Skip to content

Performance-PRO 6086Recalls & reported problems

Stryker Corporation · Class 2

Quick summary

This Stryker Performance-PRO 6086 is a wheeled stretcher (cot) used to move and load patients into ambulances. The 34 reports mostly describe difficulty loading or unloading the cot into an ambulance, positioning failures, the device tipping over, and sharp metal edges;31 reports describe malfunctions and 3 involve injuries, with no deaths. No recalls name this device, and the number of reports in the last 12 months is about the same as the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 07613327261554), received between Apr 24, 2018 and Apr 23, 2026 — classified as 3 injuries, 31 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Positioning” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Positioning14
  • Positioning Failure5
  • Device Fell4
  • Sharp Edges3
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment Of Device Component1
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Patient Involvement14
  • Injury2
  • Crushing Injury1
  • Laceration(s)1
  • Neck Pain1
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Difficult or Delayed Positioning

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24967205 ↗
  • MalfunctionApr 28, 2025Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21921708 ↗
  • MalfunctionApr 23, 2024Difficult or Delayed Positioning; Sharp Edges; Difficult or Delayed Positioning; Sharp Edges

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE AND ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162958 ↗
  • MalfunctionJan 24, 2024Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18573185 ↗
  • MalfunctionOct 24, 2023Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED COT HEIGHT CANNOT BE ADJUSTE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17999058 ↗
  • MalfunctionJul 18, 2023Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULTY LOADING AND UNLOADING TO COT INTO THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17343975 ↗
  • MalfunctionJul 18, 2023Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED THE FOWLER BEING STUCK ELEVATED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17343686 ↗
  • MalfunctionJan 19, 2023Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DIFFICULTY LOADING AND UNLOADING THE COT FROM THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16207147 ↗
  • MalfunctionOct 20, 2022Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED DIFFICULTY LOADING/UNLOADING THE COT IN THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15640443 ↗
  • MalfunctionJul 26, 2022Device Tipped Over; Device Tipped Over

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED A COT TIP. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; NO ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15104394 ↗
  • MalfunctionApr 25, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED THE COT FALSELY ENGAGES INTO THE FASTENER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193747 ↗
  • MalfunctionOct 25, 2021Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED THE FOWLER STUCK IN A RAISED POSITION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12693532 ↗
  • MalfunctionOct 25, 2021Sharp Edges; Sharp Edges

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12693363 ↗
  • MalfunctionJul 27, 2021Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED A FOWLER STUCK RAISED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12233883 ↗
  • MalfunctionApr 23, 2021Sharp Edges; Sharp Edges

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THERE WERE ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11715683 ↗
  • MalfunctionJul 27, 2020Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT, WHERE IT WAS REPORTED THE COT WAS DIFFICULT TO LOAD/UNLOAD FROM THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 10325642 ↗
  • MalfunctionApr 28, 2020Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT, WHERE IT WAS REPORTED THE COT WAS DIFFICULT TO LOAD/UNLOAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 10005535 ↗
  • MalfunctionApr 28, 2020Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT, WHERE IT WAS REPORTED THE COT HEIGHT COULD NOT BE ADJUSTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 10005526 ↗
  • MalfunctionApr 28, 2020Device Fell; Device Fell

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT, WHERE IT WAS REPORTED THE COT DROPPED.  THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 10005515 ↗
  • MalfunctionJan 27, 2020Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT, WHERE IT WAS REPORTED THE UNIT WAS DIFFICULT TO LOAD/UNLOAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9633978 ↗
  • MalfunctionOct 25, 2019Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES <NOE> 3 </NOE> MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE COT WAS DIFFICULT TO LOAD OR UNLOAD INTO THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9237935 ↗
  • MalfunctionOct 25, 2019Device Fell; Device Fell

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE COT DROPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9237356 ↗
  • MalfunctionJul 29, 2019Device Fell; Device Fell

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT, WHERE IT WAS REPORTED THE FOWLER UNEXPECTEDLY DROPPED. THERE WAS PATIENT INVOLVEMENT, HOWEVER THERE WERE NO CONSEQUENCES OR IMPACTS TO THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8835835 ↗
  • MalfunctionJul 29, 2019Difficult or Delayed Positioning; Difficult or Delayed Positioning

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED IT WAS DIFFICULT TO LOAD/UNLOAD THE COT IN THE AMBULANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8835825 ↗
  • InjuryJul 12, 2019Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE CUSTOMER THAT TWO MIDWIVES, BOTH SLIGHT IN BUILD, WERE PLACING THE PATIENT ONTO THE COT AND ATTEMPTING TO RAISE THE COT BACK UP TO THE HEIGHT OF A BED FOR THE PATIENT TO BE READMITTED TO THE MATERNITY BED. HOWEVER IN DOING SO, IT WAS ALLEGED THE TWO MIDWIVES SUSTAINED BACK INJURIES DUE TO THE MANUAL HANDLING NATURE OF THE COT AND THE FACT THE PATIENT WAS ON THE COT AT THE TIME. THE STAFF WERE UNABLE TO RAISE THE COT FROM THE LOWERED POSITION TO ALLOW FOR TRANSPORT OF THE PATIENT BACK TO THE WARD BED. THE PATIENT WAS NOT INJURED DURING THIS EVENT.

    Patient problems: Injury

    MAUDE report 8785405 ↗

Other STRETCHER, WHEELED devices

All STRETCHER, WHEELED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.