OsteoMed 218-1604Recalls & reported problems
Osteomed LLC · 1.6mm X 4mm Profile Zero Autodrive Screw · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00845694034258), received between Jun 9, 2022 and Feb 12, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 31, 2023Open, Classified
Screws for osteotomies have contain an undersize screwhead which has the potential to pass through the plate hole.
Z-2107-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Device-Device Incompatibility5
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED IN AN EMAIL FROM THE PATIENT, "I HAD SURGERY USING YOUR PARTS, RECALL Z-2107-2023. MY SKULL PLATELETS HAVE BEEN MOVING AROUND IN PIECES FOR YEARS AND IF I HAD KNOWN TO LET MY DOCTOR KNOW I COULD HAVE GOTTEN THEM REPLACED. I THOUGHT THAT IT WAS NORMAL FOR THEM TO NOT STICK TOGETHER. I'VE MADE AN APPOINTMENT WITH THE SURGEON TO GET THEM REPLACED AND I'M LOOKING INTO WHAT I CAN DO TO GET COMPENSATED FOR THE PAIN, MIGRAINES AND SEIZURES THAT I'VE BEEN HAVING DUE TO THIS."
Patient problems: Pain
MAUDE report 24328863 ↗MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW SLIPPED RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE LIFTED UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE (5) 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00032, 2207754-2022-00033, 2207754-2022-00034, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14640798 ↗MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00032, 2207754-2022-00033, AND 2207754-2022-00034 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14640757 ↗MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE (5) 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00033, 2207754-2022-00034, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14640754 ↗MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE (5) 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00032, 2207754-2022-00033, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14640744 ↗MalfunctionJun 9, 2022Device-Device Incompatibility; Device-Device Incompatibility
DURING THE CRANIOPLASTY PROCEDURE ON (B)(6) 2022, THE PHYSICIAN WAS IMPLANTING SCREWS (PART NUMBER 218-1604) INTO A PROFILE ZERO SQUARE PLATE. WHEN THE SCREW HEAD INTERFACED WITH THE PLATE, THE SCREW "SLIPPED" RIGHT THROUGH THE PLATE HOLE AND INTO THE BONE. IT WAS REPORTED IT APPEARED AS IF THE SCREW HEADS WERE SMALLER THAN THE PLATE HOLES. THIS ISSUE OCCURRED WITH SEVERAL SCREWS AS A PLATE THE SURGEON THOUGHT WAS SECURE "LIFTED" UP LEAVING THE SCREWS IN THE BONE AND UNATTACHED TO THE PLATE. THE 218-1604 SCREWS WERE REMOVED AND REPLACED WITH PART NUMBER 218-1605 SCREWS WHICH SECURED THE PLATE FLUSH TO THE BONE WITH NO ISSUES. IT WAS REPORTED THE FIVE 218-1604 SCREWS INVOLVED IN THIS EVENT ARE BEING RETURNED FOR EVALUATION. THIS REPORT IS RELATED TO REPORT NUMBERS 2207754-2022-00031, 2207754-2022-00032, 2207754-2022-00034, AND 2207754-2022-00035 FOR THE OTHER SCREWS INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14640739 ↗
Other Plate, cranioplasty, preformed, alterable devices
- OsteoMed 218-1604-SP-10Osteomed LLC · Plate, cranioplasty, preformed, alterable1 recall0 reports
- NA 56-90614Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable1 recall0 reports
- NA 78-50030Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable21 reports
- THINFLAP SYSTEM 91-6704Biomet Microfixation, Inc · Plate, cranioplasty, preformed, alterable19 reports
- NA 78-50040Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable18 reports
- NA 78-10030Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable13 reports
- NA 78-40040Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable11 reports
- NA 78-40030Stryker Leibinger GmbH & Co. KG · Plate, cranioplasty, preformed, alterable9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page