NA 0747Recalls & reported problems
Stryker Corporation · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker NA 0747 wheeled stretcher appears in 87 adverse event reports, with no deaths and 2 injuries mentioned. Reports mostly describe parts jamming, side rails dropping or failing to latch, the stretcher falling, and the mattress sliding out of place. One open recall began in May 2021 because stretchers meant for healthcare facilities were shipped to home users. Reports dropped from 12 in the previous 12 months to 1 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 87
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 87 adverse event reports that list this exact device (DI 07613327278187), received between Apr 23, 2018 and Jan 23, 2026 — classified as 2 injuries, 85 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (26 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 12, 2021Open, Classified
Medical devices intended for use in a healthcare facility were distributed to residential customers and may be in use in a home environment. The product s labeling accurately identifies the correct use environment, but home users may not be aware of or fully understand these instructions intended for healthcare professionals
Z-2205-2021 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Mechanical Jam26
- Device Fell15
- Positioning Failure15
- Fluid/Blood Leak13
- Mechanical Problem10
- Device Slipped3
- Positioning Problem2
- Sharp Edges2
Patient problems
- No Clinical Signs, Symptoms or Conditions59
- No Patient Involvement24
- No Consequences Or Impact To Patient2
- Insufficient Information1
- Laceration(s)1
- Limb Pain1
- Unspecified Musculoskeletal problem1
Recent adverse event reports
The 25 most recent of 87 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED IV POLE FALLS FROM UPRIGHT POSITION IN AN UNINTENDED DIRECTION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24168577 ↗MalfunctionOct 24, 2025Positioning Failure; Positioning Failure
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383715 ↗MalfunctionOct 24, 2025Mechanical Jam; Mechanical Jam
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383713 ↗MalfunctionOct 24, 2025Device Fell; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383703 ↗MalfunctionOct 24, 2025Device Slipped; Device Slipped
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED MATTRESS SLIDES OFF LITTER; UNSTABLE MATTRESS SUPPORT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383702 ↗MalfunctionJul 25, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608960 ↗MalfunctionJul 22, 2025Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569553 ↗MalfunctionJul 22, 2025Device Slipped; Device Slipped
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED MATTRESS SLIDES OFF LITTER. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22569550 ↗MalfunctionApr 25, 2025Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE OR STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. THERE WERE NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21909533 ↗MalfunctionApr 25, 2025Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WERE NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21909531 ↗MalfunctionJan 29, 2025Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF THE SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21267017 ↗MalfunctionJan 24, 2025Fluid/Blood Leak; Fluid/Blood Leak
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21229467 ↗MalfunctionJan 24, 2025Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DIFFICULT ENGAGING THE BRAKE OR STEER MECHANISMS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21229466 ↗MalfunctionOct 29, 2024Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20557750 ↗MalfunctionOct 25, 2024Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20539971 ↗MalfunctionOct 25, 2024Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20533519 ↗MalfunctionJul 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19819155 ↗MalfunctionJul 24, 2024Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE OR BRAKES DIFFICULT TO ENGAGE/DISENGAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19819141 ↗MalfunctionJul 24, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19819139 ↗MalfunctionApr 22, 2024Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19155518 ↗MalfunctionApr 22, 2024Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE OR BRAKES CANNOT BE ENGAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154575 ↗MalfunctionApr 22, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES(B)(4) MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154426 ↗MalfunctionJan 22, 2024Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18552249 ↗MalfunctionJan 22, 2024Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE OR BRAKES CANNOT BE ENGAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18552248 ↗MalfunctionJan 22, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18552247 ↗
Other STRETCHER, WHEELED devices
- Power-PRO 6506Stryker Corporation · STRETCHER, WHEELED5 deaths472 reports
- Procedural stretcher P8000HHill-Rom de México, S. de R.L. de C.V. · STRETCHER, WHEELED361 reports
- Prime Series 1115Stryker Corporation · STRETCHER, WHEELED1 death251 reports
- Prime Series 1115Stryker Corporation · STRETCHER, WHEELED1 recall199 reports
- Power-PRO 2, High Configuration 650705550001Stryker Corporation · STRETCHER, WHEELED2 recalls2 deaths99 reports
- NA 1089Stryker Corporation · STRETCHER, WHEELED175 reports
- NA 5050Stryker Corporation · STRETCHER, WHEELED1 recall99 reports
- Power-PRO 2 650700000000Stryker Corporation · STRETCHER, WHEELED2 recalls1 death42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page