GoBed FL28EXRecalls & reported problems
Stryker Corporation · Class 2
Quick summary
This Stryker GoBed is an electric adjustable hospital bed. Many of the 181 reports describe electrical and mechanical failures, and some describe staff not being alerted to a possible patient fall. The reports include 176 malfunctions, 5 injuries and no deaths. Reports from the last 12 months number 3, compared with 16 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 181
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 181 adverse event reports that list this exact device (DI 07613327169386), received between Mar 27, 2018 and Apr 23, 2026 — classified as 5 injuries, 176 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (43 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement43
- Mechanical Problem38
- Electrical /Electronic Property Problem29
- Unintended System Motion21
- Sharp Edges15
- Device Fell14
- No Audible Alarm14
- Positioning Failure4
Patient problems
- No Clinical Signs, Symptoms or Conditions118
- No Patient Involvement52
- Fall7
- Pain5
- No Known Impact Or Consequence To Patient4
- No Consequences Or Impact To Patient3
- Abrasion1
- Bone Fracture(s)1
Recent adverse event reports
The 25 most recent of 181 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026No Audible Alarm
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) - (B)(6) 2026. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24965126 ↗MalfunctionApr 23, 2026Sharp Edges
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24965123 ↗MalfunctionJan 26, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED LITTER/LIFT UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24183710 ↗MalfunctionOct 27, 2025No Audible Alarm; No Audible Alarm
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23391964 ↗MalfunctionOct 24, 2025Device Fell; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED LITTER/LIFT UNEXPECTED DROP/COLLAPSE. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT THAT EXPERIENCED PAIN.
Patient problems: Pain
MAUDE report 23384160 ↗MalfunctionJul 25, 2025No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608961 ↗MalfunctionJul 25, 2025Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608288 ↗MalfunctionApr 29, 2025No Audible Alarm; No Audible Alarm
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925681 ↗MalfunctionApr 29, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925663 ↗MalfunctionApr 29, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925643 ↗MalfunctionApr 29, 2025Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925637 ↗MalfunctionApr 28, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21920013 ↗MalfunctionApr 28, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21920006 ↗MalfunctionApr 28, 2025Unintended System Motion; Unintended System Motion
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919964 ↗MalfunctionJan 30, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21277793 ↗MalfunctionJan 30, 2025No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED A FAILURE TO ALARM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21277560 ↗MalfunctionJan 30, 2025Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21277556 ↗MalfunctionJan 29, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21266636 ↗MalfunctionJan 29, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 12 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INACCURATE SCALE READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21266618 ↗MalfunctionOct 29, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560508 ↗MalfunctionOct 29, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560447 ↗MalfunctionOct 29, 2024Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560445 ↗MalfunctionOct 29, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED LITTER/LIFT UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560259 ↗MalfunctionOct 29, 2024No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560257 ↗MalfunctionOct 29, 2024Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560240 ↗
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,937 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,377 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,193 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death374 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths362 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths358 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL317 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.