GoBed FL28CRecalls & reported problems

Stryker Corporation · Class 2

All models of GoBed (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Stryker GoBed hospital bed has 133 reports, many describing mechanical or electrical faults such as unexpected movement, sudden drops and weak brakes. The reports mention no deaths and 1 injury. There is 1 open recall, started in May 2021, because beds meant for healthcare settings were sent to homes, where users may not fully understand the instructions. Reports fell to 1 in the last 12 months from 8 in the previous 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
133
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 07613327169355), received between Apr 24, 2018 and Jan 26, 2026 — classified as 1 injury, 132 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (43 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2021Class IIOpen, Classified

    Medical devices intended for use in a healthcare facility were distributed to residential customers and may be in use in a home environment. The product s labeling accurately identifies the correct use environment, but home users may not be aware of or fully understand these instructions intended for healthcare professionals

    Z-2223-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect Measurement43
  • Mechanical Problem29
  • Electrical /Electronic Property Problem22
  • Unintended System Motion17
  • Device Fell12
  • No Audible Alarm5
  • Sharp Edges4
  • Positioning Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions88
  • No Patient Involvement41
  • Fall3
  • No Consequences Or Impact To Patient3
  • Bruise/Contusion2
  • Laceration(s)2
  • Bone Fracture(s)1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2026Device Fell

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24183708 ↗
  • MalfunctionOct 24, 2025Sharp Edges; Sharp Edges

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23384159 ↗
  • MalfunctionApr 29, 2025Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925716 ↗
  • MalfunctionApr 29, 2025Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925680 ↗
  • MalfunctionApr 29, 2025Unintended System Motion; Unintended System Motion

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925653 ↗
  • MalfunctionApr 29, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925619 ↗
  • MalfunctionJan 30, 2025Device Fell; Device Fell

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED LIFT COLLAPSE. ONE EVENT WAS ALLEGED TO HAVE RESULTED IN A MINOR UNSPECIFIED INJURY TO THE PATIENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21277554 ↗
  • MalfunctionJan 29, 2025Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21266622 ↗
  • MalfunctionJan 29, 2025Unintended System Motion; Unintended System Motion

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED UN-COMMANDED MOTION OF THE BED POSITIONING SYSTEMS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21266598 ↗
  • MalfunctionOct 29, 2024Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; THE PATIENT EXPERIENCED A FALL AND LACERATION.

    Patient problems: Fall; Laceration(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 20560253 ↗
  • MalfunctionOct 29, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20560250 ↗
  • MalfunctionOct 29, 2024Device Fell; Device Fell

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20560237 ↗
  • MalfunctionJul 29, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19850316 ↗
  • MalfunctionApr 23, 2024Unintended System Motion; Unintended System Motion

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162764 ↗
  • MalfunctionApr 23, 2024Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162691 ↗
  • MalfunctionApr 23, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162689 ↗
  • MalfunctionApr 23, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162688 ↗
  • MalfunctionApr 23, 2024Device Fell; Device Fell

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162682 ↗
  • MalfunctionJan 24, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18574344 ↗
  • MalfunctionJan 24, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18573511 ↗
  • MalfunctionOct 25, 2023Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED REDUCED/INADEQUATE BRAKE FORC. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18004940 ↗
  • MalfunctionOct 23, 2023Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATELY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17990273 ↗
  • MalfunctionOct 23, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17990271 ↗
  • MalfunctionJul 21, 2023Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17372733 ↗
  • MalfunctionJul 21, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17372727 ↗

FDA clearance and classification

The FDA regulates product code FNL (bed, ac-powered adjustable hospital) under 21 CFR 880.5100 as a Class II device. 21 CFR 880.5100 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the GoBed FL28C

Has the GoBed FL28C been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2223-2021 (Class II), began on May 12, 2021: Medical devices intended for use in a healthcare facility were distributed to residential customers and may be in use in a home environment. The product s labeling accurately identifies the correct use environment,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the GoBed FL28C?

The FDA's MAUDE database lists 133 adverse event reports for this device, received between Apr 24, 2018 and Jan 26, 2026. In those reports, the most common specific device problems are “Incorrect Measurement” (43), “Mechanical Problem” (29) and “Electrical /Electronic Property Problem” (22); the most common patient problems are “Fall” (3), “Bruise/Contusion” (2) and “Laceration(s)” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the GoBed FL28C?

None of the reports are classified as deaths, 1 as an injury and 132 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the GoBed FL28C FDA approved?

Devices of this type (product code FNL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.

Who makes the GoBed FL28C, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand GoBed, model FL28C). Its record shows it as no longer in commercial distribution.