
All models of REUNION (113)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327100358), received between Jul 27, 2020 and Jul 27, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2020Fracture; Fracture
PRIMARY PROCEDURE, RIGHT REVERSE SHOULDER. IT WAS REPORTED THAT AFTER IMPACTING THE GLENOSPHERE ONTO THE BASEPLATE, THE GLENOSPHERE WAS OFF-ANGLE. SURGEON INSERTED THE JACK SCREW TO DISENGAGE THE GLENOSPHERE. AFTER THE SCREW WENT IN DEEPLY, IT SNAPPED. THE PART OF THE SCREW OUTSIDE THE GLENOSPHERE WAS REMOVED. AFTER USING OTHER MEANS TO ATTEMPT TO DISENGAGE THE GLENOSPHERE, THE SURGEON REPORTED THE GLENOSPHERE WAS FIRMLY FIXED TO THE BASEPLATE AND USED AN IMPACTOR TO ADJUST THE GLENOID TO A BETTER ANGLE. AFTER CHANGING THE ANGULATION OF THE GLENOID, THE SURGEON COMPLETED THE PROCEDURE AND WAS SATISFIED WITH THE RESULT. SURGERY COMPLETED SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 10 MINUTES.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 10329742 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page