Precision Falcon 6725127090Recalls & reported problems
Stryker Corporation · Oscillating Tip Saw Cartridge · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613327143751), received between Jul 24, 2018 and Jul 22, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Fracture” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture8
- Break4
- Tear, Rip or Hole in Device Packaging2
- Delivered as Unsterile Product1
- Device Contaminated During Manufacture or Shipping1
- Material Split, Cut or Torn1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient6
- No Patient Involvement1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25907818 ↗MalfunctionJan 23, 2026Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163577 ↗MalfunctionJul 24, 2024Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping; Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY CONTAMINATED DURING MANUFACTURE OR SHIPPING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19818511 ↗MalfunctionMar 3, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT BEFORE A MEDICAL PROCEDURE, IT WAS NOTICED THAT THERE WAS A SMALL PUNCTURE HOLE IN THE PACKAGING OF THE BLADE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS IS REPORT 1 OF 2 FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16477797 ↗MalfunctionMar 3, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT BEFORE A MEDICAL PROCEDURE, IT WAS NOTICED THAT THERE WAS A SMALL PUNCTURE HOLE IN THE PACKAGING OF THE BLADE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS IS REPORT 2 OF 2 FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16477796 ↗MalfunctionAug 24, 2021Fracture; Fracture
IT WAS REPORTED THAT POST OPERATIVELY, IT WAS NOTED THAT THE BLADE BROKE DURING ITS REMOVAL FROM THE HANDPIECE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12359679 ↗MalfunctionJun 23, 2021Fracture; Material Split, Cut or Torn; Fracture; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A RIGHT TOTAL KNEE ARTHROPLASTY PROCEDURE, THE BLADE BROKE. IT WAS FURTHER REPORTED THAT THE BROKEN PIECES WERE SUCCESSFULLY REMOVED FROM THE PATIENT AND THERE WAS A FIVE MINUTE DELAY AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. IT WAS SUBSEQUENTLY DETERMINED BASED ON THE EVALUATION OF THE RETURNED DEVICE THAT THE BLADE DID NOT BREAK, THE CARTRIDGE SPLIT OPEN AT THE SIDE WELDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12052173 ↗MalfunctionOct 23, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE TIP OF THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10726967 ↗MalfunctionAug 6, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING SURGERY, THE TIP OF THE BLADE BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10373956 ↗MalfunctionFeb 11, 2020Vibration; Vibration
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BLADE WAS VIBRATING AND SHAKING UNCONTROLLABLY IN THE HANDPIECE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9694378 ↗MalfunctionOct 22, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A TOTAL HIP SURGICAL PROCEDURE, IT WAS NOTED THAT THE BLADE BROKE HALF WAY ON THE BLADE AND ONE BLADE TOOTH ALSO BROKE. IT WAS ALSO REPORTED THAT ALL PIECES OF THE BROKEN BLADE WERE RETRIEVED. IT WAS FURTHER REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9220418 ↗MalfunctionJul 9, 2019Fracture; Fracture
IT WAS REPORTED THAT THE BLADE BROKE DURING FEMORAL SURGERY. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8771188 ↗MalfunctionJan 29, 2019Break; Break
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT IN WHICH THE DEVICE BROKE DURING AN ANTERIOR TOTAL HIP ARTHROPLASTY PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Patient Involvement
MAUDE report 8286854 ↗MalfunctionOct 24, 2018Break; Break
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICES BROKE DURING A TOTAL KNEE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8001663 ↗MalfunctionJul 24, 2018Break; Break
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE BROKE DURING A TOTAL KNEE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7714603 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page