Precision Falcon 6720127105Recalls & reported problems
Stryker Corporation · Oscillating Tip Saw Cartridge · Class 1
All models of Precision Falcon (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 07613327143737), received between Feb 24, 2020 and Jul 22, 2026 — classified as 1 injury, 17 malfunctions.
The most frequently reported specific device problem is “Fracture” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture18
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- Device Embedded In Tissue or Plaque3
- Insufficient Information3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25907839 ↗MalfunctionJan 23, 2026Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT. - 2 EVENTS HAD INSUFFICIENT INFORMATION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 24163427 ↗MalfunctionOct 28, 2025Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FRACTURE, 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399873 ↗MalfunctionAug 13, 2025Fracture; Fracture
MalfunctionJul 29, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: FRACTURE. 5 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions
MAUDE report 22648105 ↗MalfunctionOct 21, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20501105 ↗MalfunctionJan 22, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18556982 ↗MalfunctionOct 23, 2023Fracture; Fracture
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD A PROLONGED SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17992800 ↗MalfunctionFeb 9, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A DISTAL FEMORAL OSTEOTOMY PROCEDURE, THE BLADE TIP BROKE FROM THE CARTRIDGE. IT WAS ALSO REPORTED THAT AN X-RAY WAS PERFORMED TO LOCATE THE BROKEN BLADE FRAGMENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16342373 ↗InjuryDec 13, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A FEMORAL AND TIBIAL OSTEOTOMY PROCEDURE, THE BLADE BROKE. IT WAS SUBSEQUENTLY REPORTED THAT THE SAW BLADE WAS STUCK INSIDE THE PATIENT'S BONE AND WAS SUCCESSFULLY RETRIEVED WITH NO ADVERSE CONSEQUENCES. IT WAS FURTHER REPORTED THAT THERE WAS A TEN MINUTE DELAY AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15968040 ↗MalfunctionNov 4, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A HTO PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15737484 ↗MalfunctionMay 16, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A PERIACETABULAR OSTEOTOMY SURGERY, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE OF THE BLADE WAS RETRIEVED FROM THE PATIENT. IT WAS REPORTED THAT THE PROCEDURE WAS DELAYED FOR 2 HOURS TO RETRIEVE THE BROKEN PIECE. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14409395 ↗MalfunctionApr 27, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THE SURGEON DECIDED TO LEAVE THE MINIMAL FRAGMENTS OF THE BLADE IN THE SURGICAL SITE. IT WAS FURTHER REPORTED THAT THERE WERE NO DELAYS OR MEDICAL INTERVENTION AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 14225512 ↗MalfunctionApr 12, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BLADE BROKE AT THE CUTTING END. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14087582 ↗MalfunctionJun 25, 2021Fracture; Fracture
IT WAS REPORTED THAT THE PRECISION CARTRIDGE BROKE DURING A TOTAL KNEE ARTHROPLASTY PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12066854 ↗MalfunctionJun 4, 2021Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA PROCEDURE, THE BLADE BROKE AT THE CUTTING END. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11938198 ↗MalfunctionAug 6, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. IT WAS FURTHER REPORTED THAT A DELAY OF 5 MINUTES OCCURRED DUE TO THIS EVENT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10373850 ↗MalfunctionFeb 24, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WAS A SHORT DELAY DUE TO AN X- RAY WHICH CONFIRMED NO BLADE FRAGMENTS IN THE PATIENT. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9745652 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page