NA 78-40030Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · PEEK Customized Cranial Implant, L · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327129298), received between Sep 5, 2018 and Jan 15, 2026 — classified as 3 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Inadequacy of Device Shape and/or Size7
- Fitting Problem2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information2
- Bacterial Infection1
- Cerebral Edema1
- No Known Impact Or Consequence To Patient1
- Wound Dehiscence1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2026Inadequacy of Device Shape and/or Size
IMPLANT DID NOT FIT THE PATIENT IN SURGERY. SURGEON USED TITANIUM MESH AS ALTERNATIVE TO THE PEEK IMPLANT. FITTING ISSUE CAUSED AN SURGICAL DELAY OF 16-60 MIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24077874 ↗MalfunctionApr 24, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE IMPLANT REQUIRED MODERATE ADJUSTMENT DUE TO SOFT TISSUE INTERFERENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21900419 ↗MalfunctionApr 24, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE IMPLANT REQUIRED MODERATE ADJUSTMENT DUE TO CEREBRAL SWELLING, SOFT TISSUE INTERFERENCE AND CHANGES OF ANATOMY/CALCIFICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21900208 ↗MalfunctionJan 21, 2025Inadequacy of Device Shape and/or Size; Fitting Problem; Inadequacy of Device Shape and/or Size; Fitting Problem
IT WAS REPORTED THAT THE IMPLANT DID NOT FIT AS PLANNED, AND IT WAS MODIFIED DURING SURGERY; NEVERTHELESS, POST-OP THERE WAS AN AESTHETIC DEFECT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21196956 ↗MalfunctionDec 17, 2024Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THE IMPLANT DID NOT FIT WELL.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20955034 ↗InjuryApr 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THERE WAS A REVISION SURGERY PERFORMED TO REVISE THE IMPLANT DUE TO PATIENT PRESENTING WITH A WOUND DEHISCENCE.
Patient problems: Bacterial Infection; Wound Dehiscence
MAUDE report 19082144 ↗InjuryMar 20, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THE SHAPE OF THE IMPLANT DID NOT FIT AS EXPECTED AND COULD NOT BE USED TO COMPLETE THE SURGERY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18943923 ↗InjuryMar 31, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT DURING THE SURGERY, IT WAS NOTICED THE PEEK IMPLANT WAS TOO SMALL AND NOT CURVED AS EXPECTED. AN ALTERNATIVE IMPLANT WAS USED. A REVISION SURGERY WAS SUCCESSFULLY PERFORMED TO IMPLANT A NEW PEEK IMPLANT.
Patient problems: Cerebral Edema; No Clinical Signs, Symptoms or Conditions
MAUDE report 13971162 ↗MalfunctionSep 5, 2018Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THERE WAS A DISCREPANCY BETWEEN THE IMPLANT AND THE CRANIAL SHAPE, THIS IMPLANT WAS NOT USED. AN ALTERNATIVE IMPLANT WAS USED. THE SURGERY WAS DELAYED FOR AN HOUR.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7846170 ↗
FDA clearance and classification
Cleared through 510(k) K203055 on Feb 9, 2021 (decision: Substantially Equivalent). Submitted by Stryker as “Stryker PEEK Customized Cranial Implant Kit”. The FDA review took 124 days. FDA record ↗
Cleared through 510(k) K190229 on Jul 11, 2019 (decision: Substantially Equivalent). Submitted by Stryker as “Stryker PEEK Customized Cranial Implant Kit”. The FDA review took 156 days. FDA record ↗
Cleared through Special 510(k) K152076 on Aug 26, 2015 (decision: Substantially Equivalent). Submitted by Stryker as “PEEK Customized Cranial Implant, PEEK Customized Cranial Implant Priority”. The FDA review took 30 days. FDA record ↗
The FDA regulates product code GWO (plate, cranioplasty, preformed, alterable) under 21 CFR 882.5320 as a Class II device. 21 CFR 882.5320 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 78-40030
Has the NA 78-40030 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613327129298), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 78-40030?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Sep 5, 2018 and Jan 15, 2026. In those reports, the most common specific device problems are “Inadequacy of Device Shape and/or Size” (7) and “Fitting Problem” (2); the most common patient problems are “Bacterial Infection” (1), “Cerebral Edema” (1) and “Wound Dehiscence” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 78-40030?
None of the reports are classified as deaths, 3 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 78-40030 FDA approved?
It was cleared through Special 510(k) K152076 on Aug 26, 2015; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 78-40030 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the NA 78-40030, and is it still sold?
The FDA's device database lists Stryker Leibinger GmbH & Co. KG as the company responsible for it (brand NA, model 78-40030). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Plate, cranioplasty, preformed, alterable devices
- OsteoMed 218-1604Osteomed LLC · Plate, cranioplasty, preformed, alterable1 recall6 reports
- OsteoMed 218-1604-SP-10Osteomed LLC · Plate, cranioplasty, preformed, alterable1 recall0 reports
- NA 56-90614Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable1 recall0 reports
- NA 78-50030Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable21 reports
- THINFLAP SYSTEM 91-6704Biomet Microfixation, Inc · Plate, cranioplasty, preformed, alterable19 reports
- NA 78-50040Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable19 reports
- NA 78-10030Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable13 reports
- NA 78-40040Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page