Precision Falcon 6720127090Recalls & reported problems
Stryker Corporation · Oscillating Tip Saw Cartridge · Class 1
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07613327143768), received between Nov 1, 2018 and Jan 23, 2026 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Fracture” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture7
- Break4
- Adverse Event Without Identified Device or Use Problem1
- Delivered as Unsterile Product1
- Device Contaminated During Manufacture or Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient4
- Laceration(s)1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD CODE NOT FOUND!!.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163388 ↗MalfunctionJan 21, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21200755 ↗MalfunctionOct 22, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501673 ↗MalfunctionJul 24, 2024Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping; Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19818514 ↗MalfunctionOct 19, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A HIGH TIBIAL OSTEOTOMY (HTO) PROCEDURE, THE BLADE BROKE AT THE TIP. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15630510 ↗MalfunctionAug 16, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A TOTAL KNEE ARTHROPLASTY PROCEDURE, THE BLADE BROKE AT THE CUTTING END. IT WAS ALSO REPORTED THAT THERE WAS A FIVE MINUTE DELAY AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15237252 ↗MalfunctionApr 11, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A TOTAL KNEE PROCEDURE, THE BLADE BROKE DURING THE FEMORAL CUT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14075353 ↗MalfunctionAug 2, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE BLADE BROKE AT THE CUTTING END DURING A TOTAL KNEE ARTHROPLASTY SURGICAL PROCEDURE. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A SPARE BLADE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8854832 ↗MalfunctionJun 11, 2019Break; Break
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT A SINGLE BLADE BROKE WHERE IT FITS INTO THE HANDPIECE. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE EVENT CAUSED A 5 MINUTE DELAY TO THE PROCEDURE. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8686616 ↗InjuryMar 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SURGEON OBSERVED POSTOPERATIVELY DURING TOURNIQUET REMOVAL THAT THE PATIENTS POPLITEAL ARTERY WAS CUT DURING A KNEE REPLACEMENT SURGERY. IT WAS ALSO REPORTED THAT THE PATIENT UNDERWENT VASCULAR SURGERY TO REPAIR THE ARTERY.
Patient problems: Laceration(s)
MAUDE report 8453340 ↗MalfunctionMar 4, 2019Break; Break
IT WAS REPORTED THAT DURING A TKA (TOTAL KNEE ARTHROPLASTY) PROCEDURE, THE BLADE BROKE AT THE CUTTING END. IT WAS FURTHER REPORTED THAT THERE WAS A TWO MINUTE DELAY AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8387338 ↗MalfunctionNov 1, 2018Break; Break
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE THE BLADE BROKE. IT WAS ALSO REPORTED THAT THE BLADE FELL INTO THE PATIENT BUT WAS RETRIEVED SUCCESSFULLY. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES OR DELAYS AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8027367 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.