OsteoMed 218-1604Recalls & reported problems

Osteomed LLC · 1.6mm X 4mm Profile Zero Autodrive Screw · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00845694034258), received between Jun 9, 2022 and Feb 12, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 31, 2023Open, Classified

    Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.

    Z-2107-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device-Device Incompatibility5
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED IN AN EMAIL FROM THE PATIENT, "I HAD SURGERY USING YOUR PARTS, RECALL Z-2107-2023. MY SKULL PLATELETS HAVE BEEN MOVING AROUND IN PIECES FOR YEARS AND IF I HAD KNOWN TO LET MY DOCTOR KNOW I COULD HAVE GOTTEN THEM REPLACED. I THOUGHT THAT IT WAS NORMAL FOR THEM TO NOT STICK TOGETHER. I'VE MADE AN APPOINTMENT WITH THE SURGEON TO GET THEM REPLACED AND I'M LOOKING INTO WHAT I CAN DO TO GET COMPENSATED FOR THE PAIN, MIGRAINES AND SEIZURES THAT I'VE BEEN HAVING DUE TO THIS."

    Patient problems: Pain

    MAUDE report 24328863 ↗
  • MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW SLIPPED RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE LIFTED UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE (5) 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00032, 2207754-2022-00033, 2207754-2022-00034, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640798 ↗
  • MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00032, 2207754-2022-00033, AND 2207754-2022-00034 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640757 ↗
  • MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE (5) 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00033, 2207754-2022-00034, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640754 ↗
  • MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE (5) 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00032, 2207754-2022-00033, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640744 ↗
  • MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00032, 2207754-2022-00034, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640739 ↗

Other Plate, cranioplasty, preformed, alterable devices

All Plate, cranioplasty, preformed, alterable devices →