TARGET M0036181650Recalls & reported problems

Stryker Corporation · XXL 360 Detachable Coil · Class 2 · Implant

All models of TARGET (114)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327128352), received between May 18, 2018 and Sep 3, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Premature Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Separation3
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2026Migration

    It was reported that during procedure, the microcatheter was withdrawn at the conclusion of the procedure; however, the subject coil was inadvertently pulled out together with the microcatheter. The procedure was completed successfully. No clinical consequences were reported to the patient due to this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50577640 ↗
  • MalfunctionJan 3, 2024Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING ANEURYSM EMBOLIZATION PROCEDURE THE SUBJECT COIL SPONTANEOUSLY DETACHED. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18436673 ↗
  • MalfunctionJan 30, 2023Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE COIL (SUBJECT DEVICE) PREMATURELY DETACHED FROM THE DELIVERY WIRE IN THE INTERNAL CAROTID ARTERY (ICA) WHILE BEING DEPLOYED. THE DELIVERY WIRE WAS REMOVED FROM THE PATIENTS BODY. THE COIL WAS NOT FULLY DEPLOYED IN THE ANEURYSM SO THE PHYSICIAN USED A STENT TO TACK DOWN THE COIL TAIL TO KEEP IT IN PLACE AND TO COMPLETE THE PROCEDURE SUCCESSFULLY WITH NO CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16268569 ↗
  • InjuryMay 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING COIL INSERTION WITH CONSIDERABLE SPEED IN AND OUT CAUSED THE COIL (SUBJECT DEVICE) TO STRETCH AT DISTAL. AFTER THAT, SUBSEQUENT ATTEMPTS WERE MADE TO RETRIEVE THE COIL (SUBJECT DEVICE); HOWEVER, IT WAS DIFFICULT. A SNARE AND FORCEPS TYPE WAS USED TO RETRIEVE THE COIL (SUBJECT DEVICE). EVENTUALLY, IT WAS RETRIEVED WITHOUT REMAINING IN THE BODY, BUT A THROMBUS WAS CONFIRMED IN THE INGUINAL REGION. THE CAUSAL RELATIONSHIP BETWEEN THE COIL (SUBJECT DEVICE) AND THE THROMBUS WAS UNKNOWN. THE PATIENT WAS RE-TREATED AT A LATER DATE. THE THROMBUS WAS SUCCESSFULLY TREATED AND THE RE-TREATMENT WAS COMPLETED. THERE WAS A SURGICAL DELAY FOR FOUR HOURS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 14379962 ↗
  • MalfunctionMay 7, 2022Premature Separation; Premature Separation

    IT WAS REPORTED THAT DURING EMBOLIZATION OF TYPE I ENDOLEAK AFTER TEVER, THE COIL (SUBJECT DEVICE) WAS PREMATURELY DETACHED DURING PLACEMENT. THE DETACHED COIL (SUBJECT DEVICE) WAS PUSHED INTO THE MASS BY THE DELIVERY WIRE AND PLACED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14318139 ↗
  • MalfunctionMay 18, 2018Break; Break

    IT WAS REPORTED THAT DURING COIL EMBOLIZATION PROCEDURE, THE FIRST COIL DETACHED IN THE MICROCATHETER INSIDE THE PATIENT. THE CUSTOMER DID NOT MAINTAIN CONTINUOUS FLUSH AS RECOMMENDED PER DFU. THE PHYSICIAN DID NOT MAKE ANY ATTEMPTS TO DETACH THE COIL USING THE DETACHMENT SYSTEM. THE PHYSICIAN REMOVED THE DETACHED COIL TO CONTINUE WITH THE PROCEDURE. A SNARE DEVICE WAS NOT USED WHEN THE SUBJECT COIL WAS REMOVED. SIMILAR EVENT OCCURRED TO ANOTHER COIL (THE SECOND REPORTED SUBJECT DEVICE) USED IN THE PROCEDURE. NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7527752 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →